A microplate assay to measure classical and alternative complement activity.

Bénédicte Puissant-Lubrano1, Françoise Fortenfant1, Peter Winterton2, Antoine Blancher3.   

Abstract

BACKGROUND: We developed and validated a kinetic microplate hemolytic assay (HA) to quantify classical and alternative complement activity in a single dilution of human plasma or serum.
METHODS: The assay is based on monitoring hemolysis of sensitized sheep (or uncoated rabbit) red blood cells by means of a 96-well microplate reader. The activity of the calibrator was evaluated by reference to 200 healthy adults. The conversion of 50% hemolysis time into a percentage of activity was obtained using a calibration curve plotted daily.
RESULTS: The linearity of the assay as well as interference (by hemolysis, bilrubinemia and lipemia) was assessed for classical pathway (CP). The within-day and the between-day precision was satisfactory regarding the performance of commercially available liposome immunoassay (LIA) and ELISA. Patients with hereditary or acquired complement deficiencies were detected (activity was measured <30%). We also provided a reference range obtained from 200 blood donors. The agreement of CP evaluated on samples from 48 patients was 94% with LIA and 87.5% with ELISA. The sensitivity of our assay was better than that of LIA, and the cost was lower than either LIA or ELISA. In addition, this assay was less time consuming than previously reported HAs.
CONCLUSIONS: This assay allows the simultaneous measurement of 36 samples in duplicate per run of a 96-well plate. The use of a daily calibration curve allows standardization of the method and leads to good reproducibility. The same technique was also adapted for the quantification of alternative pathway (AP) activity.

Entities:  

Keywords:  alternative pathway; classical pathway; complement; kinetic hemolytic assay; microplate

Mesh:

Year:  2017        PMID: 28125402     DOI: 10.1515/cclm-2016-0553

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  1 in total

1.  The oxygen carrier M101 alleviates complement activation, which may be beneficial for donor organ preservation.

Authors:  Bénédicte Puissant-Lubrano; Charlène Bouthemy; Nicolas Congy-Jolivet; Jean Milhes; Vincent Minville; Nassim Kamar; Leïla Demini; Franck Zal; Yves Renaudineau
Journal:  Front Immunol       Date:  2022-09-12       Impact factor: 8.786

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.