An uncovered risk factor of sonothrombolysis: Substantial fluctuation of ultrasound transmittance through the human skull.

Zuojun Wang1, Teppei Komatsu2, Hidetaka Mitsumura2, Norio Nakata1, Takeki Ogawa3, Yasuyuki Iguchi2, Masayuki Yokoyama4.   

Abstract

Sonothrombolysis is one of the most feasible methods for enhancing clot lysis with a recombinant tissue plasminogen activator (rt-PA) in cases of acute ischemic strokes. For safe and efficient clinical practices of sonothrombolysis, accurate estimation of ultrasound transmittance through the human skull is critical. Previously, we reported substantial and periodic fluctuation of ultrasound transmittance through a bone-phantom plate following changes to ultrasound frequency, the thickness of the bone-phantom plate, and the distance between a transducer and the bone-phantom plate. In the present study, we clarify the transmittance behavior of medium-frequency ultrasound (from 400kHz to 600kHz) through the human skull, and examine reduction of the transmittance fluctuation. For the study, we measured transmittance of sinusoidal ultrasound waves at 400kHz, 500kHz, and 600kHz at 13 temple spots on 3 human skulls by changing the distance between a transducer and the skull bone, and found substantial and periodic fluctuation in the transmittance behaviors for these sinusoidal voltage excitations. Degrees of the fluctuation varied depending on the measurement spots. A fluctuation ratio between the maximum transmittance and the minimum transmittance reached 3 in some spots. This large transmittance fluctuation is considered to be a risk factor for sonothrombolysis therapies. We examined a modulated ultrasound wave to reduce the fluctuation, and succeeded in obtaining considerable reduction. The average fluctuation ratios for 400-kHz, 500-kHz, and 600-kHz waves were 2.38, 2.38, and 2.07, respectively. We successfully reduced the ratio to 1.72 by using a periodic selection of random frequency (PSRF)-type of modulation wave. The thus obtained results indicate that attention to the fluctuation in ultrasound transmittance through the skull is necessary for safe and effective sonothrombolysis therapies, and that modulated ultrasound waves constitute a powerful method for reducing the risk of fluctuation.
Copyright © 2017 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Acute ischemic stroke; Modulation; Sonothrombolysis; Transcranial transmittance; Ultrasound

Mesh:

Substances:

Year:  2017        PMID: 28242510     DOI: 10.1016/j.ultras.2017.02.012

Source DB:  PubMed          Journal:  Ultrasonics        ISSN: 0041-624X            Impact factor:   2.890


  2 in total

1.  Acoustic properties across the human skull.

Authors:  Thomas S Riis; Taylor D Webb; Jan Kubanek
Journal:  Ultrasonics       Date:  2021-10-21       Impact factor: 2.890

2.  Clinical Importance of Temporal Bone Features for the Efficacy of Contrast-Enhanced Sonothrombolysis: a Retrospective Analysis of the NOR-SASS Trial.

Authors:  Vojtech Novotny; Aliona Nacu; Christopher E Kvistad; Annette Fromm; Gesche F Neckelmann; Andrej N Khanevski; Haakon Tobro; Ulrike Waje-Andreassen; Halvor Naess; Lars Thomassen; Nicola Logallo
Journal:  Transl Stroke Res       Date:  2017-11-08       Impact factor: 6.829

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.