Literature DB >> 28134130

Chemically tuned linear energy transfer dependent quenching in a deformable, radiochromic 3D dosimeter.

Ellen Marie Høye1, Peter S Skyt, Peter Balling, Ludvig P Muren, Vicki T Taasti, Jan Swakoń, Gabriela Mierzwińska, Marzena Rydygier, Niels Bassler, Jørgen B B Petersen.   

Abstract

Most solid-state detectors, including 3D dosimeters, show lower signal in the Bragg peak than expected, a process termed quenching. The purpose of this study was to investigate how variation in chemical composition of a recently developed radiochromic, silicone-based 3D dosimeter influences the observed quenching in proton beams. The dependency of dose response on linear energy transfer, as calculated through Monte Carlo simulations of the dosimeter, was investigated in 60 MeV proton beams. We found that the amount of quenching varied with the chemical composition: peak-to-plateau ratios (1 cm into the plateau) ranged from 2.2 to 3.4, compared to 4.3 using an ionization chamber. The dose response, and thereby the quenching, was predominantly influenced by the curing agent concentration, which determined the dosimeter's deformation properties. The dose response was found to be linear at all depths. All chemical compositions of the dosimeter showed dose-rate dependency; however this was not dependent on the linear energy transfer. Track-structure theory was used to explain the observed quenching effects. In conclusion, this study shows that the silicone-based dosimeter has potential for use in measuring 3D-dose-distributions from proton beams.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 28134130     DOI: 10.1088/1361-6560/aa512a

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  4 in total

1.  Dose response of three-dimensional silicone-based radiochromic dosimeters for photon irradiation in the presence of a magnetic field.

Authors:  Morten B Jensen; Peter Balling; Simon J Doran; Jørgen B B Petersen; Isak H Wahlstedt; Ludvig P Muren
Journal:  Phys Imaging Radiat Oncol       Date:  2020-10-19

2.  Dose-response of deformable radiochromic dosimeters for spot scanning proton therapy.

Authors:  Simon V Jensen; Lia B Valdetaro; Per R Poulsen; Peter Balling; Jørgen B B Petersen; Ludvig P Muren
Journal:  Phys Imaging Radiat Oncol       Date:  2020-11-20

Review 3.  Radiation Dosimetry by Use of Radiosensitive Hydrogels and Polymers: Mechanisms, State-of-the-Art and Perspective from 3D to 4D.

Authors:  Yves De Deene
Journal:  Gels       Date:  2022-09-19

4.  Empirical quenching correction in radiochromic silicone-based three-dimensional dosimetry of spot-scanning proton therapy.

Authors:  Lia Barbosa Valdetaro; Ellen Marie Høye; Peter Sandegaard Skyt; Jørgen Breede Baltzer Petersen; Peter Balling; Ludvig Paul Muren
Journal:  Phys Imaging Radiat Oncol       Date:  2021-04-12
  4 in total