Imaging evaluating of the implant/bone interface-an in vitro radiographic study.

Michele M Vidor1, Gabriela S Liedke2, Mariana B Vizzotto1, Heraldo L D da Silveira1, Priscila F da Silveira1, Cristiano W Araujo3,4, Heloisa E D da Silveira1.   

Abstract

OBJECTIVES: To analyze the diagnostic accuracy of conventional and digital radiographic images and the impact of digital filters in evaluating the bone-implant interface.
METHODS: Titanium implants were inserted into 74 fresh bovine ribs blocks, 37 fitting tight to the bone walls (simulating the existence of osseointegration) and 37 with a gap of 0.125 mm (simulating a failure in the osseointegration process). Periapical radiographs were taken with conventional film and two phosphor plate systems [VistaScan® (Dürr Dental, Bietigheim-Bissingen, Germany) and Express® (Instrumentarium, Tuusula, Finland)]. Digital radiographs were investigated with and without enhancement filters. Three blinded examiners assessed the images for the presence of juxtaposition in the bone-implant interface using a five-point Likert scale. Sensitivity, specificity and the area under the receiver-operating characteristic curve (AUC) and its 95% confidence interval (CI) were calculated for each variable. Intraexaminer and interexaminer agreements were analyzed using Kendall's concordance test.
RESULTS: Intraexaminer and interexaminer agreements were >0.80 for both digital and conventional images. Conventional radiographs (AUC = 0.963/CI = 0.891 to 0.993) and digital images with high enhancement filters such as Caries2 (AUC = 0.964/CI = 0.892 to 0.993), Endo (AUC = 0.952/CI = 0.875 to 0.988) and Sharpen3 (AUC = 0.894/CI = 0.801 to 0.954) showed the greatest accuracy for evaluating the bone-implant interface. Original images from both digital systems and the further enhancement filters tested showed low sensitivity for the diagnosis task tested.
CONCLUSIONS: Conventional radiographs or digital radiographs with high-pass filters could help enhance diagnosis on implant-bone interface.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 28211288      PMCID: PMC5595032          DOI: 10.1259/dmfr.20160296

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  15 in total

Review 1.  Radiographic determinants of implant performance.

Authors:  M S Reddy; I C Wang
Journal:  Adv Dent Res       Date:  1999-06

2.  A comparison of cone beam computed tomography and conventional periapical radiography at detecting peri-implant bone defects.

Authors:  Meetal Dave; Jonathan Davies; Ron Wilson; Richard Palmer
Journal:  Clin Oral Implants Res       Date:  2012-03-27       Impact factor: 5.977

3.  Accuracy of enhancement filters in measuring in vitro peri-implant bone level.

Authors:  Sergio Lins de Azevedo Vaz; Frederico Sampaio Neves; Evandro Portela Figueirêdo; Francisco Haiter-Neto; Paulo Sérgio Flores Campos
Journal:  Clin Oral Implants Res       Date:  2012-07-04       Impact factor: 5.977

4.  A digital radiographic artifact: A clinical report.

Authors:  Daniel M Schweitzer; Robert W Berg
Journal:  J Prosthet Dent       Date:  2010-06       Impact factor: 3.426

5.  Subjective image quality of digitally filtered radiographs acquired by the Dürr Vistascan system compared with conventional radiographs.

Authors:  S Yalcinkaya; A Künzel; R Willers; M Thoms; J Becker
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2006-01-19

Review 6.  Hardware and software artifacts in storage phosphor radiography.

Authors:  J W Oestmann; M Prokop; C M Schaefer; M Galanski
Journal:  Radiographics       Date:  1991-09       Impact factor: 5.333

Review 7.  Work flow with digital intraoral radiography: a systematic review.

Authors:  Ann Wenzel; Anne Møystad
Journal:  Acta Odontol Scand       Date:  2010-03       Impact factor: 2.331

8.  Uberschwinger artefact in computed radiographs.

Authors:  T H Tan; A E Boothroyd
Journal:  Br J Radiol       Date:  1997-04       Impact factor: 3.039

9.  The impact of digital image processing artefacts mimicking pathological features associated with restorations.

Authors:  D Brettle; F Carmichael
Journal:  Br Dent J       Date:  2011-08-26       Impact factor: 1.626

10.  Task-specific enhancement filters in storage phosphor images from the Vistascan system for detection of proximal caries lesions of known size.

Authors:  Francisco Haiter-Neto; Marcia Spinelli Casanova; Morten Frydenberg; Ann Wenzel
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2009-01
View more
  2 in total

1.  Photoacoustic imaging of posterior periodontal pocket using a commercial hockey-stick transducer.

Authors:  Lei Fu; Chen Ling; Zhicheng Jin; Jessica Luo; Jorge Palma-Chavez; Zhuohong Wu; Jingcheng Zhou; Jiajing Zhou; Brian Donovan; Baiyan Qi; Aditya Mishra; Tengyu He; Jesse V Jokerst
Journal:  J Biomed Opt       Date:  2022-05       Impact factor: 3.758

Review 2.  Diagnostic accuracy of imaging examinations for peri-implant bone defects around titanium and zirconium dioxide implants: A systematic review and meta-analysis.

Authors:  Mariana Murai Chagas; Solange Kobayashi-Velasco; Thais Gimenez; Marcelo Gusmão Paraiso Cavalcanti
Journal:  Imaging Sci Dent       Date:  2021-11-18
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.