Abstract

Sintered hydroxyl-apatite implants form very tight bonds with living bone but are susceptible to fatigue failure. Plasma-sprayed apatite coatings on titanium substrates overcome the fatigue problem. The static tensile substrate bond strength of the apatite coating is in excess of 85 megapascals (MPa) (12,000 psi). In a plug implant study designed to discount mechanical retention, a bone bonding shear strength of 64 MPa (9280 psi) was achieved, comparable to the strength of cortical bone. Histologic sections confirm the close bonding between apatite coating and living bone. In a canine total hip arthroplasty study, the apatite-coated implants proved far superior to the uncoated controls. Uncoated prostheses were surrounded by fibrous tissue and were easily extracted from the femur at any postoperative time. The apatite-coated implants were rigidly fixed within three weeks with demonstrable bone formation up to the implant surface. Bony defects up to 2 mm in depth were filled with bone within six weeks. The hypothetical mechanism of bone bonding is chemical. Hydroxyl-apatite coatings permit an implant fixation far superior to current methods using either cemented or cementless techniques. The plan is to study a human total hip prosthesis with hydroxyl-apatite coating for chemical fixation to bone.

Entities:  

Mesh:

Substances:

Year:  1987        PMID: 2824115

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  44 in total

1.  Orthopedic prosthesis fixation.

Authors:  J B Park
Journal:  Ann Biomed Eng       Date:  1992       Impact factor: 3.934

2.  Bilayered calcium phosphate coating to promote osseointegration of a femoral stem prosthesis.

Authors:  E Goyenvalle; N J M Guyen; E Aguado; N Passuti; G Daculsi
Journal:  J Mater Sci Mater Med       Date:  2003-03       Impact factor: 3.896

3.  An experimental study in X-ray spectroscopy of the zirconium (Ca-PSZ) - bone interface. Microanalytic evaluation of the osteogenetic response.

Authors:  M Mattioli-Belmonte; P Mengucci; N Specchia; G Gobbi; S Dubini; L Simonelli; F Greco; G Majni; G Biagini; C Rizzoli
Journal:  J Mater Sci Mater Med       Date:  1997-02       Impact factor: 3.896

4.  Topographical characterization and microstructural interface analysis of vacuum-plasma-sprayed titanium and hydroxyapatite coatings on carbon fibre-reinforced poly(etheretherketone).

Authors:  S W Ha; A Gisep; J Mayer; E Wintermantel; H Gruner; M Wieland
Journal:  J Mater Sci Mater Med       Date:  1997-12       Impact factor: 3.896

5.  Morphology and immersion behavior of plasma-sprayed hydroxyapatite/bioactive glass coatings.

Authors:  S J Ding; C P Ju; J H Lin
Journal:  J Mater Sci Mater Med       Date:  2000-03       Impact factor: 3.896

6.  Formation of apatitic calcium phosphates in a Na-K-phosphate solution of pH 7.4.

Authors:  A C Tas; F Aldinger
Journal:  J Mater Sci Mater Med       Date:  2005-02       Impact factor: 3.896

7.  Correlation between primary stability and bone healing of surface treated titanium implants in the femoral epiphyses of rabbits.

Authors:  Julie Rozé; Alain Hoornaert; Pierre Layrolle
Journal:  J Mater Sci Mater Med       Date:  2014-05-13       Impact factor: 3.896

8.  Bone remodeling and hydroxyapatite resorption in coated primary hip prostheses.

Authors:  Alphons J Tonino; Bart C H van der Wal; Ide C Heyligers; Bernd Grimm
Journal:  Clin Orthop Relat Res       Date:  2008-10-15       Impact factor: 4.176

9.  Antibacterial and bioactive silver-containing Na2O x CaO x 2SiO2 glass prepared by sol-gel method.

Authors:  M Catauro; M G Raucci; F De Gaetano; A Marotta
Journal:  J Mater Sci Mater Med       Date:  2004-07       Impact factor: 3.896

10.  Is the bone-bonding ability of a cementless total hip prosthesis enhanced by alkaline and heat treatments?

Authors:  Kazutaka So; Ayumi Kaneuji; Tadami Matsumoto; Shuichi Matsuda; Haruhiko Akiyama
Journal:  Clin Orthop Relat Res       Date:  2013-12       Impact factor: 4.176

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.