Size-specific optimization of CT protocols based on minimum detectability.

Yakun Zhang1, Christopher Smitherman2, Ehsan Samei1,2,3.   

Abstract

PURPOSE: To develop a comprehensive model of task-based performance of CT across a broad library of CT protocols, so that radiation dose and image quality can be optimized within a large multivendor clinical facility.
METHODS: Eighty adult CT protocols from the Duke University Medical Center were grouped into 23 protocol groups with similar acquisition characteristics. A size-based image quality phantom (Duke Mercury Phantom 2.0) was imaged using these protocol groups for a range of clinically relevant dose levels on two CT manufacturer platforms (Siemens SOMATOM Definition Flash and GE CT750 HD). For each protocol group, phantom size, and dose level, the images were analyzed to extract task-based image quality metrics, the task transfer function (TTF), and the noise power spectrum (NPS). The TTF and NPS were further combined with generalized models of lesion task functions to predict the detectability of the lesions in terms of areas under the receiver operating characteristic curve (Az ). A graphical user interface (GUI) was developed to present Az as a function of lesion size and contrast, dose, patient size, and protocol, as well as to derive the necessary dose to achieve a detection threshold for a targeted lesion.
RESULTS: The GUI provided the prediction of Az values modeling detection confidence for a targeted lesion, patient size, and dose. As an example, an abdomen pelvis exam for one scanner, with a reference task size/contrast of 5-mm/50-HU, and an Az of 0.9 indicated a dose requirement of 4.0, 8.9, and 16.9 mGy for patient diameters of 25, 30, and 35 cm, respectively. For a constant patient diameter of 30 cm and 50-HU lesion contrast, the minimum detected lesion size at those dose levels were predicted to be 8.4, 5.0, and 3.9 mm, respectively.
CONCLUSIONS: A CT protocol optimization platform was developed by combining task-based detectability calculations with a GUI that demonstrates the tradeoff between dose and image quality. The platform can be used to improve individual protocol dose efficiency, as well as to improve protocol consistency across various patient sizes and CT scanners.
© 2017 American Association of Physicists in Medicine.

Entities:  

Keywords:  CT protocol optimization; GUI; detectability; image quality size-specific protocol optimization; task based performance

Mesh:

Year:  2017        PMID: 28122119     DOI: 10.1002/mp.12125

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

Review 1.  Design for Additive Bio-Manufacturing: From Patient-Specific Medical Devices to Rationally Designed Meta-Biomaterials.

Authors:  Amir A Zadpoor
Journal:  Int J Mol Sci       Date:  2017-07-25       Impact factor: 5.923

2.  Effect of obesity on ability to lower exposure for detection of low-attenuation liver lesions.

Authors:  Brian R Herts; Andrew Schreiner; Frank Dong; Andrew Primak; Jennifer Bullen; Wadih Karim; Douglas Nachand; Sara Hunter; Mark E Baker
Journal:  J Appl Clin Med Phys       Date:  2020-12-28       Impact factor: 2.102

3.  A comprehensive assessment of physical image quality of five different scanners for head CT imaging as clinically used at a single hospital centre-A phantom study.

Authors:  Patrizio Barca; Fabio Paolicchi; Giacomo Aringhieri; Federica Palmas; Daniela Marfisi; Maria Evelina Fantacci; Davide Caramella; Marco Giannelli
Journal:  PLoS One       Date:  2021-01-14       Impact factor: 3.240

Review 4.  Virtual anthropology - a brief review of the literature and history of computed tomography.

Authors:  Tanya Uldin
Journal:  Forensic Sci Res       Date:  2017-09-14

5.  Individualization of computed tomography protocols for suspected pulmonary embolism: a national investigation of routines.

Authors:  Berit Dymbe; Elisabeth Vespestad Mæland; Jorunn Rønhovde Styve; Albertina Rusandu
Journal:  J Int Med Res       Date:  2020-04       Impact factor: 1.671

6.  Assessment of task-based image quality for abdominal CT protocols linked with national diagnostic reference levels.

Authors:  Anaïs Viry; Christoph Aberle; Thiago Lima; Reto Treier; Sebastian T Schindera; Francis R Verdun; Damien Racine
Journal:  Eur Radiol       Date:  2021-07-29       Impact factor: 5.315

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.