Literature DB >> 28104487

Comparison of a special designed high intensity coil to a standard round coil-TMS-setting.

Andrea Szelényi1, Julia Wölfle2, Igor Fischer2, Hans-Jakob Steiger2, Michael Sabel2.   

Abstract

BACKGROUND: Routine diagnostic Transcranial Magnetic Stimulation (TMS) is performed with a round coil (RC) for cranial and spinal root stimulation, being less successful for motor evoked potentials (MEP) of lower limb muscles. MEP elicited with a special configured flat figure-of-eight coil designed for high intensity stimulation (HI-coil) were compared to RC with regard to handling, efficiency, and physiological properties of MEP. NEW
METHODS: MEP elicited with HI-coil and 9-cm diameter RC for cortical, spinal and peripheral stimulation (PES) were compared for Motor threshold (MT), latencies and amplitudes of bilateral Abductor pollicis brevis (APB) and Abductor hallucis muscles (AH). RESULTS AND COMPARISON WITH EXISTING
METHODS: MT for HI-coil were significantly lower for cortical and spinal root stimulation compared to RC (APB: 37% vs. 48%; AH: 58% vs. 72%). MEP-latencies elicited with HI-coil and RC were without significant difference. AH-MEP amplitudes were significantly larger for HI-coil cortical (705±980μV vs. 370±280μV) and root stimulation (260±210μV vs. 151±100μV). Amplitudes elicited by PES compared to HI-coil were always significantly larger.
CONCLUSION: Results for cortical and spinal root stimulation with regards to latencies and amplitudes for APB were equivalent between HI-coil and RC. PES was superior in achieving supramaximal stimulation in terms of amplitudes. The use of the HI-coil might be advantageous for MEP of lower extremity muscles with emphasis on pathologic conditions requiring higher stimulation intensities.
Copyright © 2017 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Flat figure-of-eight coil (HI coil); Motor evoked potentials; Round coil; Transcranial magnetic stimulation

Mesh:

Year:  2017        PMID: 28104487     DOI: 10.1016/j.jneumeth.2017.01.006

Source DB:  PubMed          Journal:  J Neurosci Methods        ISSN: 0165-0270            Impact factor:   2.390


  3 in total

1.  Relationship between body mass and clinical response to repetitive transcranial magnetic stimulation (rTMS) for major depressive disorder.

Authors:  Maria Cristina Davila; Brianna Ely; Ann M Manzardo
Journal:  Ment Illn       Date:  2019-06-11

2.  Time to remission analysis for major depressive disorder after repetitive transcranial magnetic stimulation (rTMS).

Authors:  Ann M Manzardo; Brianna Ely; Maria Cristina Davila
Journal:  Ment Illn       Date:  2019-06-11

3.  Repetitive transcranial magnetic stimulation (rTMS) using different TMS instruments for major depressive disorder at a suburban tertiary clinic.

Authors:  Maria Cristina Davila; Brianna Ely; Ann M Manzardo
Journal:  Ment Illn       Date:  2019-03-22