Literature DB >> 28099995

Technical Note: Fast respiratory motion estimation using sorted singles without unlist processing: A feasibility study.

Jaewon Yang1, Mehdi Khalighi2, Thomas A Hope1,3, Karen Ordovas1, Youngho Seo1,4,5.   

Abstract

PURPOSE: The study aims to demonstrate the feasibility of fast respiratory motion estimation using singles data available as a sorted format in list-mode files acquired in an integrated positron emission tomography/magnetic resonance imaging (PET/MRI) system for a proof-of-concept.
METHODS: The derivation of singles-driven respiratory motion (SDRM) is enabled by singles recorded and binned by second for each detector crystal in PET list-mode data acquired in a SIGNA PET/MR. The proposed method is to derive a SDRM trace by summing up all singles from all detectors through the PET data acquisition. To assess the feasibility of SDRM for data-driven gating (DDG), SDRM traces were derived from the list-mode data acquired in five liver-focused 68 Ga-DOTA-TOC PET/MRI scans, and compared with the traces derived from bellows (pressure belt). Pearson's correlation coefficients and trigger time differences at peak-inhalation phases between SDRM and bellows traces were measured for quantitative evaluation.
RESULTS: The method presented the average processing time of 4.2 ± 0.42 s (range: 3.9 ~ 4.7 s) for the derivation of SDRM traces. The majority of the time was spent for reading singles data from a list-mode file (3.1 ± 0.40 s, range: 2.7 ~ 3.7s). On average, the correlation coefficient of SDRM and bellows traces was 0.69 ± 0.16 (range: 0.41 ~ 0.80) and the time offset of SDRM-driven triggers from bellows-driven triggers was 0.25 ± 0.39 s (range: -0.85 ~ 2.69 s later than bellows triggers), demonstrating the similar patterns and phases of SDRM and bellows traces.
CONCLUSIONS: We introduced PET singles-driven respiratory motion (SDRM) estimation as a proof-of-principle, using sorted singles ready for immediate processing in list-mode data. The results demonstrated the feasibility of SDRM and its potential use for gated PET with fast processing time.
© 2017 American Association of Physicists in Medicine.

Entities:  

Keywords:  PET/MRI; data-driven gating; respiratory motion; singles

Mesh:

Year:  2017        PMID: 28099995     DOI: 10.1002/mp.12115

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  A Computational Framework for Data Fusion in MEMS-Based Cardiac and Respiratory Gating.

Authors:  Mojtaba Jafari Tadi; Eero Lehtonen; Jarmo Teuho; Juho Koskinen; Jussi Schultz; Reetta Siekkinen; Tero Koivisto; Mikko Pänkäälä; Mika Teräs; Riku Klén
Journal:  Sensors (Basel)       Date:  2019-09-24       Impact factor: 3.576

2.  Impact of low injected activity on data driven respiratory gating for PET/CT imaging with continuous bed motion.

Authors:  Joseph G Meier; Radwan H Diab; Trevor M Connor; Osama R Mawlawi
Journal:  J Appl Clin Med Phys       Date:  2022-04-28       Impact factor: 2.243

  2 in total