Fast label-free detection of C-reactive protein using broad-band Mach-Zehnder interferometers integrated on silicon chips.

Aimilia Psarouli1, Athanasios Botsialas2, Alexandros Salapatas2, Gerasimos Stefanitsis3, Dimitra Nikita3, Gerhard Jobst4, Nikolaos Chaniotakis5, Dimitrios Goustouridis6, Eleni Makarona2, Panagiota S Petrou7, Ioannis Raptis2, Konstantinos Misiakos2, Sotirios E Kakabakos8.   

Abstract

An immunosensor for fast and accurate determination of C-reactive protein (CRP) in human serum samples based on an array of all-silicon broad-band Mach-Zehnder interferometers (BB-MZIs) is demonstrated. The detection was based on monitoring the spectral shifts during the binding of CRP on the antibody molecules that have been immobilized on the sensing arms of the BB-MZIs. By employing the reaction rate as the analytical signal the assay time was compressed to few minutes. The detection limit was 2.1ng/mL, the quantification limit was 4.2ng/mL and the linear dynamic range extended up to 100ng/mL. The measurements performed in human serum samples with the developed immunosensor were characterized by high repeatability and accuracy as it was demonstrated by dilution linearity and recovery experiments. In addition, the concentration values determined were in excellent agreement with those determined for the same samples by a standard clinical laboratory method. The compact size of the chip makes the proposed immunosensor attractive for incorporation into miniaturized devices for the determination of clinical analytes at the point-of-need.
Copyright © 2017 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Broad-band Mach-Zehnder interferometer; C-reactive protein; Label-free immunosensor; Monolithic integration

Mesh:

Substances:

Year:  2017        PMID: 28153283     DOI: 10.1016/j.talanta.2017.01.001

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057


  4 in total

Review 1.  Surface Plasmon Resonance (SPR) Spectroscopy and Photonic Integrated Circuit (PIC) Biosensors: A Comparative Review.

Authors:  Patrick Steglich; Giulia Lecci; Andreas Mai
Journal:  Sensors (Basel)       Date:  2022-04-09       Impact factor: 3.847

Review 2.  Diagnoses Based on C-Reactive Protein Point-of-Care Tests.

Authors:  Miroslav Pohanka
Journal:  Biosensors (Basel)       Date:  2022-05-17

3.  Development and Validation of a Reproducible and Label-Free Surface Plasmon Resonance Immunosensor for Enrofloxacin Detection in Animal-Derived Foods.

Authors:  Mingfei Pan; Shijie Li; Junping Wang; Wei Sheng; Shuo Wang
Journal:  Sensors (Basel)       Date:  2017-08-30       Impact factor: 3.576

Review 4.  Silicon Photonic Biosensors Using Label-Free Detection.

Authors:  Enxiao Luan; Hossam Shoman; Daniel M Ratner; Karen C Cheung; Lukas Chrostowski
Journal:  Sensors (Basel)       Date:  2018-10-18       Impact factor: 3.576

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.