G D Mitchell1, A A Bell1.
Abstract
Coulometry has been established as an important and reliable method for the determination of acidic compounds. The analytical method and simple apparatus described here arc applied to the precise and accurate determination of sulfur dioxide in nitrogen, specifically in compressed gas cylinders at nominal concentrations of 50 and 100 µmol/mol (ppm). This method is constant current coulometry where the magnitude of the current is set by the balance between the electrochemical generation of OH-, the flow of SO2, and the chemical reaction of the solution. The method is direct, rapid, and can be refined further to provide analysis at the nanomol/mol level.Entities:
Keywords: acid rain; analysis of gases; compressed gas; coulometry; sulfur dioxide
Year: 1991 PMID: 28184128 PMCID: PMC4927230 DOI: 10.6028/jres.096.030
Source DB: PubMed Journal: J Res Natl Inst Stand Technol ISSN: 1044-677X
Figure 1Coulometrie titration cell.
Figure 2Diagram for apparatus.
Current efficiency determinationa
| Permeating tube concentration (µmol/mol) | Coulometric concentration (µmol/mol) | Current efficiency (%) |
|---|---|---|
| 48.0 ± 0.6 | 48.6 ± 0.2 | 101.1 ± 0.6 |
.
All uncertainties are expressed at the 95% confidence level.
Coulometric titration results
| Cylinder identification | Coulometric | Coulometric | Certified | Ratio (corrected to certified) |
|---|---|---|---|---|
| A | 47.5 | 47.0 | 46.7 | 1.006 |
| B | 48.0 | 47.5 | 46.7 | 1.017 |
| C | 49.5 | 49.0 | 48.2 | 1.017 |
| D | 97.2 | 96.1 | 95.7 | 1.004 |
| E | 96.4 | 95.3 | 95.7 | 0.996 |
The uncertainty at the 95% confidence level is 6%.
The uncertainty at the 95% confidence level is 2%.
Corrected for current efficiency bias.