| Literature DB >> 28101422 |
Paritosh Pande1, Ryan L Shelton1, Guillermo L Monroy2, Ryan M Nolan1, Stephen A Boppart3.
Abstract
We report on the development of a low-cost hand-held low-coherence interferometric imaging system based on the principle of linear optical coherence tomography (Linear OCT), a technique which was first proposed in the early 2000s as a simpler alternative to the conventional time-domain and Fourier-domain OCT. A bench-top implementation of the proposed technique is first presented and validated. The axial resolution, SNR, and sensitivity roll-of of the system was estimated to be 5.2 μm and 80 dB, and 3.7 dB over a depth of 0.15 mm, respectively. After validating the bench-top system, two hand-held probe implementations for contact-based imaging and in vivo human tympanic membrane imaging are presented. The performance of the proposed system was compared with a research-grade state-of-the-art Fourier-domain low coherence interferometry (LCI) system by imaging several biological and non-biological samples. The results of this study suggest that the proposed system might be a suitable choice for applications where imaging depth and SNR can be traded for lower cost and simpler optical design.Entities:
Keywords: (110.3175) Interferometric imaging; (170.0110) Imaging systems; (170.4500) Optical coherence tomography
Year: 2016 PMID: 28101422 PMCID: PMC5231303 DOI: 10.1364/BOE.8.000338
Source DB: PubMed Journal: Biomed Opt Express ISSN: 2156-7085 Impact factor: 3.732