Results of a CIE Detector Response Intercomparison.

Douglas B Thomas1, Edward F Zalewski1.   

Abstract

A total of fifteen laboratories participated in the CIE detector response intercomparison which was designed to assess the level of agreement among participating laboratories in the absolute measurement (with respect to SI) of photodetector response in the visible spectral region. Most participants were either commercial laboratories or university laboratories with the National Institute of Standards and Technology (NIST) serving as the host laboratory. Each laboratory determined the absolute response of each of two silicon photodiode radiometers which were designed for the intercomparison by NIST. Approximately two-thirds of the laboratories reported response values which agreed with the NIST values to within ±1.0% at the two wavelengths of 488 and 633 nm.

Entities:  

Keywords:  laboratory evaluation; laser; optical radiation; photodiode; spectral response

Year:  1990        PMID: 28179789      PMCID: PMC4930017          DOI: 10.6028/jres.095.041

Source DB:  PubMed          Journal:  J Res Natl Inst Stand Technol        ISSN: 1044-677X


1. Introduction

This report provides the final results of a detector response intercomparison under the aegis of CIE Technical Committee TC 2-06 on Absolute Spectral Responsivity of Detectors. Members of the Technical Committee are listed in Appendix B of this report. The primary purpose of the intercomparison was to assess the level of agreement among participating laboratories in the absolute measurement (with respect to SI) of photodetector response (A/W) in the visible spectral region. The method chosen to accomplish this is to have these laboratories measure the absolute response of selected radiometersat two specific wavelengths near each end of the visible spectrum. The wavelengths selected are those of the helium-neon laser (632.8 nm) and the argon ion laser (488.0 nm). The intercomparison was implemented on the basis of the National Institute of Standards and Technology (NIST) serving as the host laboratory and providing (a) the radiometers to be used in the intercomparison, (b) instructions to participating laboratories in the use of the radiometers in absolute response measurements, and (c) data analysis and a report of the results of the intercomparison. The intercomparison was conducted in two stages: (1) intercomparison of U.S. laboratories and (2) intercomparison of laboratories outside of the United States. All participating laboratories except two are either commercial laboratories or university laboratories. The Electrotechnical Laboratory in Ibaraki, Japan and the Van Swinden Laboratory, The Netherlands, are national standards laboratories. The intercomparison was planned on the basis that NIST would measure the absolute response of all radiometers before shipment to the participating laboratories and then again after the radiometers were returned. The NIST absolute response value for each radiometer would be the average of the two NIST measurements.

2. The Radiometers

The radiometers used in this intercomparison were designed for ruggedness and ease of use and included commercially available silicon photodiodes. Each radiometer consists of a silicon photodiode and amplifier circuit mounted in a cylindrical aluminum housing and an external power supply. Since it was expected that the majority of the laboratories would make their measurements using lasers, the photodiodes were not protected by a window. However, to protect each photodiode during non-use, the diodes were maintained in sealed compartments. Two types of detectors were used: EG&G4 model UV-444B PN photodiodes and UDT model UV-100 inversion layer photodiodes. Radiometers PI-17, PI-19, PI-20, and PI-21 have the EG&G photodiodes while radiometers PI-25 through PI-32 have the UDT photodiodes. For the UDT photodiodes, a constant reverse bias voltage (4.5 V) was supplied by lithium batteries within each radiometer. The amplifier in each radiometer has gain settings from 104 to 109 V/A with accuracies of ±0.03% except for the 109 range where it is ±0.5%. Each participating U.S. laboratory received two radiometers: one with an EG&G photodiode and one with a UDT photodiode. The reason for requesting each laboratory to measure two radiometers is two-fold. (1) A second radiometer provides a backup for possible shipping damage and (2) there is a check on measurement repeatability. After the first stage of the intercomparison was completed involving U.S. laboratories, a decision was made to use only the radiometers with the UDT photodiodes for the intercomparison involving laboratories in other countries. This decision was made when it was determined that the EG&G photodiodes exhibited a small but significant response drift at 488 nm over long periods of time (see sec. 4).

