| Literature DB >> 28152975 |
Marc Kriege1, Christian Alflen1, Johannes Eisel1, Thomas Ott1, Tim Piepho1, Ruediger R Noppens2,3.
Abstract
BACKGROUND: Recent case reports have indicated significant cuff overinflation when using the standard filling volume based on the manufacturer's recommendations in older models of laryngeal tubes. The aim of this study was to determine the minimum cuff pressure needed to perform standardized ventilation without leakage in the new, revised model of the laryngeal tube "LTS-D".Entities:
Keywords: Airway management; Cuff pressure; Laryngeal tube; Ventilation
Mesh:
Year: 2017 PMID: 28152975 PMCID: PMC5290637 DOI: 10.1186/s12871-017-0308-4
Source DB: PubMed Journal: BMC Anesthesiol ISSN: 1471-2253 Impact factor: 2.217
Fig. 1Old (left) and revised (right) LTS-D. 1: Pilot balloon, 2: color coded connector piece, 3: opening for gastric tube insertion, 4: pharyngeal balloon, 5: multiple ventilation outlets, 6: esophageal balloon, 7: esophageal opening
Fig. 2Study flow chart (CONSORT Flow Chart)
Fig. 3Study setting (revised LTS-D Size #4). A three-way-stop cock allowed simultaneous cuff pressure monitoring and volume removal by using the manufacturer syringe. 1: cuff pressure gauge, 2: three-way-stop cock, 3: color coded syringe for cuff inflation, 4: pilot baloon, 5: inflated pharyngeal- and asophageal cuff
Demographic data and patients characters
| Data | |
|---|---|
| Age (years) | 68 [58.5–76] |
| Female/Male ratio | 33/27 |
| Height (cm) | 168 [162–178] |
| BMI (kg/m)2 | 26 [23–28] |
| Mallampati class I/II/III | 32/27/1 |
| Thyromental distance (mm) | 6 [5–6.5] |
| ASA score I/II/III | 12/26/22 |
| LTS-D Size 3/4/5 | 4/45/11 |
| Surgical duration (min) | 38 [21–62] |
| Propofol Induction doses (mg) | 200 [150–220] (mean 2.5 mg.kg−1) |
| Sufentanil Induction doses (mcg) | 10 [10–15] (mean 0.17 mcg.kg−1) |
Data are given as median [IQR] or absolute numbers
Cuff variables and ventilation variables for LTS-D
| LTS-D size | |||||
|---|---|---|---|---|---|
| #3 | #4 | #5 |
| ||
| After blocking with recommended volume | Cuff pressure [cmH2O] | 120 [110–120] | 110 [85–120] | 105 [90–120] | 0.35 |
| Cuff volume [ml] | 60 [60–60] | 80 [80–80] | 90 [90–90] | <0.001 | |
| Tidal volume [ml] | 389 [366–559] | 482 [449–559] | 640 [613–672] | <0.001 | |
| Peak pressure [cmH2O] | 11 [9.8–16.5] | 13 [12–16] | 11.5 [11–13.3] | 0.14 | |
| Cuff pressure controlled to 60 cmH2O | Cuff pressure [cmH2O] | 60 [60–60] | 60 [60–60] | 60 [60–60] | 1 |
| Cuff volume [ml] | 47 [43.5–53.8] | 70 [65–72.5] | 77 [74.5–82] | <0.001 | |
| Removed cuff volume [ml] | 13 [6–17] | 10 [8–15] | 13 [8–16] | 0.61 | |
| Tidal volume [ml] | 388 [364–424] | 479 [443–532 | 656 [617–694] | <0.001 | |
| Peak pressure [cmH2O] | 12.5 [9.8–16.86] | 13 [12–16] | 11.5 [11–14] | 0.26 | |
| Cuff pressure controlled to 50 cmH2O | Cuff pressure [cmH2O] | 50 [50–50] | 50 [50–50] | 50 [50–50] | 1 |
| Cuff volume [ml] | 47 [28–54.6] | 65 [60–70] | 74 [68.5–78] | <0.001 | |
| Removed cuff volume [ml] | 13 [5–32] | 15 [10–20] | 16 [12–22] | 0.49 | |
| Tidal volume [ml | 366 [312–397] | 477 [442–518] | 665 [627–702] | <0.001 | |
| Peak pressure [cmH2O] | 12.5 [9.8–16.8] | 13 [11–16] | 12 [11–14] | 0.53 | |
| Cuff pressure controlled to 40 cmH2O | Cuff pressure [cmH2O] | 40 [40–40] | 40 [40–40] | 40 [40–40] | 1 |
| Cuff volume [ml] | 49 [40–54] | 60 [56–65] | 68 [64–74] | <0.01 | |
| Removed cuff volume [ml] | 11 [6–20] | 20 [15–25] | 22 [17–27] | 0.09 | |
| Tidal volume [ml | 352 [336–402] | 467 [433–528] | 649 [622–689] | <0.001 | |
| Peak pressure [cmH2O] | 13 [9–18] | 13 [11–16] | 12 [11–14.5] | 0.72 | |
| Cuff pressure controlled to 30 cmH2O | Cuff pressure [cmH2O] | 30 [30–30] | 30 [30–30] | 30 [30–30] | 1 |
| Cuff volume [ml] | 40 [35–45] | 55 [52–60] | 65 [55.8–67.3] | <0.01 | |
| Removed cuff volume [ml] | 20 [15–25] | 25 [20–28] | 25 [23–34] | 0.15 | |
| Tidal volume [ml] | 347 [335–401] | 460 [434–527] | 644 [583–668] | <0.001 | |
| Peak pressure [cmH2O] | 13 [9–18]] | 12 [11–15] | 12 [10.3–14] | 0.68 | |
Data are given as median [IQR]
Sufficient ventilation for different cuff pressures
| LTS-D size | |||||
|---|---|---|---|---|---|
| Cuff pressure | Sufficient ventilation [n] |
| #3 | #4 | #5 |
| Uncontrolled | 54 (100%) | 4/4 | 40/40 | 10/10 | |
| 60 cmH2O | 52 (96%) | 0.5 | 3/4 | 39/40 | 10/10 |
| 50 cmH2O | 51 (94%) | 0.24 | 3/4 | 39/40 | 9/10 |
| 40 cmH2O | 44 (81%) | <0.01 | 3/4 | 33/40 | 8/10 |
| 30 cmH2O | 35 (65%) | <0.001 | 2/4 | 28/40 | 5/10 |
Data are given as median [IQR] or absolute numbers