Tailored interactive sequences for continuous MR-image-guided freehand biopsies of different organs in an open system at 1.0 tesla (T) - Initial experience.

Florian Streitparth1, Christian Althoff1, Martin Jonczyk1, Felix Guettler1, Martin Maurer1, Hendrik Rathke1, Keno Moritz Sponheuer1, Bernd Hamm1, Ulf K Teichgräber1, Maximilian de Bucourt1.   

Abstract

OBJECTIVES: To assess the feasibility, image quality, and accuracy of freehand biopsies of liver, bone, muscle, vertebral disc, soft tissue, and other lesions using balanced steady-state free precession (SSFP, balanced fast field echo: bFFE), spoiled and nonspoiled gradient echo (FFE), and turbo spin echo (TSE) sequences for interactive continuous navigation in an open magnetic resonance imaging (MRI) system at 1.0 tesla (T).
METHODS: Twenty-six MR-guided biopsies (five liver, five bone, four muscle, four vertebral disc, one lung, one kidney, one suprarenal gland, and five soft or other tissue) were performed in 23 patients in a 1.0-T open magnetic resonance (MR) scanner (Panorama HFO, Philips Healthcare, Best, the Netherlands). A total of 42 samples were obtained. Depending on lesion size and location, 14-18-gauge MR-compatible biopsy sets with a length of 100 or 200 mm (Somatex Medical, Teltow, Germany), 14-18-gauge MR-compatible semiautomatic biopsy guns with a length of 100 or 150 mm (Invivo, Schwerin, Germany), or 11-gauge MR-compatible bone marrow biopsy needles with a length of 100 mm (Somatex Medical, Teltow, Germany) were employed.
RESULTS: All lesions were visible with continuous interactive imaging. Our initial results indicate that bFFE is particularly suitable for fast-moving organs (pulmonary, paracardial); moving organs are targeted better with T1-weighted (T1W) TSE, T1W FFE (liver) or T2-weighted (T2W) TSE (complicated cysts, adrenal glands), and static organs are successfully approached with proton density (PD) (spine) or T1W TSE (peripheral bones, musculoskeletal system). No adverse events related to the use of MRI were obtained. No complications occurred according to the Society of Interventional Radiology (SIR) clinical practice guidelines.
CONCLUSION: Applying tailored interactive dynamic imaging sequences for continuous navigation to liver, bone, muscle, vertebral disc, soft tissue, and other lesions can improve the feasibility, image quality, and interventional accuracy of freehand MR-guided biopsies and may hence reduce the risk of complications.

Entities:  

Keywords:  MRI; biopsies; real-time image guidance

Mesh:

Year:  2017        PMID: 28076290     DOI: 10.1515/bmt-2015-0163

Source DB:  PubMed          Journal:  Biomed Tech (Berl)        ISSN: 0013-5585            Impact factor:   1.411


  1 in total

1.  MR-guided lumbar facet radiofrequency denervation for treatment of patients with chronic low back pain in an open 1.0 Tesla MRI system.

Authors:  Georg Böning; Tony Hartwig; Patrick Freyhardt; Maximilian de Bucourt; Ulf Teichgräber; Florian Streitparth
Journal:  Ann Transl Med       Date:  2021-07
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.