| Literature DB >> 28103837 |
Kai Li1, Zhongzhen Su1, Erjiao Xu1, Qiannan Huang1, Qingjing Zeng1, Rongqin Zheng2.
Abstract
BACKGROUND: To assess the accuracy of contrast-enhanced ultrasound (CEUS)-CT/MR image fusion in evaluating the radiofrequency ablative margin (AM) of hepatocellular carcinoma (HCC) based on a custom-made phantom model and in HCC patients.Entities:
Keywords: CEUS; CT; Image fusion; Phantom model; Tumor ablation
Mesh:
Substances:
Year: 2017 PMID: 28103837 PMCID: PMC5244580 DOI: 10.1186/s12885-017-3061-7
Source DB: PubMed Journal: BMC Cancer ISSN: 1471-2407 Impact factor: 4.430
Fig. 1a A spherical tumor model (2 cm in diameter) made of carrageenan in red. b Section of the AM model: the carrageenan tumor model surrounded by the 5-mm AM gel in white. c Cylindrical-shaped parenchyma model (10 cm in height and 10 cm in diameter of the upper and lower plane). d Section of the cylindrical-shaped parenchyma model. e CT image showing the tumor. The scope of AM could not accurately evaluated. f US image showing the tumor. The boundary between the tumor and AM gel was not clear
Fig. 2(A-1) and (B-1): CT images of the models. (A-2) and (B-2): US-CT image fusion of the area of ablation (left arrow). The AM was not fully encompassed by the area of ablation (right arrow, A-2). The tumor and AM were completely encompassed by the area of ablation (B-2). (A-3): The AM was not completely encompassed by the area of ablation in the gross specimen (left arrow). (B-3): The AM was completely encompassed by the area of ablation in the gross specimen. The tumor is shown in blue, and the AM gel is indicated in red (5 mm)
Fig. 3Medical images of case 1 in the clinical study. (A-1) CUES image of the area of ablation (dark). (A-2). Preoperative MR image (tumor shown in blue, and 5-mm AM shown in yellow). (A-3) CEUS-MR fusion image. The tumor and 5-mm AM were fully encompassed by the area of ablation. b MR-MR fusion image of case 1. Upper panel: preoperative MR images (tumor shown in blue, and 5-mm AM shown in yellow). Middle panel: postoperative MR images (arrow, ablation area shown in dark color). Lower panel: MR-MR fusion images showing that the tumor and the 5-mm AM were fully encompassed by the area of ablation
Fig. 4Medical images of case 2 in the clinical study. (A-2) Preoperative MR image (tumor shown in blue and 5-mm AM in yellow). (A-1) CEUS-MR fusion image showing that the area of ablation encompassed the tumor but not the entire AM due to the influence of vessels (white arrow). b MR-MR fusion image from the same patient. Upper panel: preoperative MR images (tumor shown in blue and 5-mm AM in yellow). Middle panel: postoperative MR images (arrow, area of ablation shown in dark color). Lower panel: MR-MR fusion images showing that the tumor and AM were not fully encompassed by the area of ablation (white arrow)
The ultrasound echo, CT density and color of the phantom models
| Model | Ultrasound echo | CT density | Color |
|---|---|---|---|
| Parenchyma model | Hypoechoica | Low densityb | Colorless semitransparent |
| Tumor model | Hyperechoic | High density | Red opaque |
| AM model | Hypoechoica | Low densityb | Milky opaque |
| Ablative area model | Slightly hyperechoic | -------- | Carnation semitransparent |
aThese two had the same echogenecity, bthese two had the same density
The results of AM evaluated by US-CT image fusion and gross specimen (P > 0.05)
| Gross specimen | Total | |||
|---|---|---|---|---|
| AM covered | AM not covered | |||
| US-CT image fusion | AM covered | 7 | 1 | 8 |
| AM not covered | 0 | 15 | 15 | |
| Total | 7 | 16 | 23 | |
The clinical characteristics of the patients and HCC lesions
| Characteristics | Number |
|---|---|
| Gender (Male/Female) | 26/0 |
| Age (mean ± standard deviation, years) | |
| Virus hepatitis/alcoholic liver disease/no diffuse hepatic disease | 26/0/0 |
| Liver cirrhosis (yes/no) | / |
| Tumor number (1/>1) | 22/4 |
| Tumor diameter (mean ± SD, mm) | 19.1 ± 6.5 |
HCC hepatocellular carcinoma, M median, QR interquartile range
The results of AM evaluated by CEUS-CT/MR image fusion and MR/MR image fusion
| MR/MR image fusion | Total | |||
|---|---|---|---|---|
| AM covered | AM not covered | |||
| CEUS-CT/MR image fusion | AM covered | 15 | 3 | 18 |
| AM not covered | 0 | 12 | 12 | |
| Total | 15 | 15 | 30 | |