Use of parallel-plate ionization chambers in reference dosimetry of NOVAC and LIAC® mobile electron linear accelerators for intraoperative radiotherapy: a multi-center survey.

Paolo Scalchi1, Alessia Ciccotelli2, Giuseppe Felici2, Assunta Petrucci3, Raffaella Massafra4, Valeria Piazzi5, Paola D'Avenia6, Francesca Cavagnetto7, Federica Cattani8, Raffaella Romagnoli9, Antonella Soriani10.   

Abstract

PURPOSE: LIAC® and NOVAC are two mobile linear accelerators dedicated to intraoperative radiation therapy (IORT), generating electron beams in the energy range of 3-12 MeV. Due to high dose-per-pulse (up to 70 mGy per pulse), in 2003 the Italian National Institute of Health (ISS) stated that "for the measure of dose to water in reference conditions, ionization chambers cannot be employed and no published dosimetry protocol can be used". As a consequence, ferrous sulphate (or, alternatively alanine) dosimetry was recommended. Based on a retrospective multi-center survey, a comparison with ferrous sulphate dosimetry is now used to validate the parallel-plate ionization chambers for reference dosimetry of NOVAC and LIAC.
METHODS: The IAEA TRS-398 dosimetry protocol was applied except for the reference irradiation setup and the determination of the ion-recombination correction factor ks . Moreover the depth of maximum dose (R100 ) instead of zref as measurement depth was chosen by the majority of centers, thus implying a renormalization of the beam-quality correction factor kQ,Qo , based on water-air stopping power ratios. Regarding the ks determination, a previously published method, independent of ferrous sulphate dosimetry, was adopted. All the centers participating in this study had used both ferrous sulphate dosimeters and ionization chambers in water phantoms for dosimetry under reference conditions.
RESULTS: The mean percentage difference between ionization chambers and ferrous sulphate dosimetry was -0.5% with a dispersion of 3.9% (2σ). Moreover, the uncertainty analysis allowed the agreement between ionization chambers and ferrous sulphate dosimetry to be verified. These results did not show any significant dependence on electron energy, thus indirectly confirming kQ,Qo renormalization. The results from the centers using zref as the measurement depth were similar to the other data, but further focused studies could aim at investigating possible dependences of the dose differences on the chosen reference depth.
CONCLUSION: The present study confirms that parallel-plate ionization chambers can properly and accurately substitute ferrous sulphate detectors in reference dosimetry of LIAC and NOVAC mobile linear accelerators. Therefore, we hope that the most commonly used protocols for reference dosimetry in external-beam radiotherapy will be updated in order to provide guidance in the calibration of electron beams from linear accelerators dedicated to IORT, so that users may benefit from specific, authoritative and up-to-date recommendations.
© 2016 American Association of Physicists in Medicine.

Entities:  

Keywords:  zzm321990IOERTzzm321990; electron beams; reference dosimetry

Mesh:

Substances:

Year:  2017        PMID: 28102948     DOI: 10.1002/mp.12020

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  Comparison of Local Recurrence Among Early Breast Cancer Patients Treated With Electron Intraoperative Radiotherapy vs Hypofractionated Photon Radiotherapy an Observational Study.

Authors:  Marina Guenzi; Elisabetta Bonzano; Renzo Corvò; Francesca Merolla; Alice Pastorino; Francesca Cavagnetto; Stefania Garelli; Carlo Alberto Cutolo; Daniele Friedman; Liliana Belgioia
Journal:  Front Oncol       Date:  2018-06-05       Impact factor: 6.244

2.  Verification of electron beam parameters in an intraoperative linear accelerator using dosimetric and radiobiological response methods.

Authors:  Agnieszka Dróżdż; Martyna Waluś; Marcin Zieliński; Bożena Malesa; Marta Kruszyna-Mochalska; Katarzyna Kulcenty; Beata Adamczyk; Piotr Nowaczyk; Julian Malicki; Jacek Pracz
Journal:  Rep Pract Oncol Radiother       Date:  2021-12-30
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.