3. The Intercomparison

A total of six U.S. laboratories and nine laboratories in other countries participated in the intercomparison. Each laboratory was asked to complete a questionnaire concerning detailed information about their detector response measurement system and to use standard data forms for reporting their results. Tables 1 and 2 list the U.S. laboratories and the information each submitted about their measurements at 488 and 633 nm. Tables 3 and 4 list the corresponding information for laboratories in other countries. The information submitted covered eight measurement parameters: (1) absolute base (absolute standard(s) used), (2) standard deviation of the measurements, (3) number of measurements per radiometer, (4) type of radiation source used, (5) beam diameter of the source, (6) radiant power level at the radiometer, (7) ambient temperature during measurements, and (8) estimated uncertainty (with respect to SI) of the absolute standards used. Some laboratories used a single silicon photodiode as an absolute (standard) base for their measurements. The absolute response of these photodiodes was determined using the self-calibration method [1,2]. Two laboratories made measurements only at 633 nm. Of the fifteen laboratories participating in the intercomparison, five laboratories used lasers as radiation sources at both wavelengths and five used a tungsten lamp/filter/monochromator system at both wavelengths. The remaining laboratories used various combinations of these sources. Radiant power levels ranged from 0.16 μW to 0.7 mW.
Table 1

Participating U.S. laboratories. Wavelength=488 nm

LLLNISTTEKXUDTUAZWEST
Absolute baseEGGQED2QED, UDTQED2QED2
S.D. of measurements0.24– 0.35%0.01%0.11%0.039– 1.10%0.003– 0.018%
No. of meas./Radiometer5503248
Radiation sourceTLFARLTLMBENTLF
Beam diameterOFA4 mm2 × 51 × 52 × 3
Power level3.7 × 10−8 W/cm20.5 mW0.6 µW2.0 µW4.0 µW
Ambient temp. (°C)2225–2623.021.021.0
Est. abs. uncertainty0.77%0.10%0.17%0.07%0.05%

LLL—Lawrence Livermore National Laboratory, Livermore, California.

NIST—National Institute of Standards and Technology, Gaithersburg, Maryland (Host Laboratory).

TEKX—Tektronix Corporation, Beaverton, Oregon.

UDT—United Detector Technology, Hawthorne, California.

UAZ—University of Arizona, Tucson, Arizona.

WEST—Westinghouse Electric Corporation, Baltimore, Maryland.

EGG—EG&G UV-444-BQ Photodiode.

QED—UDT QED-100 Radiometer.

QED2—UDT QED-200 Radiometer.

UDT—UDT UV-100L Photodiode.

TLF—Tungsten lamp/Filter.

ARL—Argon ion laser.

HENE—Helium-neon laser.

TLM—Tungsten lamp/Monochromator.

BEN—Bentham M300 Monochromator.

1 × 5—1 × 5 mm rectangle.

2 × 3—2.5 × 3.5 mm rectangle.

2 × 5—2.5 × 5.0 mm rectangle.

OFA—Overfill of aperture.

Table 2

Participating U.S. laboratories. Wavelength = 633 nm

LLLNISTTEKXUDTUAZWEST
Absolute baseEGGQED2QED, UDTQED2QED2UDT, QED2 EGG
S.D. of measurements0.28%0.012%0.015– 0.12%0.3– 0.4%0.018– 0.004%0.26– 0.14%
No. of meas./Radiometer55062643
Radiation sourceTLFHENETLM, HENEBEN, HENETLFHENE
Beam diameterOFA4 mm2 × 5, 2.5 mm1 × 5, 4 mm2 × 32 mm
Power level4.9 × 10−8 W/cm20.5 mW0.6 µW, 0.5 µW2.0 µ.W, 0.5 mW11 µW0.44 mW
Ambient temp. (°C)2225–2623.021.021.020.8
Est. abs. uncertainty0.77%0.10%0.17%0.07%0.05%0.05%

LLL—Lawrence Livermore National Laboratory, Livermore California.

