Literature DB >> 28150873

Dual echo Dixon imaging with a constrained phase signal model and graph cuts reconstruction.

Eric G Stinson1, Joshua D Trzasko1, Joel G Fletcher1, Stephen J Riederer1.   

Abstract

PURPOSE: The purpose of this work is to derive and demonstrate constrained-phase dual-echo Dixon imaging within a maximum likelihood framework solved with a regularized graph-cuts-guided optimization. THEORY AND METHODS: Dual-echo Dixon reconstruction is fundamentally underdetermined; however, adopting a constrained-phase signal model reduces the number of unknowns and the nonlinear problem can be solved under a maximum likelihood framework. Period shifts in the field map (manifesting as fat/water signal swaps) must also be corrected. Here, a regularized cost function promotes a smooth field map and is solved with a graph-cuts-guided greedy binary optimization. The reconstruction shown here is compared to two other prevalent Dixon reconstructions in experimental phantom and human studies.
RESULTS: Reconstructed images of the water and fat signal are shown for a phantom study, and in vivo studies of foot/ankle, pelvis, and CE-MRA of the thighs. The method shown here compared favorably with the other two methods. Large field inhomogeneities on the order of 20 ppm were resolved, thereby avoiding the fat and water signal swaps present in images reconstructed with the other methods.
CONCLUSION: Constrained-phase dual-echo Dixon imaging solved with a regularized graph-cuts-guided optimization has been derived and demonstrated to successfully separate water and fat images in the presence of large magnetic field inhomogeneities. Magn Reson Med 78:2203-2215, 2017.
© 2017 International Society for Magnetic Resonance in Medicine. © 2017 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  Dixon; constrained phase; fat-water; graph cuts

Mesh:

Substances:

Year:  2017        PMID: 28150873      PMCID: PMC5818249          DOI: 10.1002/mrm.26620

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  23 in total

1.  Effect of windowing and zero-filled reconstruction of MRI data on spatial resolution and acquisition strategy.

Authors:  M A Bernstein; S B Fain; S J Riederer
Journal:  J Magn Reson Imaging       Date:  2001-09       Impact factor: 4.813

2.  Multicoil Dixon chemical species separation with an iterative least-squares estimation method.

Authors:  Scott B Reeder; Zhifei Wen; Huanzhou Yu; Angel R Pineda; Garry E Gold; Michael Markl; Norbert J Pelc
Journal:  Magn Reson Med       Date:  2004-01       Impact factor: 4.668

3.  Three-dimensional water/fat separation and T2* estimation based on whole-image optimization--application in breathhold liver imaging at 1.5 T.

Authors:  Johan Berglund; Joel Kullberg
Journal:  Magn Reson Med       Date:  2011-12-21       Impact factor: 4.668

4.  Dual-echo Dixon imaging with flexible choice of echo times.

Authors:  Holger Eggers; Bernhard Brendel; Adri Duijndam; Gwenael Herigault
Journal:  Magn Reson Med       Date:  2011-01       Impact factor: 4.668

5.  A single-point Dixon technique for fat-suppressed fast 3D gradient-echo imaging with a flexible echo time.

Authors:  Jingfei Ma
Journal:  J Magn Reson Imaging       Date:  2008-04       Impact factor: 4.813

6.  Multipoint Dixon technique for water and fat proton and susceptibility imaging.

Authors:  G H Glover
Journal:  J Magn Reson Imaging       Date:  1991 Sep-Oct       Impact factor: 4.813

Review 7.  Chemical shift encoding-based water-fat separation methods.

Authors:  Holger Eggers; Peter Börnert
Journal:  J Magn Reson Imaging       Date:  2014-01-21       Impact factor: 4.813

8.  Simple proton spectroscopic imaging.

Authors:  W T Dixon
Journal:  Radiology       Date:  1984-10       Impact factor: 11.105

9.  Constraining the initial phase in water-fat separation.

Authors:  Mark Bydder; Takeshi Yokoo; Huanzhou Yu; Michael Carl; Scott B Reeder; Claude B Sirlin
Journal:  Magn Reson Imaging       Date:  2010-12-14       Impact factor: 2.546

10.  ESPIRiT--an eigenvalue approach to autocalibrating parallel MRI: where SENSE meets GRAPPA.

Authors:  Martin Uecker; Peng Lai; Mark J Murphy; Patrick Virtue; Michael Elad; John M Pauly; Shreyas S Vasanawala; Michael Lustig
Journal:  Magn Reson Med       Date:  2014-03       Impact factor: 4.668

View more
  4 in total

1.  Robust and efficient pharmacokinetic parameter non-linear least squares estimation for dynamic contrast enhanced MRI of the prostate.

Authors:  Soudabeh Kargar; Eric A Borisch; Adam T Froemming; Akira Kawashima; Lance A Mynderse; Eric G Stinson; Joshua D Trzasko; Stephen J Riederer
Journal:  Magn Reson Imaging       Date:  2017-12-24       Impact factor: 2.546

2.  Whole-body MRI for metastatic cancer detection using T2 -weighted imaging with fat and fluid suppression.

Authors:  Xinzeng Wang; Ali Pirasteh; James Brugarolas; Neil M Rofsky; Robert E Lenkinski; Ivan Pedrosa; Ananth J Madhuranthakam
Journal:  Magn Reson Med       Date:  2018-02-14       Impact factor: 4.668

3.  Time-resolved contrast-enhanced MR angiography with single-echo Dixon fat suppression.

Authors:  Eric G Stinson; Joshua D Trzasko; Norbert G Campeau; James F Glockner; John Huston; Phillip M Young; Stephen J Riederer
Journal:  Magn Reson Med       Date:  2018-02-27       Impact factor: 4.668

4.  Technical Aspects of Contrast-enhanced MR Angiography: Current Status and New Applications.

Authors:  Stephen J Riederer; Eric G Stinson; Paul T Weavers
Journal:  Magn Reson Med Sci       Date:  2017-08-31       Impact factor: 2.471

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.