Implementation and application of moving average as continuous analytical quality control instrument demonstrated for 24 routine chemistry assays.

Huub H van Rossum1, Hans Kemperman1.   

Abstract

BACKGROUND: General application of a moving average (MA) as continuous analytical quality control (QC) for routine chemistry assays has failed due to lack of a simple method that allows optimization of MAs. A new method was applied to optimize the MA for routine chemistry and was evaluated in daily practice as continuous analytical QC instrument.
METHODS: MA procedures were optimized using an MA bias detection simulation procedure. Optimization was graphically supported by bias detection curves. Next, all optimal MA procedures that contributed to the quality assurance were run for 100 consecutive days and MA alarms generated during working hours were investigated.
RESULTS: Optimized MA procedures were applied for 24 chemistry assays. During this evaluation, 303,871 MA values and 76 MA alarms were generated. Of all alarms, 54 (71%) were generated during office hours. Of these, 41 were further investigated and were caused by ion selective electrode (ISE) failure (1), calibration failure not detected by QC due to improper QC settings (1), possible bias (significant difference with the other analyzer) (10), non-human materials analyzed (2), extreme result(s) of a single patient (2), pre-analytical error (1), no cause identified (20), and no conclusion possible (4).
CONCLUSIONS: MA was implemented in daily practice as a continuous QC instrument for 24 routine chemistry assays. In our setup when an MA alarm required follow-up, a manageable number of MA alarms was generated that resulted in valuable MA alarms. For the management of MA alarms, several applications/requirements in the MA management software will simplify the use of MA procedures.

Entities:  

Keywords:  moving average; quality assurance; quality control

Mesh:

Year:  2017        PMID: 28076303     DOI: 10.1515/cclm-2016-0696

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  4 in total

1.  Integrating moving average control procedures into the risk-based quality control plan in small-volume medical laboratories.

Authors:  Vera Lukić; Svetlana Ignjatović
Journal:  Biochem Med (Zagreb)       Date:  2022-06-15       Impact factor: 2.515

2.  Optimizing moving average control procedures for small-volume laboratories: can it be done?

Authors:  Vera Lukić; Svetlana Ignjatović
Journal:  Biochem Med (Zagreb)       Date:  2019-10-15       Impact factor: 2.313

3.  Moving average procedures as an additional tool for real-time analytical quality control: challenges and opportunities of implementation in small-volume medical laboratories.

Authors:  Vera Lukić; Svetlana Ignjatović
Journal:  Biochem Med (Zagreb)       Date:  2021-12-15       Impact factor: 2.313

4.  Comparison and optimization of various moving patient-based real-time quality control procedures for serum sodium.

Authors:  Yuanyuan Li; Qian Yu; Xiaoyan Zhang; Xiaoling Chen
Journal:  J Clin Lab Anal       Date:  2021-09-14       Impact factor: 2.352

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.