Literature DB >> 28164516

HIL Interferences on Three Hemostasis Analyzers and Contribution of a Preanalytical Module for Routine Coagulation Assays.

Carole Nagant, Laurence Rozen, Anne Demulder.   

Abstract

BACKGROUND: Preanalytical issues are a major part of routine coagulation laboratory errors. Automation in detection of preanalytical problems, including hemolysis, icterus and lipemia (HIL), improper tube fill volume, and undue clotting, has recently been implemented on specific hemostasis instruments. The aim of this study was to assess the added value of a new preanalytical module integrated into hemostasis analyzers compared to visual inspection of samples.
METHODS: The detection of preanalytical issues was performed by visual or manual inspection of the samples and by the new preanalytical module integrated on the ACL TOP 550. Additional tests were done to evaluate the interference of hemoglobin (Hb), bilirubin or triglycerides (TG). Plasma pools containing the interference substance were tested for routine coagulation assays on the STA-R Evolution, CS-5100, and ACL TOP 550.
RESULTS: Visual or manual inspection detected statistically less samples with preanalytical issues than the new preanalytical module integrated on the ACL TOP 550 (3.5% vs. 6.6%, p < 0.001). The majority of the samples were rejected for poor filling. HIL interferences appeared on assays when the concentration of Hb, bilirubin or TG exceeded a certain threshold that was analyzer and reagent dependent.
CONCLUSIONS: Automatic and standardized check of routine coagulation samples by ACL TOP 550 increased the accuracy and consistency in detection of preanalytical issues as compared to visual inspection only. The main advantages were the detection of insufficient filled tubes and icteric samples that are not detected visually.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 28164516     DOI: 10.7754/Clin.Lab.2016.160313

Source DB:  PubMed          Journal:  Clin Lab        ISSN: 1433-6510            Impact factor:   1.138


  2 in total

1.  A collaborative study by the Working Group on Hemostasis and Thrombosis of the Italian Society of Clinical Biochemistry and Clinical Molecular Biology (SIBioC) on the interference of haemolysis on five routine blood coagulation tests by evaluation of 269 paired haemolysed/non-haemolysed samples.

Authors:  Chiara Novelli; Matteo Vidali; Bruno Brando; Benedetto Morelli; Giovanna Andreani; Marina Arini; Paola Calzoni; Roberta Giacomello; Barbara Montaruli; Emanuela Muccini; Angela Papa; Paola Pradella; Lucia Ruocco; Fosca Siviero; Filomena Gemma Viola; Mario Zanchetta; Lorena Zardo; Giuseppe Lippi
Journal:  Biochem Med (Zagreb)       Date:  2018-10-15       Impact factor: 2.313

2.  Clinical Evaluation of the Pre-Analytical Capabilities of Hemostasis Instrument.

Authors:  Mingyu Yang; Xiaoning Gui; Run Wang; Shiju Jiang; Jing Zhou; Jian Chen; Meiling Wang; Jiwei Ning; Linzi Miao; Hongwei Liu; Xiaomei Tang; Chenxue Qu
Journal:  Clin Appl Thromb Hemost       Date:  2022 Jan-Dec       Impact factor: 3.512