Literature DB >> 28100979

Late Cardiac Rupture after Amplatzer Septal Occluder Implantation.

Ahmet Arnaz, Yasemin Turkekul, Yusuf Yalcinbas, Arda Saygili, Tayyar Sarioglu.   

Abstract

Transcatheter device implantation has become an attractive alternative to surgery in the closure of atrial septal defects in selected patients. However, it can lead to early and late sequelae, some of them life threatening. For example, 79 days before her admission to our emergency department with sudden-onset respiratory distress and respiratory arrest (leading to cardiac tamponade and rupture), a 22-year-old woman had undergone percutaneous closure of an atrial septal defect. We describe the damage and its treatment. Although the adverse effects of transcatheter device implantation are rare, physicians should know that these events can be life threatening. Further data are needed to prevent such sequelae and to design new devices. It is of utmost importance that patients and their family members be informed both of possible sequelae and of life-saving interventions to be administered at early diagnosis.

Entities:  

Keywords:  Balloon occlusion/adverse effects/instrumentation/standards; device removal; heart atria/injuries; heart septal defects, atrial/therapy; minimally invasive surgical procedures/adverse effects; prostheses and implants/adverse effects; septal occluder device/adverse effects; treatment outcome; wounds, penetrating/etiology

Mesh:

Year:  2016        PMID: 28100979      PMCID: PMC5179165          DOI: 10.14503/THIJ-15-5635

Source DB:  PubMed          Journal:  Tex Heart Inst J        ISSN: 0730-2347


  6 in total

1.  Surgery for complications of trans-catheter closure of atrial septal defects: a multi-institutional study from the European Congenital Heart Surgeons Association.

Authors:  George E Sarris; George Kirvassilis; Prodromos Zavaropoulos; Emre Belli; Hakan Berggren; Thierry Carrel; Juan V Comas; Antonio F Corno; Willem Daenen; Duccio Di Carlo; Tjark Ebels; Jose Fragata; Leslie Hamilton; Viktor Hraska; Jeffrey Jacobs; Stojan Lazarov; Constantine Mavroudis; Dominique Metras; Jean Rubay; Christian Schreiber; Giovanni Stellin
Journal:  Eur J Cardiothorac Surg       Date:  2010-03-28       Impact factor: 4.191

Review 2.  Cardiac perforation after device closure of atrial septal defects with the Amplatzer septal occluder.

Authors:  Abhay Divekar; Tidimogo Gaamangwe; Nasir Shaikh; Michael Raabe; John Ducas
Journal:  J Am Coll Cardiol       Date:  2005-04-19       Impact factor: 24.094

3.  Closure of atrial septal defects in children: surgery versus Amplatzer device implantation.

Authors:  Jacek Bialkowski; Blandyna Karwot; Malgorzata Szkutnik; Pawel Banaszak; Jacek Kusa; Janusz Skalski
Journal:  Tex Heart Inst J       Date:  2004

4.  Erosion of Amplatzer septal occluder device after closure of secundum atrial septal defects: review of registry of complications and recommendations to minimize future risk.

Authors:  Zahid Amin; Ziyad M Hijazi; John L Bass; John P Cheatham; William E Hellenbrand; Charles S Kleinman
Journal:  Catheter Cardiovasc Interv       Date:  2004-12       Impact factor: 2.692

Review 5.  Transcatheter device closure of atrial septal defects: a safety review.

Authors:  John Moore; Sanjeet Hegde; Howaida El-Said; Robert Beekman; Lee Benson; Lisa Bergersen; Ralf Holzer; Kathy Jenkins; Richard Ringel; Jonathan Rome; Robert Vincent; Gerard Martin
Journal:  JACC Cardiovasc Interv       Date:  2013-05       Impact factor: 11.195

6.  Late cardiac perforation following transcatheter atrial septal defect closure.

Authors:  Ourania Preventza; Sridhar Sampath-Kumar; John Wasnick; Jeffrey P Gold
Journal:  Ann Thorac Surg       Date:  2004-04       Impact factor: 4.330

  6 in total
  4 in total

1.  Clinical Use of the Amplatzer Device in the Management of Intracardiac Defects: A Single-Center Experience.

Authors:  Seema S Rao; Tyler L Scaff; John Gurley; Sibu Saha
Journal:  Int J Angiol       Date:  2018-01-30

Review 2.  Atrial septal defect (ASD) device trans-catheter closure: limitations.

Authors:  Alessia Faccini; Gianfranco Butera
Journal:  J Thorac Dis       Date:  2018-09       Impact factor: 2.895

3.  A case report of a late left atrial appendage perforation 4 months after occluder implant: reason for or caused by a resuscitation?

Authors:  Karsten Schenke; Stephan Geidel; Christian Keller; Gerian Grönefeld
Journal:  Eur Heart J Case Rep       Date:  2019-10-04

4.  Atrio-aortic erosion caused by Amplatzer Atrial Septal Occluder - a case report.

Authors:  Christine E Kamla; Joscha Buech; Philipp M Doldi; Christian Hagl; Gerd Juchem; Alexey Dashkevich
Journal:  J Cardiothorac Surg       Date:  2021-03-20       Impact factor: 1.637

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.