Measured Head CT/CTA Skin Dose and Intensive Care Unit Patient Cumulative Exposure.

R D Nawfel1,2, G S Young3,4,2.   

Abstract

BACKGROUND AND
PURPOSE: Estimates of cumulative CT/CTA radiation dose based on volumetric CT dose index have raised concern that neurological intensive care unit patient exposures may reach thresholds for deterministic skin injury. Because the accuracy of volumetric CT dose index for this purpose in unknown, we set out to directly measure head CT and CTA peak skin dose, assess the relationship of volumetric CT dose index to measured peak skin dose, and determine whether multiple CT/CTA exposures in typical patients in the neurological intensive care unit produce cumulative doses approaching or exceeding single-dose deterministic thresholds for skin injury.
MATERIALS AND METHODS: In a prospective study from 2011-2013, nanoDot optical stimulated luminescence dosimeters were used to measure head CT/CTA peak skin dose in 52 patients (28 female, 24 male; mean age, 63 years) divided equally between 2 CT scanners. Volumetric CT dose index and dose-length product were recorded for each examination. Peak skin dose was also measured on an acrylic skull phantom in each scanner. A 2-tailed, unpaired t test was used to compare mean patient skin doses between the 2 scanners. The measured peak skin doses were then used to calculate cumulative peak skin dose in 4 typical patients in intensive care units who received multiple CT/CTA scans.
RESULTS: Head CT/CTA peak skin dose agreed between scanners in patients and phantoms: (scanner 1 CT/CTA: patients, 39.2 ± 3.7 mGy and 98.9 ± 5.3 mGy, respectively, versus phantom, 40.0 mGy and 105.4 mGy, respectively; scanner 2 CT/CTA: patients, 42.9 ± 9.4 mGy and 98.8 ± 7.4 mGy, respectively, versus phantom, 37.6 mGy and 95.2 mGy, respectively). Volumetric CT dose index overestimated peak skin dose by a factor of 1.4-1.9 depending on examination and CT scanner. Cumulative doses in 4 patients in the intensive care unit estimated from measured CT/CTA peak skin dose ranged from 1.9-4.5 Gy.
CONCLUSIONS: Directly measured radiation skin doses from head CT/CTA patient examinations are substantially lower than volumetric CT dose index. Measured peak skin dose confirms that multiple head CT/CTA examinations in representative patients in the neurological intensive care unit may produce cumulative doses exceeding the single-dose deterministic threshold for skin injury.
© 2017 by American Journal of Neuroradiology.

Entities:  

Mesh:

Year:  2017        PMID: 28104641      PMCID: PMC7959989          DOI: 10.3174/ajnr.A5040

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  22 in total

Review 1.  What are the risks from medical X-rays and other low dose radiation?

Authors:  B F Wall; G M Kendall; A A Edwards; S Bouffler; C R Muirhead; J R Meara
Journal:  Br J Radiol       Date:  2006-04       Impact factor: 3.039

Review 2.  CT dosimetry: comparison of measurement techniques and devices.

Authors:  John A Bauhs; Thomas J Vrieze; Andrew N Primak; Michael R Bruesewitz; Cynthia H McCollough
Journal:  Radiographics       Date:  2008 Jan-Feb       Impact factor: 5.333

3.  The linear no-threshold relationship is inconsistent with radiation biologic and experimental data.

Authors:  Maurice Tubiana; Ludwig E Feinendegen; Chichuan Yang; Joseph M Kaminski
Journal:  Radiology       Date:  2009-04       Impact factor: 11.105

4.  Estimating peak skin and eye lens dose from neuroperfusion examinations: use of Monte Carlo based simulations and comparisons to CTDIvol, AAPM Report No. 111, and ImPACT dosimetry tool values.

Authors:  Di Zhang; Chris H Cagnon; J Pablo Villablanca; Cynthia H McCollough; Dianna D Cody; Maria Zankl; John J Demarco; Michael F McNitt-Gray
Journal:  Med Phys       Date:  2013-09       Impact factor: 4.071

Review 5.  Does scientific evidence support a change from the LNT model for low-dose radiation risk extrapolation?

Authors:  Dietrich Averbeck
Journal:  Health Phys       Date:  2009-11       Impact factor: 1.316

6.  Peak skin and eye lens radiation dose from brain perfusion CT based on Monte Carlo simulation.

Authors:  Di Zhang; Chris H Cagnon; J Pablo Villablanca; Cynthia H McCollough; Dianna D Cody; Donna M Stevens; Maria Zankl; John J Demarco; Adam C Turner; Maryam Khatonabadi; Michael F McNitt-Gray
Journal:  AJR Am J Roentgenol       Date:  2012-02       Impact factor: 3.959

7.  Intravenous heparin for the treatment of intraluminal thrombus in patients with acute ischemic stroke: a case series.

Authors:  Maxim Mokin; Tareq Kass-Hout; Omar Kass-Hout; Vladan Radovic; Adnan H Siddiqui; Elad I Levy; Kenneth V Snyder
Journal:  J Neurointerv Surg       Date:  2012-02-19       Impact factor: 5.836

8.  Dose exposure of patients undergoing comprehensive stroke imaging by multidetector-row CT: comparison of 320-detector row and 64-detector row CT scanners.

Authors:  S Diekmann; E Siebert; R Juran; M Roll; W Deeg; H-C Bauknecht; F Diekmann; R Klingebiel; G Bohner
Journal:  AJNR Am J Neuroradiol       Date:  2010-01-28       Impact factor: 3.825

9.  Radiation exposure in patients with subarachnoid hemorrhage: a quality improvement target.

Authors:  Judith M Wong; Allen L Ho; Ning Lin; Georgios A Zenonos; Christopher B Martel; Kai Frerichs; Rose Du; William B Gormley
Journal:  J Neurosurg       Date:  2013-04-26       Impact factor: 5.115

10.  Acute brain infarct: detection and delineation with CT angiographic source images versus nonenhanced CT scans.

Authors:  Erica C S Camargo; Karen L Furie; Aneesh B Singhal; Luca Roccatagliata; Mary E Cunnane; Elkan F Halpern; Gordon J Harris; Wade S Smith; Ramon G Gonzalez; Walter J Koroshetz; Michael H Lev
Journal:  Radiology       Date:  2007-06-20       Impact factor: 11.105

View more
  3 in total

1.  Regarding "Measured Head CT/CTA Skin Dose and Intensive Care Unit Patient Cumulative Exposure".

Authors:  I Ikuta; A Mustafa; M H Johnson
Journal:  AJNR Am J Neuroradiol       Date:  2017-05-25       Impact factor: 3.825

2.  Reply.

Authors:  R D Nawfel; G S Young
Journal:  AJNR Am J Neuroradiol       Date:  2017-05-25       Impact factor: 3.825

3.  Application of multiple injections of contrast agent in head and neck CT arteriovenous angiography in children: Study protocol clinical trial (SPIRIT Compliant).

Authors:  Shilong Tang; Guanping Zhang; Zhuo Chen; Xianfan Liu; Xiao Fan; Daisong Liu; Ling He
Journal:  Medicine (Baltimore)       Date:  2020-04       Impact factor: 1.817

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.