Validity of rotational laxity coupled with anterior translation of the knee: A cadaveric study comparing radiostereometry and the Rotab®.

A Senioris1, T Rousseau2, M L'Hermette3, S Gouzy4, F Duparc5, F Dujardin6.   

Abstract

BACKGROUND: In current practice, anterior cruciate ligament (ACL) tears can be diagnosed using several devices to measure anterior tibial translation and rotational knee laxity, but these measures are never collected together. The Rotab®, which yields simultaneous measurements of anterior tibial translation and passive lower limb rotation under stress, would therefore be advantageous in current practice, but its reliability has never been tested.
AIM OF STUDY: To evaluate the accuracy and reliability of the Rotab® compared to the reference system, radiostereometric analysis (RSA).
METHODS: This anatomical study was conducted on 14 lower limbs collected from fresh cadavers. Simultaneous measurements of anterior tibial translation and rotation were obtained with both systems, with antero-posterior (AP) forces of 134N and 250N using the Rotab®. Measurements were made on intact ACL and then repeated after ACL section. Variables were analyzed in univariate analysis by ANOVA, and the intraclass correlation coefficient (ICC) between the systems was determined by the Bland and Altman method.
RESULTS: The difference between the two methods for evaluating anterior tibial translation was 0.05±0.98mm at 134N and 0.29±1.04mm at 250N. The correlation between the tests was high (r134=r250=0.97, p=0.8). The difference between the two methods for rotational laxity was 0.69±2.7° at 134N and 0.5±0.6° at 250N. The Rotab® showed a significant difference only at 250N for rotational laxity after the ACL tear.
CONCLUSION: The Rotab® is a reliable device to measure rotational laxity coupled with anterior translation of the knee.
Copyright © 2017 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Anterior cruciate ligament; Knee arthrometer; Radiostereometry; Reliability; Rotab®

Mesh:

Year:  2017        PMID: 28139332     DOI: 10.1016/j.knee.2017.01.009

Source DB:  PubMed          Journal:  Knee        ISSN: 0968-0160            Impact factor:   2.199


  2 in total

1.  Comparison of a simplified skin pointer device compared with a skeletal marker for knee rotation laxity: A cadaveric study using a rotation-meter.

Authors:  Ken Lee Puah; Andy Khye Soon Yew; Siaw Meng Chou; Denny Tijauw Tjoen Lie
Journal:  World J Orthop       Date:  2018-06-18

2.  Preoperative Knee Instability Affects Residual Instability as Evaluated by Quantitative Pivot-Shift Measurements During Double-Bundle ACL Reconstruction.

Authors:  Yusuke Kawanishi; Masahiro Nozaki; Makoto Kobayashi; Sanshiro Yasuma; Hiroaki Fukushima; Atsunori Murase; Tetsuya Takenaga; Masahito Yoshida; Gen Kuroyanagi; Yohei Kawaguchi; Yuko Nagaya; Hideki Murakami
Journal:  Orthop J Sports Med       Date:  2020-10-19
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.