Luke Porra1, Hans Swan2, Chien Ho1.
Abstract
Introduction: Acoustic Radiation Force Impulse (ARFI) Quantification measures shear wave velocities (SWVs) within the liver. It is a reliable method for predicting the severity of liver fibrosis and has the potential to assess fibrosis in any part of the liver, but previous research has found ARFI quantification in the right lobe more accurate than in the left lobe. A lack of standardised applied transducer force when performing ARFI quantification in the left lobe of the liver may account for some of this inaccuracy. The research hypothesis of this present study predicted that an increase in applied transducer force would result in an increase in SWVs measured.Entities:
Keywords: ARFI; applied transducer force; left lobe; liver
Year: 2015 PMID: 28191250 PMCID: PMC5024960 DOI: 10.1002/j.2205-0140.2015.tb00208.x
Source DB: PubMed Journal: Australas J Ultrasound Med ISSN: 1836-6864
Figure 1a) A mechanical device used to regulate applied transducer force. b) Spirit level used to ensure crossbar was horizontal.
Figure 2Schematic of mechanical device. The amount of applied transducer force applied to the abdomen increased when the mass at the 400 mm point from fulcrum was increased. Standardised levels of mass at 400 mm from fulcrum allowed the regulation of applied transducer force to the abdomen.
Participant demographics.
| Male / Female | 3 / 25 |
|---|---|
| Age of participant (years) | 39 ± 14.1 (18−70) |
| BMI (kg/m2) | 25.6 ± 4.0 (19−35) |
| Thickness of abdominal wall (mm) | 32.8 ± 13.1 (14−76) |
Mean, SD and range for each demographic
Measurement success rate for each applied transducer force.
| Force (g) | Participants | SWV measurements Total Valid Invalid | Success rate (SR) | ||
|---|---|---|---|---|---|
| 5 | 28 | 143 | 140 | 3 | 0.98 (0.71−1.00) |
| 249 | 28 | 145 | 140 | 5 | 0.97 (0.71−1.00) |
| 495 | 28 | 144 | 140 | 4 | 0.98 (0.71−1.00) |
| 1000 | 28 | 140 | 140 | 0 | 1.00 (1.00−1.00) |
| 2015 | 26 | 140 | 130 | 0 | 1.00 (1.00−1.00) |
| 2972 | 24 | 140 | 120 | 0 | 1.00 (1.00−1.00) |
Mean and range for success rate at each applied transducer force.
Results
| Force (g) | SWV (m/s) | Depth (cm) | Interquartile range (IQR) |
|---|---|---|---|
| 5 | 1.37 ± 0.29 (0.90−2.16) | 4.5 ± 1.2 (2.7−8.0) | 0.39 ± 0.18 (0.14−0.92) |
| 249 | 1.51 ± 0.31 (0.94−2.07) | 3.9 ± 1.2 (2.0−8.0) | 0.38 ± 0.17 (0.11−0.75) |
| 495 | 1.45 ± 0.33 (0.79−2.43) | 3.8 ± 1.1 (2.0−7.3) | 0.34 ± 0.16 (0.10−0.63) |
| 1000 | 1.24 ± 0.27(0.81−1.86) | 3.7 ± 1.0 (1.9−6.9) | 0.26 ± 0.17 (0.04−0.76) |
| 2015 | 1.19 ± 0.32 (0.81−2.21) | 3.5 ± 0.9 (2.1−6.2) | 0.29 ± 0.16 (0.06−0.66) |
| 2972 | 1.25 ± 0.31 (0.79−1.92) | 3.5 ± 09(1.8−6.1) | 0.27 ± 0.12 (0.06−0.52) |
Mean, SD and range for SWV, depth of ROI from skin, and IQR at each applied transducer force.