Material-Dependent Implant Artifact Reduction Using SEMAC-VAT and MAVRIC: A Prospective MRI Phantom Study.

Lukas Filli1, Lukas Jud, Roger Luechinger, Daniel Nanz, Gustav Andreisek, Val M Runge, Sebastian Kozerke, Nadja A Farshad-Amacker.   

Abstract

OBJECTIVE: The aim of this study was to compare the degree of artifact reduction in magnetic resonance imaging achieved with slice encoding for metal artifact correction (SEMAC) in combination with view angle tilting (VAT) and multiacquisition variable resonance image combination (MAVRIC) for standard contrast weightings and different metallic materials.
METHODS: Four identically shaped rods made of the most commonly used prosthetic materials (stainless steel, SS; titanium, Ti; cobalt-chromium-molybdenum, CoCr; and oxidized zirconium, oxZi) were scanned at 3 T. In addition to conventional fast spin-echo sequences, metal artifact reduction sequences (SEMAC-VAT and MAVRIC) with varying degrees of artifact suppression were applied at different contrast weightings (T1w, T2w, PDw). Two independent readers measured in-plane and through-plane artifacts in a standardized manner. In addition, theoretical frequency-offset and frequency-offset-gradient maps were calculated. Interobserver agreement was assessed using intraclass correlation coefficient.
RESULTS: Interobserver agreement was almost perfect (intraclass correlation coefficient, 0.86-0.99). Stainless steel caused the greatest artifacts, followed by CoCr, Ti, and oxZi regardless of the imaging sequence. While for Ti and oxZi rods scanning with weak SEMAC-VAT showed some advantage, for SS and CoCr, higher modes of SEMAC-VAT or MAVRIC were necessary to achieve artifact reduction. MAVRIC achieved better artifact reduction than SEMAC-VAT at the cost of longer acquisition times. Simulations matched well with the apparent geometry of the frequency-offset maps.
CONCLUSIONS: For Ti and oxZi implants, weak SEMAC-VAT may be preferred as it is faster and produces less artifact than conventional fast spin-echo. Medium or strong SEMAC-VAT or MAVRIC modes are necessary for significant artifact reduction for SS and CoCr implants.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 28092272     DOI: 10.1097/RLI.0000000000000351

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  5 in total

Review 1.  Understanding Magnetic Resonance Imaging of Knee Cartilage Repair: A Focus on Clinical Relevance.

Authors:  Daichi Hayashi; Xinning Li; Akira M Murakami; Frank W Roemer; Siegfried Trattnig; Ali Guermazi
Journal:  Cartilage       Date:  2017-06-05       Impact factor: 4.634

Review 2.  Imaging update in arthroplasty.

Authors:  Nicholas Ong; Izwan Zailan; Ankit Tandon
Journal:  J Clin Orthop Trauma       Date:  2021-10-21

3.  Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences.

Authors:  Tim Hilgenfeld; Marcel Prager; Franz S Schwindling; Johann M E Jende; Peter Rammelsberg; Martin Bendszus; Sabine Heiland; Alexander Juerchott
Journal:  J Vis Exp       Date:  2018-05-17       Impact factor: 1.355

Review 4.  [Metallic Artifacts on MR Imaging and Methods for Their Reduction].

Authors:  Hye Jung Choo; Sun Joo Lee; Young Han Lee
Journal:  Taehan Yongsang Uihakhoe Chi       Date:  2020-01-31

5.  Evaluating Compressed SENSE (CS) MRI Metal Artifact Reduction Using Pig L-Spine Phantom and Transplant Patients: Focused on the CS-SEMAC (SPIR), mDixon(O-MAR) and STIR Techniques.

Authors:  Eun-Hoe Goo; Sung-Soo Kim
Journal:  Tomography       Date:  2022-09-15
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.