Zenith Bifurcated Iliac Side Branch Device: Mid-term Results and Assessment of Risk Factors for Intraoperative Thrombosis.

Charline Delay1, Sébastien Deglise2, Anne Lejay1, Yannick Georg1, Mathieu Roussin1, Mickaël Schaeffer3, François Saucy2, Fabien Thaveau1, Jean-Marc Corpataux2, Nabil Chakfe4.   

Abstract

BACKGROUND: The aim of this study is to evaluate the short- and mid-term results of the Zenith bifurcated iliac side branch device (ZBIS) in the treatment of common iliac artery (CIA) aneurysms, and to assess risk factors for intraoperative internal iliac artery (IIA) thrombosis.
METHODS: All patients who underwent endovascular treatment of either an isolated CIA aneurysm or an aortoiliac aneurysm using the ZBIS device in the departments of vascular surgery of Strasbourg (France) and Lausanne (Switzerland) between January 2010 and December 2014 were retrospectively collected.
RESULTS: Thirty-one implantations were performed: 30 patients underwent 31 endovascular CIA aneurysm treatments with the ZBIS device. Mean operative time was 188 min. Technical success was obtained in 26 implantations (84%). In 5 implantations (16%), the final angiogram revealed an IIA thrombosis. Thirty-day mortality was 3.2%. Thirty-day morbidity was 13.3%. Mean follow-up was 15 months. Overall survival was 96% at 1 year and 89% at 2 years. In intention-to-treat analysis, primary patency of the internal iliac side branch was 84% at 1 year and 76% at 2 years (5 peroperative IIA occlusions and 1 late occlusion). Freedom from reintervention was 89% at 1 and 2 years. One case of type III endoleak and 2 cases of type II endoleaks were identified. Only type III endoleak required an additional intervention with a covered stent. Aneurysm diameter decreased in 15 implantations (48%) and remained stable in 16 implantations (52%). Clinical, radiological, and peroperative parameters were analyzed to identify risk factor for intraoperative thrombosis of the internal iliac side branch. Notion of intraoperative difficulties (any additional procedure that was not initially planned and increasing the operating time) appeared as a risk factor in multivariate analysis (P < 0.01, standard deviation 1.27, odds ratio 30.6).
CONCLUSIONS: The main findings of our study is that the procedure can be difficult to perform in particular conditions and can lead to peroperative failure in these cases, highlighting the need for adequate patients screening. When technical success is obtained, outcomes can be considered as satisfactory.
Copyright © 2017 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2017        PMID: 28238918     DOI: 10.1016/j.avsg.2016.08.048

Source DB:  PubMed          Journal:  Ann Vasc Surg        ISSN: 0890-5096            Impact factor:   1.466


  2 in total

1.  Preservation of internal iliac artery flow during endovascular aortic aneurysm repair in a patient with bilateral absence of common iliac artery.

Authors:  Minh-Anh Pham; Thanh-Phong Le
Journal:  J Vasc Surg Cases Innov Tech       Date:  2021-01-28

2.  The outcomes of internal iliac artery preservation during endovascular or open surgery treatment for aortoiliac aneurysms.

Authors:  Rafael de Athayde Soares; Marcelo Fernando Matielo; Francisco Cardoso Brochado; Amanda Thurler Palomo; Rodrigo Andrade Lourenço; Caroline Tanaka; Roberto Sacilotto
Journal:  J Vasc Bras       Date:  2020-12-11
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.