Poornima Ramanan1, M J Espy1, Reeti Khare2, M J Binnicker3.
Abstract
A real-time RT-PCR assay was designed to detect and differentiate norovirus genogroups I (GI) and II (GII), with primers and probes targeting the nonstructural polyprotein gene. Stool samples (n = 100) submitted for routine testing by the BioFire FilmArray® GI panel were also tested by the norovirus GI/GII real-time PCR assays. When compared to the FilmArray GI panel, the norovirus real-time PCR assay demonstrated a sensitivity of 77.5% (62/80) and specificity of 95% (19/20). Specimens yielding discordant results (n = 19) were tested at two outside laboratories for adjudication. Following discordant resolution, the adjusted sensitivity and specificity of the norovirus real-time PCR assays were 96.9% (63/65) and 100% (35/35), respectively. These results suggest that the real-time PCR assays are able to accurately detect and differentiate norovirus GI/GII from clinical stool specimens. Furthermore, our report highlights a potential issue with the specificity of the BioFire FilmArray® norovirus assay, which warrants additional investigation.Entities:
Keywords: Norovirus; PCR
Mesh:
Substances:
Year: 2017 PMID: 28159444 DOI: 10.1016/j.diagmicrobio.2017.01.009
Source DB: PubMed Journal: Diagn Microbiol Infect Dis ISSN: 0732-8893 Impact factor: 2.803