Shenghui Mei1,2, Jiaqing Wang1, Leting Zhu1, Ruiling Chen1,2, Xingang Li1,2, Kai Chen3, Guangqiang Chen3, Jianxin Zhou3, Qiang Wang3, Zhigang Zhao1,2.
Abstract
Vancomycin (VCM) is clinically used in treating patients with postoperative intracranial infections. The cerebrospinal fluid (CSF) concentration of VCM varies greatly among patients. To guide the dosage regimens, monitoring of VCM in CSF is needed. However a method for analysis of VCM in human CSF is lacking. An ultraperformance liquid chromatography tandem mass spectrometry (UPLC-MS/MS) was developed and validated for analysis of VCM in human CSF, and the agreement of UPLC-MS/MS and chemiluminescence immunoassay (CLIA) in the analysis of CSF VCM was evaluated. The ion transitions were m/z 725.5 > 144.1 for VCM and m/z 455.2 > 308.2 for methotrexate (internal standard). The agreement between UPLC-MS/MS and CLIA was evaluated by Bland-Altman plot in 179 samples. The calibration range of the UPLC-MS/MS method was 1-400 mg/L. The inaccuracy and imprecision were -0.69-10.80% and <4.95%. The internal standard normalized recovery and matrix factor were 86.14-99.31 and 85.84-92.07%, respectively. The measurements of CLIA and UPLC-MS/MS were strongly correlated (r > 0.98). The 95% limit of agreement of the ratio of CLIA to UPLC-MS/MS was 61.66-107.40%. Further studies are warranted to confirm the results.Entities:
Keywords: UPLC-MS/MS; cerebrospinal fluid; chemiluminescence immunoassay; method comparison; method development and validation; vancomycin
Mesh:
Substances:
Year: 2017 PMID: 28139829 DOI: 10.1002/bmc.3939
Source DB: PubMed Journal: Biomed Chromatogr ISSN: 0269-3879 Impact factor: 1.902