Measurement of cochlear length using the 'A' value for cochlea basal diameter: A feasibility study.

Nicholas L Deep1, Brittany E Howard1, Sarah O Holbert1, Joseph M Hoxworth2, David M Barrs1.   

Abstract

OBJECTIVES: To determine whether the cochlea basal diameter (A value) measurement can be consistently and precisely obtained from high-resolution temporal bone imaging for use in cochlear length estimation.
METHODS: A feasibility study at a tertiary referral center was performed using the temporal bone CTs of 40 consecutive patients. The distance from the round window to the lateral wall was measured for each cochlea by two independent reviewers, a neuroradiologist and an otolaryngologist. The interrater reliability was calculated using the intraclass correlation coefficient (ICC) and the Bland-Altman plot.
RESULTS: Forty patients (19 males, 21 females) for a total of 80 cochleae were included. Interrater reliability on the same ear had a high level of agreement by both the ICC and the Bland-Altman plot. ICCs were 0.90 (95% CI: 0.82, 0.94) for the left ear and 0.96 (95% CI: 0.92, 0.98) for the right ear. Bland-Altman plot confirmed interrater reliability with all 96% of measurements falling within the 95% limits of agreement.
CONCLUSIONS: Measurement between the round window and lateral cochlear wall can be consistently and reliably obtained from high-resolution temporal bone CT scans. Thus, it is feasible to utilize this method to estimate the cochlear length of patients undergoing cochlear implantation.

Entities:  

Keywords:  A value; Cochlear duct length; Cochlear implantation

Mesh:

Year:  2017        PMID: 28235385     DOI: 10.1080/14670100.2017.1292689

Source DB:  PubMed          Journal:  Cochlear Implants Int        ISSN: 1467-0100


  3 in total

1.  CT-scan contouring technique allows for direct and reliable measurements of the cochlear duct length: implication in cochlear implantation with straight electrode-arrays.

Authors:  Thi Hau Vu; Chiara Perazzini; Mathilde Puechmaille; Aurélie Bachy; Aurélien Mulliez; Louis Boyer; Thierry Mom; Jean Gabrillargues
Journal:  Eur Arch Otorhinolaryngol       Date:  2019-04-22       Impact factor: 2.503

2.  Is Cochlear Length Related to Congenital Sensorineural Hearing Loss: Preliminary Data.

Authors:  Mehmet Bilgin Eser; Başak Atalay; Mahmut Tayyar Kalcıoğlu
Journal:  J Int Adv Otol       Date:  2021-01       Impact factor: 1.017

3.  A novel cochlear measurement that predicts inner-ear malformation.

Authors:  Tawfiq Khurayzi; Fida Almuhawas; Abdulrahman Alsanosi; Yassin Abdelsamad; Úna Doyle; Anandhan Dhanasingh
Journal:  Sci Rep       Date:  2021-04-01       Impact factor: 4.379