Reduce beam hardening artifacts of polychromatic X-ray computed tomography by an iterative approximation approach.

Hongli Shi, Zhi Yang, Shuqian Luo.   

Abstract

BACKGROUND: The beam hardening artifact is one of most important modalities of metal artifact for polychromatic X-ray computed tomography (CT), which can impair the image quality seriously.
OBJECTIVE: An iterative approach is proposed to reduce beam hardening artifact caused by metallic components in polychromatic X-ray CT.
METHODS: According to Lambert-Beer law, the (detected) projections can be expressed as monotonic nonlinear functions of element geometry projections, which are the theoretical projections produced only by the pixel intensities (image grayscale) of certain element (component). With help of a prior knowledge on spectrum distribution of X-ray beam source and energy-dependent attenuation coefficients, the functions have explicit expressions. Newton-Raphson algorithm is employed to solve the functions. The solutions are named as the synthetical geometry projections, which are the nearly linear weighted sum of element geometry projections with respect to mean of each attenuation coefficient. In this process, the attenuation coefficients are modified to make Newton-Raphson iterative functions satisfy the convergence conditions of fixed pointed iteration(FPI) so that the solutions will approach the true synthetical geometry projections stably. The underlying images are obtained using the projections by general reconstruction algorithms such as the filtered back projection (FBP). The image gray values are adjusted according to the attenuation coefficient means to obtain proper CT numbers.
RESULTS: Several examples demonstrate the proposed approach is efficient in reducing beam hardening artifacts and has satisfactory performance in the term of some general criteria. In a simulation example, the normalized root mean square difference (NRMSD) can be reduced 17.52% compared to a newest algorithm.
CONCLUSIONS: Since the element geometry projections are free from the effect of beam hardening, the nearly linear weighted sum of them, the synthetical geometry projections, are almost free from the effect of beam hardening. By working out the synthetical geometry projections, the proposed approach becomes quite efficient in reducing beam hardening artifacts.

Keywords:  Newton-Raphson algorithm; X-ray CT; beam hardening; fixed pointed iteration; metal artifact

Mesh:

Substances:

Year:  2017        PMID: 28157119     DOI: 10.3233/XST-16187

Source DB:  PubMed          Journal:  J Xray Sci Technol        ISSN: 0895-3996            Impact factor:   1.535


  2 in total

1.  Chest Computed Tomography Images in Neonatal Bronchial Pneumonia under the Adaptive Statistical Iterative Reconstruction Algorithm.

Authors:  Ying Sun; Liao Wu; Zhaofang Tian; Tianping Bao
Journal:  J Healthc Eng       Date:  2021-10-27       Impact factor: 2.682

2.  Analytical expressions for the reconstructed image of a homogeneous cylindrical sample exhibiting a beam hardening artifact in X-ray computed tomography.

Authors:  Tsukasa Nakano; Yoshito Nakashima
Journal:  J Xray Sci Technol       Date:  2018       Impact factor: 1.535

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.