NIST—National Institute of Standards and Technology, Gaithersburg, Maryland (Host Laboratory).

TEKX—Tektronix Corporation, Beaverton, Oregon.

UDT—United Detector Technology, Hawthorne, California.

UAZ—University of Arizona, Tucson, Arizona.

WEST—Westinghouse Electric Corporation, Baltimore, Maryland.

EGG—EG&G UV-444-BQ Photodiode.

QED—UDT QED-100 Radiometer.

QED2—UDT QED-200 Radiometer.

UDT—UDT UV-100L Photodiode.

TLF—Tungsten lamp/Filter.

ARL—Argon ion laser.

TLM—Tungsten lamp/Monochromator.

HENE—Helium-neon laser.

BEN—Bentham M300 Monochromator.

1 × 5—1 × 5 mm rectangle.

2 × 3—2.5 × 3.5 mm rectangle.

2 × 5—2.5 × 5.0 mm rectangle.

OFA—Overfill of aperture.

Table 3

Participating laboratories in other countries. Wavelength=488 nm

CIPETLHAMLCIELNEMATKROCUDIVSL
Absolute basePSPHAM2ASPTSPHAM2HAM3EGGQED2
S.D. of measurements0.17– 0.21%0.04%0.6%0.11– 0.18%0.02%0.52%0.12– 0.04%0.007– 0.011%
No. of meas./Radiometer104627,331058–1275
Radiation sourceARLTLMTLMTLMARLTLMARLARL
Beam diameter4 mm2 × 35 mm6 mm3 mm7 mm0.6 mm4 mm
Power level0.02 mW0.16 µW20 µW1.5 µW0.19 mW2.4 × 10−3 W/m20.3 mW0.7 mW
Ambient temp. (°C)1823232325232123
Est. abs. uncertainty0.17%0.07%0.20%0.11 0.22%0.07%0.17%0.50%0.20%

CIP—Central Institute of Physics, Magurele-Bucharest, Romania.

ETL—Electrotechnical Laboratory, Ibaraki, Japan.

HAM—Hamamatsu Photonics K.K., Hamamatsu City, Japan.

LCIE—L.C.I.E., Fontenay-aux-Roses, France.

LNE—Laboratoire National D’Essais, Paris, France.

MAT—Matsushita Electric Industrial Co. Ltd., Moriguchi Osaka, Japan.

KROC—PR C Krochmann GMBH, Berlin, West Germany.

UDI—University College, Dublin, Ireland.

VSL—Van Swinden Laboratory, Delft, The Netherlands.

PSP—pn Silicon photodiode (Romanian).

HAM2—Hamamatsu S 1337 Photodiode.

HAM3—Hamamatsu S 1227 Photodiode.

ASP—Silicon photodiode.

TSP—Three silicon photodiodes.

EGG—EG&G UV-444B Photodiode.

QED2—UDT QED-200 Radiometer.

ARL—Argon ion laser.

TLM—Tungsten lamp/Monochromator.

2 × 3—2 × 3 mm rectangle.

Table 4

Participating laboratories in other countries. Wavelength = 63 3 nm

CIPETLHAMLCIELNEMATKROCUDIVSL
Absolute basePSPHAM1HAM2ASPTSPHAM2HAM3EGGQED2
S.D. of measurements0.17–0.14%0.02%0.07–0.04%0.6%0.10–0.08%0.02%0.3–0.15%0.07–0.08%0.02–
No. of meas./Radiometer10310621,28105675
Radiation sourceHENEHENEHENETLMTLMHENETLMHENEHENE
Beam diameter4 mm1 mm1.5 mm5 mm6 mm3 mm7 mm0.6 mm4 mm
Power level0.1 mW40 µW25 µW30µW2 µW0.3 mW1.1 × 10−2 W/m20.6 mW0.7 mW
Ambient temp. (°C)182325232325231823
Est. abs. uncertainty0.17%0.07%0.17%0.20%0.12 0.09%0.07%0.17%0.08%0.20%

CIP—Central Institute of Physics, Magurele-Bucharest, Romania.

