The electrophoretic mobility and the zeta potential (ζ) of fluorescently labeled colloidal silica rods, with an aspect ratio of 3.8 and 6.1, were determined with microelectrophoresis measurements using confocal microscopy. In the case where the colloidal particles all move at the same speed parallel to the direction of the electric field, we record a xyz-stack over the whole depth of the capillary. This method is faster and more robust compared to taking xyt-series at different depths inside the capillary to obtain the parabolic flow profile, as was done in previous work from our group. In some cases, rodlike particles do not move all at the same speed in the electric field, but exhibit a velocity that depends on the angle between the long axis of the rod and the electric field. We measured the orientation-dependent velocity of individual silica rods during electrophoresis as a function of κa, where κ-1 is the double layer thickness and a is the radius of the rod associated with the diameter. Thus, we determined the anisotropic electrophoretic mobility of the silica rods with different sized double layers. The size of the double layer was tuned by suspending silica rods in different solvents at different electrolyte concentrations. We compared these results with theoretical predictions. We show that even at already relatively high κa when the Smoluchowski limiting law is assumed to be valid (κa > 10), an orientation dependent velocity was measured. Furthermore, we observed that at decreasing values of κa the anisotropy in the electrophoretic mobility of the rods increases. However, in low polar solvents with κa < 1, this trend was reversed: the anisotropy in the electrophoretic mobility of the rods decreased. We argue that this decrease is due to end effects, which was already predicted theoretically. When end effects are not taken into account, this will lead to strong underestimation of the experimentally determined zeta potential.
The electrophoretic mobility and the zeta potential (ζ) of fluorescently labeled colloidal silica rods, with an aspect ratio of 3.8 and 6.1, were determined with microelectrophoresis measurements using confocal microscopy. In the case where the colloidal particles all move at the same speed parallel to the direction of the electric field, we record a xyz-stack over the whole depth of the capillary. This method is faster and more robust compared to taking xyt-series at different depths inside the capillary to obtain the parabolic flow profile, as was done in previous work from our group. In some cases, rodlike particles do not move all at the same speed in the electric field, but exhibit a velocity that depends on the angle between the long axis of the rod and the electric field. We measured the orientation-dependent velocity of individual silica rods during electrophoresis as a function of κa, where κ-1 is the double layer thickness and a is the radius of the rod associated with the diameter. Thus, we determined the anisotropic electrophoretic mobility of the silica rods with different sized double layers. The size of the double layer was tuned by suspending silica rods in different solvents at different electrolyte concentrations. We compared these results with theoretical predictions. We show that even at already relatively high κa when the Smoluchowski limiting law is assumed to be valid (κa > 10), an orientation dependent velocity was measured. Furthermore, we observed that at decreasing values of κa the anisotropy in the electrophoretic mobility of the rods increases. However, in low polar solvents with κa < 1, this trend was reversed: the anisotropy in the electrophoretic mobility of the rods decreased. We argue that this decrease is due to end effects, which was already predicted theoretically. When end effects are not taken into account, this will lead to strong underestimation of the experimentally determined zeta potential.
Charged particle dispersions,
with particles ranging from the nano-
to micrometer range, are present in everyday life. One could think
of small DNA strands, proteins, virusses, micelles, or charged human
made colloids, such as those present in e-ink and printer toners.
