Point-of-care testing for coagulation function: CoaguChek® XS System versus standard laboratory testing in pediatric patients with normal and abnormal coagulation function.

Hiromi Kako1,2, Vidya T Raman3,4, Dmitry Tumin3, Julie Rice3, Joseph D Tobias3,4.   

Abstract

PURPOSE: Intraoperative abnormalities of coagulation function may occur for various reasons. In most scenarios, treatment is directed by laboratory parameters. Unfortunately, standard laboratory testing may take 1-2 h. The purpose of the current study was to evaluate a point-of-care testing device (CoaguChek® XS System) in pediatric patients.
METHODS: Patients ranging in age from 2 to 18 years, undergoing posterior spinal fusion (PSF) or cardiac surgery using cardiopulmonary bypass (CPB) were eligible for inclusion. After CPB and/or the surgical procedure, 2.8 ml of blood was obtained and simultaneously tested on both the standard laboratory apparatus and the CoaguChek® XS System.
RESULTS: The study cohort consisted of 100 patients (50 PSF and 50 cardiac cases) with 13 cases excluded, leaving 87 patients (49 PSF and 38 cardiac cases) for analysis. In PSF cases, reference laboratory international normalized ratio (INR) ranged from 0.98 to 1.77 while CoaguChek® XS INR ranged from 1.0 to 1.3. The correlation coefficient was 0.69. The results of the Bland-Altman analysis showed a bias of 0.09, precision of 0.1, and 95% limits of agreement ranging from -0.11 to 0.28. In cardiac cases, reference INR ranged from 1.68 to 14.19, while CoaguChek® XS INR ranged from 1.4 to 7.9. The correlation coefficient was 0.35. The results of the Bland-Altman analysis showed a bias of -1.8, precision of 2.1, and 95% limits of agreement ranging from -6.0 to 2.4.
CONCLUSIONS: INR values obtained from CoaguChek® XS showed a moderate correlation with reference laboratory values within the normal range. However, in the presence of coagulopathy, the discrepancy was significantly greater, thereby making the CoaguChek® XS clinically unreliable.

Entities:  

Keywords:  CoaguChek® XS System; Coagulation function; International normalized ratio; Point-of-care testing

Mesh:

Year:  2017        PMID: 28213667     DOI: 10.1007/s00540-017-2317-3

Source DB:  PubMed          Journal:  J Anesth        ISSN: 0913-8668            Impact factor:   2.078


  6 in total

Review 1.  Practice guidelines for perioperative blood management: an updated report by the American Society of Anesthesiologists Task Force on Perioperative Blood Management*.

Authors: 
Journal:  Anesthesiology       Date:  2015-02       Impact factor: 7.892

2.  Agreement between Coaguchek XS and STA-R Evolution (Hepato Quick) INR results depends on the level of INR.

Authors:  Joseph S Biedermann; Frank W G Leebeek; Peter N Buhre; Sacha de Lathouder; Jan P F Barends; Moniek P M de Maat; Felix J M van der Meer; Marieke J H A Kruip
Journal:  Thromb Res       Date:  2015-07-05       Impact factor: 3.944

3.  Comparison of the international normalized ratio levels obtained by portable coagulometer and laboratory in a clinic specializing in oral anticoagulation.

Authors:  Z M Meneghelo; C M Q Barroso; I L Liporace; A P Cora
Journal:  Int J Lab Hematol       Date:  2015-03-09       Impact factor: 2.877

4.  The accuracy of coagulation tests during spinal fusion and instrumentation.

Authors:  T T Horlocker; G A Nuttall; M B Dekutoski; S C Bryant
Journal:  Anesth Analg       Date:  2001-07       Impact factor: 5.108

5.  Comparison of point-of-care testing (POCT): i-STAT(®) international normalized ratio (INR) vs reference laboratory INR in pediatric patients undergoing major surgery.

Authors:  Nelly Spielmann; Jacqueline Y Mauch; Caveh Madjdpour; Markus Schmugge; Manuela Albisetti; Markus Weiss; Thorsten Haas
Journal:  Paediatr Anaesth       Date:  2011-05-04       Impact factor: 2.556

Review 6.  Multidisciplinary approach to the challenge of hemostasis.

Authors:  Jerrold H Levy; Richard P Dutton; J Claude Hemphill; Aryeh Shander; David Cooper; Michael J Paidas; Craig M Kessler; John B Holcomb; Jeffrey H Lawson
Journal:  Anesth Analg       Date:  2009-12-10       Impact factor: 5.108

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.