| Literature DB >> 28184129 |
G D Mitchell1, A A Bell1.
Abstract
The coulometric method presented here is a reliable method for the direct analysis of CO2/air cylinder gas mixtures. It is based on Faraday's laws of electrolysis and therefore no external standardization is required. A series of CO2/air cylinder gas mixtures ranging in concentration from 300 to 375 µmol/mol (ppm) were analyzed and the results compared to those results obtained by non-dispersive infrared (NDIR) analysis with traceability to gravimetric standards. The coulometric method is rapid, sensitive, precise, and with the proper experimental controls, will yield accurate results.Entities:
Keywords: N,N-dimethylformatide; carbon dioxide; coulometry; current efficiency; cylinder gases; monoethanolamine
Year: 1991 PMID: 28184129 PMCID: PMC4927231 DOI: 10.6028/jres.096.031
Source DB: PubMed Journal: J Res Natl Inst Stand Technol ISSN: 1044-677X
Comparison of CO2 coulometrica and NDIR data
| Cylinder number | Direct coloumetric concentration µmol/mol | Corrected | NDIR concentration | Ratio Coul/NDIR |
|---|---|---|---|---|
| 1 | 307.3 | 303.3 | 303.8 | 0.998 |
| 2 | 337.5 | 333.1 | 335.9 | 0.992 |
| 3 | 338.1 | 333.7 | 335.9 | 0.993 |
| 4* | 346.4 | 341.9 | 341.7 | 1.001 |
| 5 | 344.8 | 340.3 | 342.8 | 0.993 |
| 6 | 348.2 | 343.7 | 343.2 | 1.001 |
| 7 | 349.8 | 345.2 | 347.6 | 0.993 |
| 8 | 352.7 | 348.1 | 351.5 | 0.990 |
| 9 | 380.2 | 375.3 | 375.2 | 1.000 |
| 10 | 380.7 | 375.8 | 375.3 | 1.001 |
Uncertainty ± 4.2 µmol/mol.
Direct coulometric data corrected for current efficiency. .
Uncertainty ±0.4 µmol/mol.
Figure 1Chronological study.