ETL—Electrotechnical Laboratory, Ibaraki, Japan.

HAM—Hamamatsu Photonics K.K., Hamamatsu City, Japan.

LCIE—L.C.I.E., Fontenay-aux-Roses, France.

LNE—Laboratoire National D’Essais, Paris, France.

MAT—Matsushita Electric Industrial Co. Ltd., Moriguchi Osaka, Japan.

KROC—PRC Krochmann GMBH, Berlin, West Germany.

UDI—University College, Dublin, Ireland.

VSL—Van Swinden Laboratory, Delft, The Netherlands.

PSP—pn Silicon photodiode (Romanian).

HAM1—Hamamatsu S 1723 Photodiode.

HAM2—Hamamatsu S 1337 Photodiode.

ASP—Silicon photodiode.

TSP—Three silicon photodiodes.

HAM3—Hamamatsu S 1227 Photodiode.

EGG—EG&G UV-444B Photodiode.

QED2—UDT QED-200 Radiometer.

HENE—Helium-neon laser.

TLM—Tungsten lamp/Monochromator.

The measurement system used at NIST for this intercomparison consists of He-Ne and Argon ion lasers, laser stabilizer, spatial filter, beam splitter, and a silicon photodiode monitor detector. Three UDT QED-200 absolute radiometers [3] were used as base standards. Figure 1 is an illustration of the system components. The NIST procedures for determining the absolute response of the intercomparison radiometers consisted basically of two steps: (1) measuring the ratio of the photocurrent of each UDT QED-200 radiometer to the photocurrent of the monitor detector at a particular laser power setting and (2) measuring the ratio of the photocurrent of the intercomparison radiometers to the monitor detector at the same power level in (1). Since the UDT QED-200 radiometers are 100% quantum efficient (with voltage bias) at the wavelengths and power levels stated, the power (watts) can be accurately measured and the absolute response (amperes/watt) of each of the intercomparison radiometers can be determined. Details concerning the system and the measurement procedure are further described in [4]. The NIST absolute base was compared to other international standards laboratories in a recent detector response intercomparison sponsored by the Consultative Committee on Photometry and Radiometry (CCPR) [5]. In the CCPR intercomparison, the absolute response of a select group of silicon photodiode radiometers were measured by 10 international standards laboratories and also by NIST which served as the host laboratory. The ratios of the NIST response values to the mean of the response values of the other participating laboratories were 1.0011±0.0035 and 1.0014±0.0037 at the two wavelengths of 488 and 633 nm, respectively.
Figure 1

NIST laser based detector calibration facility.

Since the absolute response values reported by each of the participating laboratories were compared to the response values determined at NIST, it was essential for NIST to measure the response of each set of radiometers before it was shipped to the participating laboratory and then measured again after the radiometers were returned. The before and after measurements by NIST were made to determine if any significant changes occurred in the radiometers during shipment.

4. Data Analysis

Tables 5 and 6 list the laboratory designations, date of measurement, radiometer descriptions, and absolute responsivities reported by the U.S. laboratories and laboratories in other countries, respectively. Each set of response values for a participating laboratory includes the corresponding before and after values determined by NIST. The NIST value for each radiometer was taken as the average of the before and after respective values. The before and after NIST values indicate that some of the radiometers had undergone a small but significant change in response between shipments to and from the laboratories. For example, at 488 nm, the response value for radiometer PI-20 (laboratory C, table 5) decreased from 0.2814 to 0.2787 over the period 7/87 to 2/88 as measured by NIST. This is a decrease of 0.96%. All ratios reported represent an average of the before and after values.
Table 5

