Selena Cisana, Nicoletta Cerri, Alessandro Bosetti, Andrea Verzeletti1, Venusia Cortellini.
Abstract
AIM: To report on the successful analysis of amplicons obtained with PowerPlex® Fusion 6C System, a highly robust 27-plex genotyping kit developed for human identification laboratories, on the Applied Biosystems® 3500 Genetic Analyzer.Entities:
Mesh:
Substances:
Year: 2017 PMID: 28252872 PMCID: PMC5346900 DOI: 10.3325/cmj.2017.58.26
Source DB: PubMed Journal: Croat Med J ISSN: 0353-9504 Impact factor: 1.351
Figure 1Peak height ratio analysis. The graph displays the range of peak height ratio for serially dilution in two replicate polymerase chain reactions (PCRs) for each locus. The mean of peak height ratio are also shown by red plot. The homozygous peaks and short tandem repeats (STR) on the Y-chromosome (Y-STR) were excluded.
Amplification efficiency with increasing of urea concentration*
| Sample | Count | %Full |
|---|---|---|
| 0 M | 47 | 100 |
| 1 M | 47 | 100 |
| 2 M | 31 | 66 |
| 5 M | 5 | 11 |
*In “Sample” column are indicated the inhibitor concentrations. The untreated control is designed as “0 M”. “Count” column is the mean number of allele detected, and “%Full” column indicates the approximate percentage of a full profile for the DNA that was represented by the mean allele counts. The full profile contained 47 alleles total.
Figure 2Different effect of inhibitor on amplification efficiency for each locus. The graph shows the different percentage of allele detected for each locus in different inhibited samples (1M, 2M, and 5M of urea). The untreated control “0M” gave full alleles (not included).
Figure 3DNA mixture study. The average percentage of the alleles detected at a given mixture ratio is shown. Error bars represent the standard deviation of replicate results.
Case-type samples evaluated with the PowerPlex® Fusion 6C System and comparison of case-type samples previously genotyped with other kits (AmpFlSTR® Identifiler Plus, AmpFlSTR® NGM PCR kits, PowerPlex® 16 HS System or PowerPlex® ESX 17 System)
| Sample* | PowerPlex® Fusion 6C Alleles number | Previous typing kit | Alleles number | Concordance |
|---|---|---|---|---|
| Buccal swab | 47 full profile | AmpFlSTR® NGM PCR | 37 | YES |
| Blood sample | 37 full profile | AmpFlSTR® Identifiler® Plus | 25 | YES |
| Blood sample | 38 full profile | PowerPlex® 16 HS System | 3 | YES |
| Swab of glove | 24 | PowerPlex® 16 HS System | 27 | YES |
| Swab of glove | 17 | PowerPlex® 16 HS System | 18 | YES |
| Swab of glove | 24 | PowerPlex® 16 HS System | 13 | YES |
| Swab of glove | 7 | PowerPlex® 16 HS System | 0 | N/A |
| Swab of glasses | 11 | PowerPlex® ESX 17 System | 27 | YES |
| Swab of glasses | 9 | PowerPlex® ESX 17 System | 30 | YES |
| Swab of mask | 42 | PowerPlex® ESX 17 System | 13 | YES |
| Swab of mask | 20 | PowerPlex® ESX 17 System | 13 | YES |
| Swab of mask | 0 | PowerPlex® ESX 17 System | 0 | N/A |
| Swab of neckband | 16 | PowerPlex® ESX 17 System | 29 | YES |
| Human tissue | 45 full profile | AmpFlSTR® Identifiler® Plus | 27 | YES |
| Human tissue | 8 | AmpFlSTR® Identifiler® Plus | 19 | YES |
| Blood stain (cloth) | 47 full profile | AmpFlSTR® Identifiler® Plus | 29 | YES |
| Blood stain (cloth) | 45 full profile | AmpFlSTR® Identifiler® Plus | 29 | YES |
| Blood stain (cloth) | 46 full profile | AmpFlSTR® Identifiler® Plus | 30 | YES |
| Case-type mixtures | 69 | AmpFlSTR® Identifiler® Plus | 43 | YES |
| Case-type mixtures | 48 | AmpFlSTR® Identifiler® Plus | 40 | YES |
| Case-type mixtures | 70 | AmpFlSTR® Identifiler® Plus | 43 | YES |
| Case-type mixtures | 67 | AmpFlSTR® NGM PCR | 53 | YES |
*For each sample, the total number of alleles called with each kit is shown. Where loci had overlap between kits, they were examined for concordance.
Figure 4Precision of WEN ILS 500. The average size (bp) of each allele was plotted against the standard deviation observed across 16 ladders run during 5 injections on the same day on Applied Biosystems® 3500 Genetic Analyzer.