An approach for estimating measurement uncertainty in medical laboratories using data from long-term quality control and external quality assessment schemes.

Andrea Padoan1, Giorgia Antonelli1, Ada Aita1, Laura Sciacovelli1, Mario Plebani1.   

Abstract

BACKGROUND: The present study was prompted by the ISO 15189 requirements that medical laboratories should estimate measurement uncertainty (MU).
METHODS: The method used to estimate MU included the: a) identification of quantitative tests, b) classification of tests in relation to their clinical purpose, and c) identification of criteria to estimate the different MU components. Imprecision was estimated using long-term internal quality control (IQC) results of the year 2016, while external quality assessment schemes (EQAs) results obtained in the period 2015-2016 were used to estimate bias and bias uncertainty.
RESULTS: A total of 263 measurement procedures (MPs) were analyzed. On the basis of test purpose, in 51 MPs imprecision only was used to estimate MU; in the remaining MPs, the bias component was not estimable for 22 MPs because EQAs results did not provide reliable statistics. For a total of 28 MPs, two or more MU values were calculated on the basis of analyte concentration levels. Overall, results showed that uncertainty of bias is a minor factor contributing to MU, the bias component being the most relevant contributor to all the studied sample matrices.
CONCLUSIONS: The model chosen for MU estimation allowed us to derive a standardized approach for bias calculation, with respect to the fitness-for-purpose of test results. Measurement uncertainty estimation could readily be implemented in medical laboratories as a useful tool in monitoring the analytical quality of test results since they are calculated using a combination of both the long-term imprecision IQC results and bias, on the basis of EQAs results.

Entities:  

Keywords:  ISO 15189; external quality assessment schemes (EQAs); internal quality controls (IQC); measurement procedures (MPs); measurement uncertainty (MU); medical laboratory accreditation

Mesh:

Year:  2017        PMID: 28245184     DOI: 10.1515/cclm-2016-0896

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  3 in total

1.  Measurement uncertainty of blood ethanol concentration in drink-driving cases in an emergency laboratory.

Authors:  Yasemin Ustundağ; Kağan Huysal
Journal:  Biochem Med (Zagreb)       Date:  2017-10-15       Impact factor: 2.313

2.  ISO 15189 Accreditation: Navigation Between Quality Management and Patient Safety.

Authors:  Mario Plebani; Laura Sciacovelli
Journal:  J Med Biochem       Date:  2017-07-14       Impact factor: 3.402

Review 3.  The top-down approach to measurement uncertainty: which formula should we use in laboratory medicine?

Authors:  Flávia Martinello; Nada Snoj; Milan Skitek; Aleš Jerin
Journal:  Biochem Med (Zagreb)       Date:  2020-04-15       Impact factor: 2.313

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.