Procedure for the Selection and Validation of a Calibration Model II-Theoretical Basis.

Brigitte Desharnais1,2, Félix Camirand-Lemyre3, Pascal Mireault4, Cameron D Skinner5.   

Abstract

In the first part of this paper (I-Description and application), an automated, stepwise and analyst-independent process for the selection and validation of calibration models was put forward and applied to two model analytes. This second part presents the mathematical reasoning and experimental work underlying the selection of the different components of this procedure. Different replicate analysis designs (intra/inter-day and intra/inter-extraction) were tested and their impact on test results was evaluated. For most methods, the use of intra-day/intra-extraction measurement replicates is recommended due to its decreased variability. This process should be repeated three times during the validation process in order to assess the time stability of the underlying model. Strategies for identification of heteroscedasticity and their potential weaknesses were examined and a unilateral F-test using the lower limit of quantification and upper limit of quantification replicates was chosen. Three different options for model selection were examined and tested: ANOVA lack-of-fit (LOF), partial F-test and significance of the second-order term. Examination of mathematical assumptions for each test and LC-MS-MS experimental results lead to selection of the partial F-test as being the most suitable. The advantages and drawbacks of ANOVA-LOF, examination of the standardized residuals graph and residuals normality testing (Kolmogorov-Smirnov or Cramer-Von Mises) for validation of the calibration model were examined with the last option proving the best in light of its robustness and accuracy. Choosing the correct calibration model improves QC accuracy, and simulations have shown that this automated scheme has a much better performance than a more traditional method of fitting with increasingly complex models until QC accuracies pass below a threshold.
© The Author 2017. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

Mesh:

Year:  2017        PMID: 28158619     DOI: 10.1093/jat/bkx002

Source DB:  PubMed          Journal:  J Anal Toxicol        ISSN: 0146-4760            Impact factor:   3.367


  3 in total

1.  Experimental and statistical protocol for the effective validation of chromatographic analytical methods.

Authors:  Eugenio Alladio; Eleonora Amante; Cristina Bozzolino; Fabrizio Seganti; Alberto Salomone; Marco Vincenti; Brigitte Desharnais
Journal:  MethodsX       Date:  2020-05-16

2.  Targeted and untargeted detection of fentanyl analogues and their metabolites in hair by means of UHPLC-QTOF-HRMS.

Authors:  Alberto Salomone; Daniele Di Corcia; Pierre Negri; Maria Kolia; Eleonora Amante; Enrico Gerace; Marco Vincenti
Journal:  Anal Bioanal Chem       Date:  2020-10-15       Impact factor: 4.142

3.  Targeted and untargeted quantification of quorum sensing signalling molecules in bacterial cultures and biological samples via HPLC-TQ MS techniques.

Authors:  Federica Dal Bello; Michael Zorzi; Riccardo Aigotti; Davide Medica; Vito Fanelli; Vincenzo Cantaluppi; Eleonora Amante; Viviana Teresa Orlandi; Claudio Medana
Journal:  Anal Bioanal Chem       Date:  2020-11-18       Impact factor: 4.142

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.