Literature DB >> 28200212

Head-to-head comparison of adaptive statistical and model-based iterative reconstruction algorithms for submillisievert coronary CT angiography.

Dominik C Benz1, Tobias A Fuchs1, Christoph Gräni1, Annina A Studer Bruengger1, Olivier F Clerc1, Fran Mikulicic1, Michael Messerli1, Julia Stehli1, Mathias Possner1, Aju P Pazhenkottil1, Oliver Gaemperli1, Philipp A Kaufmann1, Ronny R Buechel1.   

Abstract

Aims: Iterative reconstruction (IR) algorithms allow for a significant reduction in radiation dose of coronary computed tomography angiography (CCTA). We performed a head-to-head comparison of adaptive statistical IR (ASiR) and model-based IR (MBIR) algorithms to assess their impact on quantitative image parameters and diagnostic accuracy for submillisievert CCTA. Methods and results: CCTA datasets of 91 patients were reconstructed using filtered back projection (FBP), increasing contributions of ASiR (20, 40, 60, 80, and 100%), and MBIR. Signal and noise were measured in the aortic root to calculate signal-to-noise ratio (SNR). In a subgroup of 36 patients, diagnostic accuracy of ASiR 40%, ASiR 100%, and MBIR for diagnosis of coronary artery disease (CAD) was compared with invasive coronary angiography. Median radiation dose was 0.21 mSv for CCTA. While increasing levels of ASiR gradually reduced image noise compared with FBP (up to - 48%, P < 0.001), MBIR provided largest noise reduction (-79% compared with FBP) outperforming ASiR (-59% compared with ASiR 100%; P < 0.001). Increased noise and lower SNR with ASiR 40% and ASiR 100% resulted in substantially lower diagnostic accuracy to detect CAD as diagnosed by invasive coronary angiography compared with MBIR: sensitivity and specificity were 100 and 37%, 100 and 57%, and 100 and 74% for ASiR 40%, ASiR 100%, and MBIR, respectively.
Conclusion: MBIR offers substantial noise reduction with increased SNR, paving the way for implementation of submillisievert CCTA protocols in clinical routine. In contrast, inferior noise reduction by ASiR negatively affects diagnostic accuracy of submillisievert CCTA for CAD detection. Published on behalf of the European Society of Cardiology. All rights reserved.
© The Author 2017. For permissions, please email: journals.permissions@oup.com.

Entities:  

Keywords:  adaptive statistical iterative reconstruction; model-based iterative reconstruction; submillisievert coronary CT angiography

Mesh:

Year:  2018        PMID: 28200212     DOI: 10.1093/ehjci/jex008

Source DB:  PubMed          Journal:  Eur Heart J Cardiovasc Imaging        ISSN: 2047-2404            Impact factor:   6.875


  6 in total

1.  Coronary artery volume index: a novel CCTA-derived predictor for cardiovascular events.

Authors:  Georgios Benetos; Ronny R Buechel; Marisa Gonçalves; Dominik C Benz; Elia von Felten; Georgios P Rampidis; Olivier F Clerc; Michael Messerli; Andreas A Giannopoulos; Cathérine Gebhard; Tobias A Fuchs; Aju P Pazhenkottil; Philipp A Kaufmann; Christoph Gräni
Journal:  Int J Cardiovasc Imaging       Date:  2020-01-01       Impact factor: 2.357

2.  Clinical value of a new generation adaptive statistical iterative reconstruction (ASIR-V) in the diagnosis of pulmonary nodule in low-dose chest CT.

Authors:  Hui Tang; Zhentang Liu; Zhijun Hu; Taiping He; Dou Li; Nan Yu; Yongjun Jia; Hong Shi
Journal:  Br J Radiol       Date:  2019-09-06       Impact factor: 3.039

3.  Diagnostic accuracy of coronary opacification derived from coronary computed tomography angiography to detect ischemia: first validation versus single-photon emission computed tomography.

Authors:  Dominik C Benz; Fran Mikulicic; Christoph Gräni; Marvin Grossmann; Andreas A Giannopoulos; Michael Messerli; Catherine Gebhard; Oliver Gaemperli; Ronny R Buechel; Philipp A Kaufmann; Aju P Pazhenkottil
Journal:  EJNMMI Res       Date:  2017-11-25       Impact factor: 3.138

4.  Diagnostic accuracy of low-radiation coronary computed tomography angiography with low tube voltage and knowledge-based model reconstruction.

Authors:  Joohee Lee; Tae Hoon Kim; Byoung Kwon Lee; Young Won Yoon; Hyuck Moon Kwon; Bum Kee Hong; Pil-Ki Min; Eui-Young Choi; Chi Suk Oh; Chul Hwan Park
Journal:  Sci Rep       Date:  2019-02-04       Impact factor: 4.379

5.  Prior-image-based CT reconstruction using attenuation-mismatched priors.

Authors:  Hao Zhang; Dante Capaldi; Dong Zeng; Jianhua Ma; Lei Xing
Journal:  Phys Med Biol       Date:  2021-03-17       Impact factor: 3.609

6.  Coronary computed tomography angiography using model-based iterative reconstruction algorithms in the detection of significant coronary stenosis: how the plaque type influences the diagnostic performance.

Authors:  Antonio Vizzuso; Riccardo Righi; Aldo Carnevale; Michela Zerbini; Giorgio Benea; Melchiore Giganti
Journal:  Pol J Radiol       Date:  2019-12-09
  6 in total