BACKGROUND: The increasing complexity of percutaneous coronary intervention (PCI) for chronic total coronary occlusions (CTO) leads to a significant increase of radiation exposure for both patient and operator.
OBJECTIVE: To study the potential of modified settings of the X-ray equipment combined with operator protocols to reduce radiation dose despite increasing procedural complexity. PATIENTS AND METHODS: We analyzed a consecutive cohort of 984 PCIs for CTOs in 863 patients between January 2010 and July 2015. During that period, the X-ray equipment was changed from an analog to a digital detector system, and a subsequent filter and imaging modification was implemented. The fluoroscopy settings were reduced from 15 pulses/s to 7.5, and then to 6. The cine framerate was reduced from 15 to 7.5/s. For the last time period, with optimized settings, procedural, and lesion related factors influencing the radiation exposure were analyzed.
RESULTS: The lesion complexity increased from a J-CTO score of 1.64 to 2.33 with an increase of retrograde procedures from 21.6 to 50.4%. With a similar fluoroscopy time, the dose area product was reduced from period 1 to 2 by 20%, and further by 7% to period 3. There was a significant reduction of Air Kerma from period 2 to 3 from 3.5 to 2.7 Gy. The operator exposure was reduced by more than half. The patient's weight and the complexity of the procedure were the main determinants of radiation exposure.
CONCLUSIONS: The radiation exposure for patient and operator was decreased considerably during the three observation periods despite an increase in lesion and procedural complexity. Rigorous implementation of radiation device settings did reduce radiation exposure without impeding procedural success.
© 2017 Wiley Periodicals, Inc. © 2017 Wiley Periodicals, Inc.

Entities:  

Keywords:  chronic coronary occlusion; percutaneous transluminal intervention; radiation exposure

Mesh:

Year:  2017        PMID: 28112448     DOI: 10.1002/ccd.26886

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  5 in total

Review 1.  Chronic Total Occlusion Interventions: Update on Current Tips and Tricks.

Authors:  Peter Tajti; Iosif Xenogiannis; Dimitris Karmpaliotis; Khaldoon Alaswad; Farouc A Jaffer; M Nicholas Burke; Imre Ungi; Emmanouil S Brilakis
Journal:  Curr Cardiol Rep       Date:  2018-10-22       Impact factor: 2.931

2.  Reduction of operator radiation exposure using a passive robotic device during fluoroscopy-guided arterial puncture: an experimental study in a swine model.

Authors:  Muhammad Umair Ahmad Khan; Chang-Hwan Yoon; Byung-Ju Yi
Journal:  Eur Radiol Exp       Date:  2019-05-29

Review 3.  Chronic Total Occlusion Percutaneous Coronary Intervention in Patients With Prior Coronary Artery Bypass Graft: Current Evidence and Future Perspectives.

Authors:  Lei Guo; Haichen Lv; Xiaomeng Yin
Journal:  Front Cardiovasc Med       Date:  2022-04-11

Review 4.  Chronic Total Occlusion Percutaneous Coronary Intervention: Evidence and Controversies.

Authors:  Peter Tajti; Emmanouil S Brilakis
Journal:  J Am Heart Assoc       Date:  2018-01-12       Impact factor: 5.501

5.  Reduction of radiation exposure to operating physician and assistant using a real-time auditory feedback dosimeter during femoral artery puncturing: a study on swine model.

Authors:  Muhammad Umair Ahmad Khan; Byung-Ju Yi
Journal:  Eur Radiol Exp       Date:  2019-09-23
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.