U.S. laboratories

LaboratoryDateResponsivity (488 nm)Responsivity (633 nm)
PI-20PI-25
NIST  7/87  0.4551  0.4152
LAB A  7/87  0.4555  0.4154
NIST  7/87  0.4548  0.4153
PI-21PI-27PI-21PI-27
NIST  7/87  0.2830  0.2994  0.4550  0.4140
LAB B  9/87  0.2857  0.2990  0.4608  0.4208
NIST12/87  0.2824  0.2986  0.4546  0.4141
PI-20PI-25PI-20PI-25
NIST  7/87  0.2814  0.2982  0.4548  0.4153
LABC  9/87  0.2802  0.2987  0.4547  0.4151
NIST  2/88  0.2787  0.2984  0.4545  0.4153
PI-19PI-28PI-19PI-28
NIST  7/87  0.2596  0.2965  0.4472  0.4165
LAB D11/87  0.2570  0.3006  0.4468  0.4168
NIST  2/88  0.2551  0.2969  0.4467  0.4169
PI-17PI-30PI-17PI-30
NIST  7/87  0.2849  0.3021  0.4580  0.4108
LAB E12/87  0.2799  0.3003  0.4521  0.4027
NIST  2/88  0.2830  0.3025  0.4576  0.4105
Table 6

Laboratories in other countries

LaboratoryDateResponsivity (488 nm)Responsivity (633 nm)
PI-28PI-31PI-28PI-31
NIST  2/88  0.2969  0.3014
NIST  8/88  0.4169  0.4125
LAB F10/88  0.2964  0.3008  0.4166  0.4124
NIST12/88  0.4169  0.4129
NIST  1/89  0.2967  0.3009
PI-28PI-31PI-28PI-31
NIST  2/88  0.2969  0.3014
NIST  8/88  0.4169  0.4125
LAB G11/88  0.2973  0.3019  0.4170  0.4129
NIST12/88  0.4169  0.4129
NIST  1/89  0.2967  0.3009
PI-28PI-31
NIST  8/88  0.4169  0.4125
LAB H12/88  0.4178  0.4138
NIST12/88  0.4169  0.4129
PI-25PI-32PI-25PI-32
NIST  2/88  0.2983  0.2599
NIST  8/88  0.4153  0.4515
LAB I11/88  0.2999  0.2622  0.4171  0.4538
NIST12/88  0.4148  0.4514
NIST  1/89  0.2987  0.2605
PI-28PI-31PI-28PI-31
NIST12/88  0.4169  0.4129
NIST  1/89  0.2967  0.3009
LAB J  5/89  0.2972  0.3020  0.4172  0.4130
NIST  8/89  0.4168  0.4127
NIST  9/89  0.2966  0.3009
PI-28PI-31PI-28PI-31
NIST12/88  0.4169  0.4129
NIST  1/89  0.2967  0.3009
LAB K  7/89  0.2980  0.3030  0.4160  0.4110
NIST  8/89  0.4168  0.4127
NIST  9/89  0.2966  0.3009
PI-25PI-32PI-25PI-32
NIST12/88  0.4148  0.4514
NIST  1/89  0.2987  0.2605
LAB L  6/89  0.2995  0.2617  0.4153  0.4511
NIST  8/89  0.4148  0.4512
NIST  9/89  0.2982  0.2607
PI-27PI-29PI-27PI-29
NIST  8/88  0.4145  0.4149
NIST  2/89  0.2989  0.2991
LAB M  4/89  0.3043  0.3049  0.4205  0.4216
NIST  8/89  0.4144  0.4147
NIST  9/89  0.2992  0.2988
PI-26PI-30PI-26PI-30
NIST  7/88  0.2667  0.3020
NIST  8/88  0.4454  0.4108
LAB N12/88  0.2805  0.3144  0.4583  0.4240
NIST  8/89  0.4446  0.4107
NIST  9/89  0.2672  0.3023
Since three of the four radiometers with the EG&G type photodiode showed small but significant decreases in response at 488 nm over a 7-month period, it was decided to use only the radiometers with the UV-100 type photodiodes for the second phase of the intercomparison (foreign laboratories). Table 7 is a listing of the participating laboratories by code letter, the absolute response values reported by each laboratory, the absolute response values as determined by NIST, and the ratios of the response values.
Table 7

