Examining the influence of humidity on reference ionization chamber performance.

Malcolm R McEwen1, Jaswinder Taank2.   

Abstract

PURPOSE: International dosimetry protocols require measurements made with a vented ionization chamber to be corrected for the influence of air density by using the standard temperature-pressure correction factor. The effect of humidity, on the other hand, is generally ignored with the provision that the relative humidity (RH) is within certain limits (typically 20% to 80%). However, there is little experimental data in the published literature as to the true effect of humidity on modern reference-class ionization chambers. This investigation used two different radiation beams - a Co-60 irradiator and an Sr-90 check source - to examine the effect of humidity on several Farmer-type ionization chambers.
METHODS: An environmental cabinet controlled the humidity. For the Co-60 beam, the irradiation was external, whereas for the Sr-90 measurements, the source itself was placed within the cabinet. Extensive measurements were carried out to ensure that the experimental setup provided reproducible readings. Four chamber types were investigated: IBA FC65-G, IBA FC65-P, PTW 30013 and Exradin A19. The different wall materials provided potentially different mechanical responses (i.e., in terms of expansion/contraction) to the water content in the air. The relative humidity was varied between 8% and 98% and measurements were made with both increasing and decreasing humidity to investigate any possible hysteresis effects.
RESULTS: Measurements in Co-60 were consistent with the published data obtained with primary standard cavity chambers in ICRU Report 31 (i.e., a very small variation <0.1% for RH >10%). The measurements in the Sr-90 field showed no dependence with the relative humidity, within the measurement uncertainties (0.05%, k = 1). Very good repeatability of the ionization current was obtained over successive wet/dry cycles, no hysteresis was observed, and there was no dependence on chamber type.
CONCLUSION: These results indicate that humidity has no significant effect on these particular types of ionization chambers, consistent with recommendations in current megavoltage dosimetry protocols. © Her Majesty the Queen in Right of Canada 2016. Reproduced with the permission of the Minister of Science.

Entities:  

Keywords:  correction factors; humidity; ionization chamber; temperature

Mesh:

Year:  2017        PMID: 28133754     DOI: 10.1002/mp.12057

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  An on-site dosimetry audit for high-energy electron beams.

Authors:  Leon de Prez; Stan Heukelom; Bartel Jansen; Wim Jansen; Jeroen van de Kamer; Wenze van Klink; Elfried Kok; Thijs Perik; Jacco de Pooter; Frits Wittkämper
Journal:  Phys Imaging Radiat Oncol       Date:  2018-02-27

2.  The reliability of radiation dose display of a computed tomography scanner.

Authors:  V Forss; H Yli-Ollila; J Vatanen; P Kölhi; V-P Poutanen; A Palomäki
Journal:  Eur J Radiol Open       Date:  2021-04-05
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.