| Literature DB >> 28235530 |
Dhritiman Chakrabarti1, Rohini Mrugank Surve2, Deepti Bs3, Dheeraj Masapu4.
Abstract
Bispectral index is an accepted depth of anaesthesia monitor for guiding intraoperative hypnotic agent administration. Frontalis EMG displayed on BIS monitor may increase due to twitching of frontalis muscle. EMG increases are also known to cause artefactual increases in BIS values. We report a case of artefactual increase of EMG and subsequently BIS values, due to electrical artefact from cranial nerve stimulator being used to identify the facial nerve. An explanation of the effect of stimulator signal on BIS EMG and BIS values has been provided.Entities:
Keywords: Artefact; Bispectral index; Facial nerve monitoring
Mesh:
Substances:
Year: 2016 PMID: 28235530 DOI: 10.1016/j.jclinane.2016.11.005
Source DB: PubMed Journal: J Clin Anesth ISSN: 0952-8180 Impact factor: 9.452