Structural, morphological, compositional, and mechanical changes of palatal implants after use: a retrieval analysis.

Marc Schätzle1, Spiros Zinelis2, Goran Markic1, George Eliades2, Theodore Eliades1.   

Abstract

PURPOSE: The aim of this study was to characterize the surface, elemental, and mechanical alterations of orthodontic palatal implants after intraoral aging. MATERIALS AND
METHOD: Nineteen consecutively retrieved implants (RET) after orthodontic treatment and three unused implants used as control (CON) were included in this study. Both groups were characterized non-destructively by Stereomicroscopy, Optical Profilometry (Sa, Sq, Sz, Sc), and SEM/EDX analysis and then destructively after metallogaphic preparation employing instrumented indentation testing (HM, EIT, ηIT, and HV) and SEM/EDX at bone-implant interface.
RESULTS: All retrieved implants showed a loss of gloss with the formation of bone-like formation on the majority of them. However, no differences in surface roughness parameters were identified between macroscopically intact and retrieved regions of implants. The elements precipitated on the surface were O, C, Ca, and P while traces of Na, K, Al, S, Cl, and Mg were also identified. The surface of control sample is characterized by small pits while only Ti and Al traces were identified by EDX analysis. The presence of all the aforementioned elements apart from Ti and Al on the retrieved implants' surface should be appended to the contact of implant with bone and biological fluids while Interfacial analysis revealed a well-formed bone-implant interface. However, no significant differences were found for all mechanical properties tested between RET and CON groups.
CONCLUSIONS: The results of this study indicate that retrieved palatal implant surface has undergone morphological and elemental alterations probably associated with the osseointegration process during service. Insertion and functional loading did not affect the mechanical properties of implants tested.
© The Author 2017. Published by Oxford University Press on behalf of the European Orthodontic Society. All rights reserved. For permissions, please email: journals.permissions@oup.com

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 28199512     DOI: 10.1093/ejo/cjx001

Source DB:  PubMed          Journal:  Eur J Orthod        ISSN: 0141-5387            Impact factor:   3.075


  1 in total

1.  Side effects of twistflex retainers-3D evaluation of tooth movement after retainer debonding.

Authors:  Isabel Knaup; Jenny Rosa Bartz; Ulrike Schulze-Späte; Rogério Bastos Craveiro; Christian Kirschneck; Michael Wolf
Journal:  J Orofac Orthop       Date:  2020-12-01       Impact factor: 1.938

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.