| Literature DB >> 28127163 |
Iram Zaidi1, Rani Somani2, Shipra Jaidka3, Muhamad Nishad4, Shikha Singh1, Divya Tomar5.
Abstract
AIM: The aim of this study was to compare and evaluate the efficacy of different diagnostic aids for diagnosis of dental caries and to compare the validity in terms of sensitivity and specificity of all four diagnostic modalities for diagnosis of caries.Entities:
Keywords: Caries detector dye (Kuraray); DIAGNOdent; Intraoral camera; Occlusal lesion; Visual examination.
Year: 2016 PMID: 28127163 PMCID: PMC5233698 DOI: 10.5005/jp-journals-10005-1385
Source DB: PubMed Journal: Int J Clin Pediatr Dent ISSN: 0974-7052
Table 1: Criteria used in visual examination (Ekstrand et al)[12]
| V0 | No or slight change in enamel translucency after prolonged air drying (≥5 sec) | ||
| V1 | Opacity hardly visible on the wet surface, but distinctly visible after drying. | ||
| V2 | Opacity distinctly visible without air drying | ||
| V3 | Localized enamel breakdown in opaque or discolored enamel and/or gray discoloration from the underlying dentin | ||
| V4 | Cavitation in opaque or discolored enamel exposing the dentin |
Fig. 1:Diagnosis of caries by visual examination
Fig. 2:Examination with intraoral camera
Fig. 3:Diagnosis with DIAGNOdent
Table 2: Criteria used in examination with DIAGNOdent (Lussi et al)[13]
| 0-14 | No caries | |||
| 15-20 | Enamel caries | |||
| 21-99 | Dentinal caries |
Fig. 4:Dye application
Fig. 5:Operative intervention (pit and fissure opening)
Table 3: Criteria used in examination after pit and fissure opening (Ekstrand et al)[14]
| D0 | No caries | |||
| D1-D2 | Enamel caries | |||
| D3 | Caries up to half of dentin | |||
| D4 | Caries beyond half of dentin |
Table 4: Sensitivity and specificity of each diagnostic method for enamel (D1-D2) and dentin (D3-D4) caries
| Visual | |||||||||
| examination | |||||||||
| Wet | 0.60 | 0.57 | 0.50 | 0.52 | |||||
| Dry | 0.65 | 0.61 | 0.57 | 0.60 | |||||
| Intraoral | 0.75 | 0.80 | 0.63 | 0.74 | |||||
| camera | |||||||||
| Wet | 0.81 | 0.86 | 0.66 | 0.79 | |||||
| Dry | |||||||||
| DIAGNOdent | 0.91 | 0.75 | 0.72 | 0.71 | |||||
| DIAGNOdent | 0.71 | 0.70 | 0.61 | 0.68 | |||||
| with dye | |||||||||
Graph 1:The comparison of sensitivity and specificity of all four tests for enamel caries
Graph 2:The comparison of sensitivity and specificity of all four tests for dentin caries
Table 5: Measurement of interrater agreement (kappa statistics) of all four diagnostic tests
| Measure of agreement kappa | 0.03 | 0.06 | 0.680 | 0.720 | 0.816 | 0.524 | |||||||
Graph 3:The comparison of kappa of all four tests