| Literature DB >> 28179765 |
Abstract
A procedure for the calibration of instruments for the detection of 222Rn in air is described. The method is based on the alpha-spectrometric determination of the concentration in air of 218Po in the calibration chamber. The calibration chamber is described, together with the method of maintaining a high aerosol concentration. The 218Po concentration at steady state in the chamber is found to be 98% of the 222Rn concentration typically. An assessment of the sources of uncertainty in the method presented indicate that the 222Rn concentration in the chamber can be determined with an overall uncertainty of about 7% at the 95% confidence level.Entities:
Keywords: aerosol generation; alphaspectrometry; calibration chamber; radon daughter measurement; radon gas measurement; statistical uncertainty
Year: 1990 PMID: 28179765 PMCID: PMC4930038 DOI: 10.6028/jres.095.014
Source DB: PubMed Journal: J Res Natl Inst Stand Technol ISSN: 1044-677X
Decay schemes for 222Rn (radon) and 220Rn (thoron), short-lived daughters only
| Nuclide | Half-life | Mode of decay | MeV | |
|---|---|---|---|---|
| 222Rn | 3.823 d | 5.48 | ||
| 218Po | 3.11 min | 6.00 | ||
| 214Pb | 26.8 min | |||
| 214Bi | 19.7 min | |||
| 214Po | 164 µs | 7.69 | ||
|
| ||||
| 220Rn | 55 s | 6.28 | ||
| 216Po | 0.16 s | 6.78 | ||
| 212Pb | 10.6 h | |||
| 212Bi | 60.5 min | 6.94 | 25% | |
| 6.08 | 10% | |||
| other | 1% | |||
| 212Po | 0.3 µs | 8.78 | 64% | |
214Po is always in equilibrium with 214Bi.
Range of parameter values that account for the radon daughter conditions in the exposure chamber
| Detachment probability for 214Pb on formation | 0.83 |
| Ratio of deposition rate constants for attached and unattached radon daughters | 0.01 |
| Ventilation rate | 0.006–0.008 h−1 |
| Rate constant for attachment of radon daughters to aerosols | 80–190 h−1 |
| Deposition rate constant for unattached radon daughters | 3–5 h−1 |
| Ratio of activity concentrations of 218Po to radon | 0.967–0.990 |
| 0.98 | |
| 0.96–1.00 |
Assessed uncertainties in the measurement of radon activity concentration
| Systematic uncertainties | Per cent | |
|---|---|---|
| Flow rate: | Semi-range | 2 |
| Reference 241Am source: | Semi-range | 2 |
| Modelling: | Semi-range | 2 |
| Overall systematic, 95% confidence interval: | ±3.9% | |