Response ratios

Laboratory(RAD. #)488 nm
633 nm
Resp. Lab.Resp. NISTLab./NISTResp. Lab.Resp. NISTLab./NIST
A (PI-20)0.45550.45501.0011
A (PI-25)0.41540.41521.0005
B (PI-21)0.28570.28271.01060.46080.45481.0132
B (PI-27)0.29900.29901.00000.42080.41411.0162
C (PI-20)0.28020.28060.99860.45470.45471.0000
C (PI-25)0.29870.29831.00130.41510.41530.9995
D (PI-19)0.25700.25701.00000.44680.44690.9998
D (PI-28)0.30060.29671.01310.41680.41671.0002
E (PI-17)0.27990.28360.98700.45210.45780.9875
E (PI-30)0.30030.30230.99340.40270.41060.9808
F (PI-28)0.29640.29680.99870.41660.41690.9993
F (PI-31)0.30080.30110.99900.41240.41270.9993
G (PI-28)0.29730.29681.00170.41700.41691.0002
G(PI-31)0.30190.30111.00270.41290.41271.0005
H (PI-28)0.41780.41691.0022
H (PI-31)0.41380.41271.0027
I (PI-25)0.29990.29851.00470.41710.41511.0048
I (PI-32)0.26220.26021.00770.45380.45151.0051
J (PI-28)0.29720.29661.00200.41720.41681.0010
J (PI-31)0.30200.30091.00370.41300.41281.0005
K (PI-28)0.29800.29661.00470.41600.41680.9981
K (PI-31)0.30300.30091.00700.41100.41280.9956
L (PI-25)0.29950.29841.00370.41530.41481.0012
L (PI-32)0.26170.26061.00420.45110.45130.9996
M (PI-27)0.30430.29901.01770.42050.41441.0147
M (PI-29)0.30490.29901.01970.42160.41481.0164
N (PI-26)0.28050.26701.05060.45830.44501.0299
N (PI-30)0.31440.30221.04040.42400.41081.0321
Figures 2 and 3 are plots of the ratios of the response values (A/W) determined by each of the participating laboratories to the respective response values (A/W) determined by NIST at 488 and 633 nm. The solid line on each plot is the mean of all the ratios at the respective wavelength and the dashed lines are the standard deviation of the mean. Table 8 is a summary of the standard deviations of the measurements and the estimated uncertainty (with respect to SI) of the absolute standards used by each of the participating laboratories. Also listed are the before/after change in absolute response for each detector as measured by NIST and the absolute response ratio uncertainty. The absolute response ratio uncertainty is the quadrature summation of the measurement and absolute uncertainties of each participant laboratory, the before/after response change for each radiometer, and the NIST measurement and absolute uncertainties. The error bars in figures 2 and 3 indicate the absolute response ratio uncertainty for each laboratory.
Figure 2

Ratio of the participant laboratory spectral response to that determined by NIST at 488 nm. The error bars indicate the quadrature summation of the measurement and absolute uncertainties of each participant laboratory, the before/after response change for each radiometer, and the NIST measurement and absolute uncertainties. The dashed lines indicate the standard deviation of the ratio values.

Figure 3

Ratio of the participant laboratory spectral response to that determined by NIST at 633 nm. The error bars indicate the quadrature summation of the measurement and absolute uncertainties of each participant laboratory, the before/after response change for each radiometer, and the NIST measurement and absolute uncertainties. The dashed lines indicate the standard deviation of the ratio values.

