Literature DB >> 28069170

A sensitive gas chromatography detector based on atmospheric pressure chemical ionization by a dielectric barrier discharge.

Ansgar T Kirk1, Torben Last2, Stefan Zimmermann2.   

Abstract

In this work, we present a novel concept for a gas chromatography detector utilizing an atmospheric pressure chemical ionization which is initialized by a dielectric barrier discharge. In general, such a detector can be simple and low-cost, while achieving extremely good limits of detection. However, it is non-selective apart from the use of chemical dopants. Here, a demonstrator manufactured entirely from fused silica capillaries and printed circuit boards is shown. It has a size of 75×60×25mm3 and utilizes only 2W of power in total. Unlike other known discharge detectors, which require high-purity helium, this detector can theoretically be operated using any gas able to form stable ion species. Here, purified air is used. With this setup, limits of detection in the low parts-per-billion range have been obtained for acetone.
Copyright © 2017 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Atmospheric pressure chemical ionization; Detector; Dielectric barrier discharge; Ion mobility; Ionization detector

Mesh:

Substances:

Year:  2016        PMID: 28069170     DOI: 10.1016/j.chroma.2016.12.071

Source DB:  PubMed          Journal:  J Chromatogr A        ISSN: 0021-9673            Impact factor:   4.759