Cyclic and Load to Failure Properties of All-Suture Anchors in Synthetic Acetabular and Glenoid Cancellous Bone.

Nathan P Douglass1, Anthony W Behn1, Marc R Safran2.   

Abstract

PURPOSE: To evaluate the cyclic displacement, maximum load to failure, and failure mode of multiple all-suture anchors (ASAs) in 2 different densities of sawbones cancellous bone substitute.
METHODS: Anchors tested included the Suturefix Ultra 1.7 mm, JuggerKnot 1.45 mm (No. 1 and No. 2 MaxBraid) and 2.9 mm, Y-Knot Flex 1.3 mm and 1.8 mm, Iconix 1, 2, 25, and 3, Q-Fix 1.8 mm, and Bioraptor 2.3 PK. The Bioraptor served as a non-all-suture-based control. Seven to eleven anchors were tested in both 20 and 30 pounds per cubic foot (pcf) test blocks that were chosen to simulate glenoid and acetabular cancellous bone, respectively. After a 40 N deployment force, anchors were cyclically loaded at 0.5 Hz from 10 to 50 N and then 10 to 100 N for 200 cycles each. Surviving specimens were pulled to failure at 10 mm/s. Displacement, stiffness, maximum load, and failure mode were recorded. Welch t-tests and Welch analysis of variance with Games-Howell post hoc tests were used for statistical analysis.
RESULTS: In higher density blocks, 11 of 12 anchors had significantly (P < .05) higher maximum loads to failure, and 8 anchors showed significantly lower post-cyclic displacement. The Q-Fix 1.8 displayed the lowest post-cyclic displacement in both densities (0.1 ± 0.2 mm, mean ± standard deviation, in both densities). All other groups exhibited at least 2.8 mm and 0.6 mm post-cyclic displacement in 20 and 30 pcf, respectively. The Bioraptor did not survive cyclic testing in 20 pcf and had 0.6 ± 0.3 mm post-cyclic displacement in 30 pcf.
CONCLUSIONS: ASAs show better fixation in higher density synthetic bone. The cyclic displacement and maximum load of ASAs vary widely depending on anchor design and bone density. Most anchors fail by suture anchor pullout. In general, the Bioraptor 2.3 PK outperformed ASAs in higher density test blocks with mixed results in lower density test blocks. CLINICAL RELEVANCE: ASAs show mixed results compared with a traditional suture anchor. They perform better in higher density bone substitute.
Copyright © 2017 Arthroscopy Association of North America. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 28132809     DOI: 10.1016/j.arthro.2016.11.022

Source DB:  PubMed          Journal:  Arthroscopy        ISSN: 0749-8063            Impact factor:   4.772


  6 in total

1.  Suture Tape-Reinforced Human Dermal Allograft Used for Superior Capsule Reconstruction Demonstrates Improved Ability to Withstand Elongation.

Authors:  Cody S Lee; Manoj Reddy; Bryan Scott; Daniel Curtis; Farid Amirouche; Aravind Athiviraham
Journal:  Arthrosc Sports Med Rehabil       Date:  2020-10-24

2.  The Effect of Torque Differences for All-Suture Anchor Fixation Strength: A Biomechanical Analysis.

Authors:  Lucca Lacheta; Jon Miles; Brenton Douglass; Peter Millett
Journal:  Arthrosc Sports Med Rehabil       Date:  2021-02-25

3.  Prospective Analysis of Arthroscopic Hip Anatomic Labral Repair Utilizing Knotless Suture Anchor Technology: The Controlled-Tension Anatomic Technique at Minimum 2-Year Follow-up.

Authors:  David R Maldonado; Sarah L Chen; Jeffery W Chen; Jacob Shapira; Philip J Rosinksy; Shawn Annin; Ajay C Lall; Benjamin G Domb
Journal:  Orthop J Sports Med       Date:  2020-07-27

4.  All-suture anchor and unicortical button show comparable biomechanical properties for onlay subpectoral biceps tenodesis.

Authors:  Alexander Otto; Sebastian Siebenlist; Joshua B Baldino; Matthew Murphy; Lukas N Muench; Julian Mehl; Elifho Obopilwe; Mark P Cote; Andreas B Imhoff; Augustus D Mazzocca
Journal:  JSES Int       Date:  2020-09-21

5.  Clinical Effect of Suture Anchor and Double-Pulley Technique in the Treatment of Inferior Patellar Fracture.

Authors:  Hang Yu; Hui Dong; Binjia Ruan; Xiaohang Xu; Yongxiang Wang; Le Hu
Journal:  Comput Math Methods Med       Date:  2021-12-31       Impact factor: 2.238

Review 6.  The Clinical and Biomechanical Performance of All-Suture Anchors: A Systematic Review.

Authors:  Selim Ergün; Umut Akgün; F Alan Barber; Mustafa Karahan
Journal:  Arthrosc Sports Med Rehabil       Date:  2020-05-28
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.