Table 8

Summary of uncertainties

LaboratoryMeasurement standard deviation (1 sigma)Absolute uncertainty (1 sigma)Before/After response change (%/100)Ratio uncertainty (1 sigma)
488 nm

LLL  0.00240.00770.00210.0084
  0.00350.00770.00270.0089
N1ST  0.00010.0010
  0.00010.0010
TEKX  0.00110.00170.00960.0099
  0.00110.00170.00070.0024
UDT    0.000390.00070.01750.0175
  0.01100.00070.00130.0111
UAZ    0.000030.00050.00670.0068
    0.000180.00050.00130.0017
CIP  0.00170.00170.00190.0032
  0.00210.00170.00100.0031
ETL  0.00040.00070.00070.0015
  0.00040.00070.00170.0021
LCIE  0.00600.00200.00030.0064
  0.00600.00200.00000.0064
LNE  0.00110.00110.00030.0019
  0.00180.00220.00000.0030
MAT  0.00020.00070.00070.0014
  0.00020.00070.00170.0021
KROC  0.00520.00170.00100.0057
  0.00520.00170.00100.0057
UDI  0.00120.00500.00130.0054
  0.00040.00500.00230.0056
VSL    0.000070.00200.00170.0028
    0.000110.00200.00080.0024

633 nm

LLL  0.00280.00770.00090.0083
  0.00280.00770.00020.0083
NIST    0.000120.0010
    0.000120.0010
TEKX    0.000150.00170.00070.0021
  0.00120.00170.00000.0023
UDT  0.00400.00070.00110.0043
  0.00300.00070.00100.0034
UAZ    0.000180.00050.00090.0014
    0.000040.00050.00070.0013
WEST  0.00260.00050.00070.0029
  0.00140.00050.00020.0018
CIP  0.00170.00170.00180.0032
  0.00140.00170.00020.0024
ETL  0.00020.00070.00000.0012
  0.00020.00070.00100.0016
HAM  0.00070.00170.00000.0021
  0.00040.00170.00100.0022
LCIE  0.00600.00200.00020.0064
  0.00600.00200.00050.0064
LNE  0.00100.00120.00020.0019
  0.00080.00090.00050.0016
MAT  0.00020.00070.00000.0012
  0.00020.00070.00100.0016
KROC  0.00300.00170.00020.0036
  0.00150.00170.00050.0025
UDI  0.00070.00080.00120.0019
  0.00080.00080.00020.0015
VSL  0.00020.00200.00000.0022
  0.00020.00200.00050.0023

5. Conclusion

In general, it can be concluded that most of the response values reported by the laboratories were in good agreement with NIST. At 488 nm, the mean of all participating laboratories was 0.71% higher than the corresponding NIST values with a standard deviation of 1.39%. Similarly, at 633 nm, the mean of all laboratory values was higher than the NIST values by 0.36% with a standard deviation of 1.07%. All laboratories participating in this intercomparison (except laboratory N) reported values at both wavelengths within ±2.0% of the NIST values and nine of the 14 laboratories reported values at both wavelengths within ±1.0% of the NIST values. This can be considered good agreement among the laboratories when one considers the variety of sources, procedures, and testing environments involved in this intercomparison.
  3 in total

1.  Spectral response self-calibration and interpolation of silicon photodiodes.

Authors:  J Geist; E F Zalewski; A R Schaefer
Journal:  Appl Opt       Date:  1980-11-15       Impact factor: 1.980

2.  Silicon photodiode device with 100% external quantum efficiency.

Authors:  E F Zalewski; C R Duda
Journal:  Appl Opt       Date:  1983-09-15       Impact factor: 1.980

3.  Silicon photodiode absolute spectral response self-calibration.

Authors:  E F Zalewski; J Geist
Journal:  Appl Opt       Date:  1980-04-15       Impact factor: 1.980

  3 in total