BACKGROUND: Coronal alignment is an important factor for the function and longevity of total knee arthroplasty (TKA). Coronal bowing of the lower extremity is common among Asians and it may pose a risk for malalignment of the lower leg and malposition of component. HYPOTHESIS: We hypothesized that coronal bowing itself has a risk for malalignment of the lower leg and malposition of femoral/tibial components and that navigation TKA is beneficial for patients with coronal bowing. We investigated the incidence of femoral/tibial bowing in patients treated with TKA and compared the radiographic parameters between the navigation group and the conventional group. Additionally, the influence of coronal bowing on these radiographic parameters was investigated.
MATERIALS AND METHODS: We enrolled 35 patients with knee osteoarthritis and 70 bilateral simultaneous TKAs. The patients underwent TKA with the use of a computer tomography-free navigation in one knee and conventional TKA in the contralateral knee. Preoperative coronal bowing were measured, and the subjects were divided into 2 subgroups, i.e. the bowing group and the non-bowing group. Lateral bowing was expressed as plus (+) and medial bowing was expressed as minus (-). Various radiographic parameters, including coronal bowing, lower leg alignment, component position, and outliers were compared between the navigation group and the conventional group.
RESULTS: Femoral bowing varied from -7.4° to 10.9° with an average of 3.0°. Tibial bowing varied from -4.1° to 4.6° with an average of 0.4°. The femoral component was placed more properly in the navigation group. Number of outlier regarding to the coronal femoral component angle to the femoral mechanical axis was 14 cases (37.8%) in the bowing group and 6 cases (18.2%) in the non-bowing group (P=0.04). DISCUSSION: In conclusion, coronal femoral bowing has an important effect on femoral bone cut in TKA. The navigated TKA was more consistent than conventional TKA in aiding proper alignments of femoral component. LEVEL OF EVIDENCE: Level II, comparative prospective study.
Copyright © 2017 Elsevier Masson SAS. All rights reserved.

Entities:  

Keywords:  Alignment; Coronal bowing; Navigation; Total knee arthroplasty (TKA)

Mesh:

Year:  2017        PMID: 28087396     DOI: 10.1016/j.otsr.2016.11.017

Source DB:  PubMed          Journal:  Orthop Traumatol Surg Res        ISSN: 1877-0568            Impact factor:   2.256


  4 in total

1.  The entry point of intramedullary tibia cutting guide should vary according to the individual tibia morphology in TKA.

Authors:  Sung-Mok Oh; Seong-Il Bin; Bum-Sik Lee; Jong-Min Kim
Journal:  Arch Orthop Trauma Surg       Date:  2019-12-16       Impact factor: 3.067

2.  The accuracy of an extramedullary femoral cutting system in total knee arthroplasty in patients with severe coronal femoral bowing: a radiographic study.

Authors:  Qianjin Wang; Xiaofeng Zhang; Tianshu Shi; Zhengyuan Bao; Bin Wang; Yao Yao; Dengxian Wu; Zheng Liu; Honggang Cai; Dongyang Chen; Jin Dai; Qing Jiang; Zhihong Xu
Journal:  J Orthop Surg Res       Date:  2022-05-07       Impact factor: 2.677

3.  The relationship between the ''Fujisawa point'' and anatomical femorotibial angle following simulated open wedge high tibial osteotomy.

Authors:  Hideo Kobayashi; Suguru Saito; Yasushi Akamatsu; Ken Kumagai; Shuntaro Nejima; Yutaka Inaba
Journal:  BMC Musculoskelet Disord       Date:  2022-08-15       Impact factor: 2.562

4.  Lateral Coronal Bowing of Femur and/or Tibia Amplifies the Varus Malalignment of Lower Limb as well as Increases Functional Disability in Patients with Knee Osteoarthritis.

Authors:  Mayur Nayak; Vijay Kumar; Rahul Yadav; Deep Narayan Srivastava; Hemant Pandit; Rajesh Malhotra
Journal:  Indian J Orthop       Date:  2020-11-22       Impact factor: 1.251