Literature DB >> 28065638

Iterative metal artefact reduction in CT: can dedicated algorithms improve image quality after spinal instrumentation?

J Aissa1, C Thomas2, L M Sawicki2, J Caspers2, P Kröpil2, G Antoch2, J Boos2.   

Abstract

AIM: To investigate the value of dedicated computed tomography (CT) iterative metal artefact reduction (iMAR) algorithms in patients after spinal instrumentation.
MATERIALS AND METHODS: Post-surgical spinal CT images of 24 patients performed between March 2015 and July 2016 were retrospectively included. Images were reconstructed with standard weighted filtered back projection (WFBP) and with two dedicated iMAR algorithms (iMAR-Algo1, adjusted to spinal instrumentations and iMAR-Algo2, adjusted to large metallic hip implants) using a medium smooth kernel (B30f) and a sharp kernel (B70f). Frequencies of density changes were quantified to assess objective image quality. Image quality was rated subjectively by evaluating the visibility of critical anatomical structures including the central canal, the spinal cord, neural foramina, and vertebral bone.
RESULTS: Both iMAR algorithms significantly reduced artefacts from metal compared with WFBP (p<0.0001). Results of subjective image analysis showed that both iMAR algorithms led to an improvement in visualisation of soft-tissue structures (median iMAR-Algo1=3; interquartile range [IQR]:1.5-3; iMAR-Algo2=4; IQR: 3.5-4) and bone structures (iMAR-Algo1=3; IQR:3-4; iMAR-Algo2=4; IQR:4-5) compared to WFBP (soft tissue: median 2; IQR: 0.5-2 and bone structures: median 2; IQR: 1-3; p<0.0001). Compared with iMAR-Algo1, objective artefact reduction and subjective visualisation of soft-tissue and bone structures were improved with iMAR-Algo2 (p<0.0001).
CONCLUSION: Both iMAR algorithms reduced artefacts compared with WFBP, however, the iMAR algorithm with dedicated settings for large metallic implants was superior to the algorithm specifically adjusted to spinal implants.
Copyright © 2016 The Royal College of Radiologists. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2017        PMID: 28065638     DOI: 10.1016/j.crad.2016.12.006

Source DB:  PubMed          Journal:  Clin Radiol        ISSN: 0009-9260            Impact factor:   2.350


  5 in total

1.  Reduction of artifacts caused by orthopedic hardware in the spine in spectral detector CT examinations using virtual monoenergetic image reconstructions and metal-artifact-reduction algorithms.

Authors:  Nils Große Hokamp; V Neuhaus; N Abdullayev; K Laukamp; S Lennartz; A Mpotsaris; J Borggrefe
Journal:  Skeletal Radiol       Date:  2017-09-21       Impact factor: 2.199

2.  CT metal artifacts in patients with total hip replacements: for artifact reduction monoenergetic reconstructions and post-processing algorithms are both efficient but not similar.

Authors:  Kai Roman Laukamp; Simon Lennartz; Victor-Frederic Neuhaus; Nils Große Hokamp; Robert Rau; Markus Le Blanc; Nuran Abdullayev; Anastasios Mpotsaris; David Maintz; Jan Borggrefe
Journal:  Eur Radiol       Date:  2018-05-03       Impact factor: 5.315

3.  Metal artifacts in patients with large dental implants and bridges: combination of metal artifact reduction algorithms and virtual monoenergetic images provides an approach to handle even strongest artifacts.

Authors:  Kai Roman Laukamp; David Zopfs; Simon Lennartz; Lenhard Pennig; David Maintz; Jan Borggrefe; Nils Große Hokamp
Journal:  Eur Radiol       Date:  2019-01-16       Impact factor: 5.315

4.  Low-dose CT with tin filter combined with iterative metal artefact reduction for guiding lung biopsy.

Authors:  Jing Zhang; Meiling Liu; Daihong Liu; Xiaoqin Li; Meng Lin; Yong Tan; Yuesheng Luo; Xiangfei Zeng; Hong Yu; Hesong Shen; Xiaoxia Wang; Leilei Liu; Yuchuan Tan; Jiuquan Zhang
Journal:  Quant Imaging Med Surg       Date:  2022-02

5.  Reduction of CT artifacts from cardiac implantable electronic devices using a combination of virtual monoenergetic images and post-processing algorithms.

Authors:  Lenhard Pennig; David Zopfs; Roman Gertz; Johannes Bremm; Charlotte Zaeske; Nils Große Hokamp; Erkan Celik; Lukas Goertz; Marcel Langenbach; Thorsten Persigehl; Amit Gupta; Jan Borggrefe; Simon Lennartz; Kai Roman Laukamp
Journal:  Eur Radiol       Date:  2021-02-25       Impact factor: 5.315