Evaluation of interarm blood pressure differences using the Microlife WatchBP Office in a clinical setting.

Christoffer Krogager1, Esben Laugesen, Niklas B Rossen, Per L Poulsen, Mogens Erlandsen, Klavs W Hansen.   

Abstract

OBJECTIVE: The aim of this study was to evaluate the usefulness of Microlife WatchBP Office and the effect of increasing the number of measurements in the clinical evaluation of systolic interarm difference (IAD). PATIENTS AND METHODS: Office blood pressure was measured simultaneously on both arms in 339 patients (85% diabetic) using the Microlife WatchBP Office, a fully automatic, oscillometric device. The patients included were all scheduled for ambulatory blood pressure measurement at the outpatient clinic of endocrinology at Silkeborg Regional Hospital, Denmark. Two successive sets of three individual measurements were made. A statistical analysis of variance was carried out on the measurements.
RESULTS: In the first set of measurements, the mean IAD was -0.3 mmHg and the prevalence of IAD greater than or equal to 10 mmHg was 9.1%. Only 7.6% of the patients with an IAD less than 10 mmHg in the first set of measurements had an IAD greater than or equal to 10 mmHg in the second set of measurements. The 95% limits of agreement for the mean IAD for a single set of three measurements were ±13.16 mmHg. The probability of detecting an IAD more than 10 mmHg only increased slightly with an increasing number of measurements.
CONCLUSION: A single set of triplicate measurements using Microlife WatchBP is an acceptable method for evaluating IAD as more measurements do little to improve the probability of detecting an IAD more than 10 mmHg because of high intraindividual variation.

Entities:  

Mesh:

Year:  2017        PMID: 28234747     DOI: 10.1097/MBP.0000000000000246

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  2 in total

1.  The cuffless SOMNOtouch NIBP device shows poor agreement with a validated oscillometric device during 24-h ambulatory blood pressure monitoring.

Authors:  Jakob Nyvad; Kent L Christensen; Niels Henrik Buus; Mark Reinhard
Journal:  J Clin Hypertens (Greenwich)       Date:  2020-12-17       Impact factor: 3.738

2.  Clinical relevance of double-arm blood pressure measurement and prevalence of clinically important inter-arm blood pressure differences in Indian primary care.

Authors:  Gurpreet S Wander; Sinead T J McDonagh; M Srinivasa Rao; R Alagesan; J C Mohan; Ajit Bhagwat; A K Pancholia; M Viswanathan; Manoj Bhavrilal Chopda; A Purnanand; P L N Kapardhi; Arun R Vadavi; R Selvaraj; Pankaj Aneja; Suhas Hardas; Neil Bordoloi; N Sivakadaksham; Nilesh Goswami; Christopher E Clark; Willem J Verberk
Journal:  J Clin Hypertens (Greenwich)       Date:  2022-07-10       Impact factor: 2.885

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.