Literature DB >> 28156008

A Device Histogram-Based Simple Predictor of Mortality Risk in ICD and CRT-D Patients: The Heart Rate Score.

Bruce L Wilkoff1, Mark Richards2, Arjun Sharma3, Nicholas Wold3, Paul Jones3, David Perschbacher3, Brian Olshansky4.   

Abstract

BACKGROUND: We hypothesized that survival in implantable cardioverter defibrillator (ICD) and cardiac resynchronization therapy defibrillator (CRT-D) patients is predicted by baseline Heart Rate Score.
METHODS: Heart Rate Score is determined from the atrial paced and sensed histogram of a DDD ICD or CRT-D, and defined as percent of beats in the histogram in the tallest 10 beats/min range bin. It was calculated at initial remote monitoring for patients enrolled in LATITUDE® without persistent atrial fibrillation, and with pulse generators implanted in 2006-2011. Univariate, multivariate, and Kaplan-Meier analyses determined the impact of Heart Rate Score on survival.
RESULTS: Of 57,893 ICDs and 67,929 CRT-Ds followed for 2.4 ± 1.5 years, each 10% increase in Heart Rate Score was associated with decreased survival (CRT-D hazard ratio [HR] 1.07 95%, confidence interval 1.06-1.07, P < 0.0001; ICD HR 1.05, 95% confidence interval 1.04-1.06, P < 0.0001). Multivariate analysis showed survival decreased with increasing age, atrial fibrillation, presence of a shock in first-year follow-up, and increasing programmed lower pacing rate in ICD and CRT-D patients. Increased percent right ventricular pacing predicted mortality in ICD patients, while male gender and lower percent left ventricular pacing predicted mortality in CRT patients. Heart Rate Score predicted survival independent of those variables. Heart Rate Score correlates with heart rate variability (standard deviation of average R-R intervals [SDANN]) when both are obtainable, but SDANN was only present in 6% of patients with Heart Rate Score >70%.
CONCLUSION: A simple device histogram measure, Heart Rate Score, predicts survival in ICD and CRT-D patients independent of the available variables, and even when SDANN is unavailable.
© 2017 Wiley Periodicals, Inc.

Entities:  

Keywords:  Heart Rate Histograms; cardiac resynchronization; chronotropic incompetence; defibrillators; noninvasive assessment of risk

Mesh:

Year:  2017        PMID: 28156008     DOI: 10.1111/pace.13036

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  3 in total

1.  Survival after left ventricular assist device implantation correlates with a novel device-based measure of heart rate variability: the heart rate score.

Authors:  Thomas Czermak; Valentina Seitelberger; Christian Hagl; Patrick-Nicolas Samson-Himmelstjerna; Sven Groß; Sebastian Sadoni; Oliver Heyn; Antonia Kellnar; Bonnie Hartrampf; Eric Lemmermöhle; Ina Klier; Raphael Rehms; Sabine Hoffmann; Heidi L Estner; Stephanie Fichtner; Korbinian Lackermair
Journal:  Interact Cardiovasc Thorac Surg       Date:  2021-07-26

2.  Impact of a poor functional capacity on the clinical outcomes in patients with a pacemaker implantation -Results from the Japanese Heart Rhythm Society Registry.

Authors:  Takanori Arimoto; Eiichi Watanabe; Ritsuko Kohno; Kenji Shimeno; Kan Kikuchi; Atsushi Doi; Kanki Inoue; Takashi Nitta; Akihiko Nogami; Haruhiko Abe; Ken Okumura
Journal:  J Arrhythm       Date:  2020-12-09

3.  Heart rate score, a measure related to chronotropic incompetence in pacemaker patients.

Authors:  Arjun D Sharma; Mark Richards; Brian Olshansky; Nicholas Wold; Paul Jones; David Perschbacher; Bruce L Wilkoff
Journal:  Heart Rhythm O2       Date:  2021-02-20
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.