Repeatability of whole-cornea measurements using an anterior segment imaging device based on OCT and Placido-disk.

Noelia Martínez-Albert1, José J Esteve-Taboada1, Robert Montés-Micó1.   

Abstract

BACKGROUND: Current advances in refractive surgery require high accuracy of corneal measurements. We aim to evaluate the repeatability of topographic measurements using an integrated Placido disk - OCT device.
METHODS: Thirty right eyes of thirty healthy and young subjects were included. The topographic parameters of simulated keratometry, asphericity, minimum pachymetry, white-to-white distance, spherical aberration and corneal elevation were studied. Three measurements were taken by the same observer using the Visante omni device. The coefficient of repeatability (CoR), coefficient of variation (CoV) and intraclass correlation coefficient (ICC) were calculated to assess the repeatability. The Bland-Altman graphical method was also applied.
RESULTS: No significant differences between repeated measurements were found. The mean difference for simulated keratometry was between 0.01 and 0.02 D. For asphericity was -0.01 ± 0.02, and for minimum pachymetry, spherical aberration and WTW was equal or less than 0.02 mm. The mean differences for corneal elevation, in millimeters and diopters, were between 0.00 and 0.01. Overall, most topographic parameters had a CoV less than 0.4%, a CoR less than 0.3 and an ICC higher than 0.98.
CONCLUSIONS: The Visante omni device provides good repeatability for topographical measurements of simulated keratometry, asphericity, pachymetry, spherical aberration and corneal elevation.

Entities:  

Keywords:  Placido disk; cornea; optical coherence tomography; repeatability; topography

Mesh:

Year:  2017        PMID: 28097917     DOI: 10.1080/17434440.2017.1281739

Source DB:  PubMed          Journal:  Expert Rev Med Devices        ISSN: 1743-4440            Impact factor:   3.166


  2 in total

1.  Repeatability assessment of anterior segment biometric measurements under accommodative and nonaccommodative conditions using an anterior segment OCT.

Authors:  Noelia Martínez-Albert; Jose J Esteve-Taboada; Robert Montés-Micó
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2017-11-02       Impact factor: 3.117

2.  The precision and agreement of corneal thickness and keratometry measurements with SS-OCT versus Scheimpflug imaging.

Authors:  Yune Zhao; Ding Chen; Giacomo Savini; Qing Wang; Hongfang Zhang; Yili Jin; Benhao Song; Rui Ning; Jinhai Huang; Chenyang Mei
Journal:  Eye Vis (Lond)       Date:  2020-06-09
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.