Their surface potential is an important parameter to characterize
their stability and interactions. The always present van der Waals
forces are attractive if the particles are identical.[1] In the case where there are no capping molecules present
on the particle surface that can provide so-called steric stabilization,[2] most colloidal particles rely on stabilizing
repulsive forces that are caused by excess charges present on/in the
particles. Therefore, measuring electrophoretic mobilities of charged
particles is of importance for fundamental studies in a broad range
of fields,[3−10] in biophysics,[11−14] but also in more applied work in industry.[15−19]There are fewer papers in which the electrophoretic motion of rodlike
particles, the subject of the present paper, is evaluated experimentally
in solution, compared to papers for spheres, and most of the published
work is from the field of biophysics using anisotropic biomolecules
or particles. At relatively thin double layers, κa ≫ 10, where κ–1 is the double layer
thickness and a is the rod radius, it is well-known
that the electrophoretic mobility of a particle does not depend on
its shape and it does not rotate in the applied electric field.[20] For moderate κa, Henry
already predicted in 1931 that the electrophoretic mobility of infinitely
long spherocylinders is dependent on the orientation of the cylinder
with respect to the applied electric field.[21] In general, therefore, a rod moves at an angle with respect to the
field. Henry neglected the relaxation effect of the double layer,
so that the result is only valid for zeta potentials (ζ) that
are smaller than about 25 mV (ζ < kBT/e). A theory for infinitely
long cylinders, where the relaxation effect was included, was developed
by Stigter.[22,23] However, in practice, rodlike
particles such as DNA strands, fd-virus, microtubules, f-actin, or silica rods are finite and need a more accurate
description. Using numerical calculations to describe the mobility
of finite cylinders, research by both Allison et al. and Buitenhuis,
found that with smaller aspect ratio the mobility of spherocylinders
decreases at small κa, because end effects
become more pronounced.[24,25]Experimentally,
most electrophoresis measurements were done using
capillairy zone electrophoresis[13] or electrophoretic
light scattering.[25,26] Using these techniques, only
orientationally averaged electrophoretic mobilities of the rodlike
particles can be obtained. On the other hand, using optical microscopy,
one can measure the electrophoretic mobility of single rodlike particles
and its dependence on the orientation with respect to the applied
electric field. In this way the mobilities, which in general can be
decomposed into two values, a mobility in the direction parallel and
perpendicular to the rod, can be measured simultaneously.[11,12] In addition, the strong sectioning capabilities and increased resolution
of confocal microscopy, allow for the investigation of the electrophoresis
of concentrated dispersions on the single particle level.[3−5] Here we make use of the fact that with confocal microscopy it has
recently been shown that both coordinates and orientations of rodlike
silica model colloids can be obtained as well.[27] Here, we use silica rods with aspect ratios of 3.8 and
6.1 to measure the anisotropic electrophoretic motion as a function
of κa. Silica rods are well approximated by
short and rigid spherocylinders, in contrast to the long and semiflexible
microtubules and fd-virus used earlier.[11,12,25] This makes them ideal for use
as model particles.In this paper, we first describe a method
to faster measure electrophoretic
mobility profiles of colloidal particles using a confocal microscope,
as compared to methods used by our group previously.[3,4] Measuring electrophoretic mobility profiles, using optical microscopy
in closed capillaries, was previously done by measuring xyt-series at 10 different depths in the capillary, which is time-consuming.[3−5] Using cross-correlation between subsequent frames in a xyz-stack, over the whole depth of the capillary, for particles that
move all at the same speed parallel to the electric field, we were
able to significantly decrease the measurement time needed for microelectrophoresis
for rodlike particles in the Smoluchowski limit of high κa. This method is also valid for spherical particles. Second,
we describe our findings on the anisotropic electrophoretic mobility
of colloidal silica rods as a function of the double layer thickness
κa, by recording xyt-series
at the stationary plane zstat. Using confocal
microscopy and particle fitting algorithms,[27,28] we were able to measure the anistropic electrophoretic motion on
the single particle level. We dispersed the silica rods in different
(mixtures of) solvents of different dielectric constants (ϵm), which changes both the surface potential of the particles
and the double layer thickness, κ–1. To gain
additional control over the double layer thickness in that particular
solvent, we also added salt containing a large cation.We show
that, even at already relatively high κa, where
the Smoluchowski limiting law is often assumed to be valid,
an orientation dependent velocity is found. Furthermore, at decreasing
values of κa the anisotropy in mobility increases.
However, in low polar solvents with lower values of κa, the opposite trend was observed. The anisotropy in mobility
again decreases. We argue that is due to the fact that end effects
start to play a role, which was already predicted theoretically.[29]
Experimental Methods
Model
System
We used FITC labeled silica rods (ϵr ∼ 4.5, nD21 = 1.45) that were prepared according
to the method of Kuijk and co-workers.[30,31] Additionally,
SR29 rods, which were stable in cyclohexyl chloride (CHC, >98%,
Merck),
were modified with octadecyltrimethoxylsilane (OTMOS, 90%, Sigma-Aldrich).[32] Data on the particles used in this study, can
be found in Table and Figure S1. The mean dimensions of
the particles and their standard deviations were determined from transmission
electron microscopy (TEM) images made with a Philips Tecnai 10 or
12 microscope (Fei Company). Typically, 80 particles were measured
by hand using the program iTem (Soft Imaging System GmbH, version
5).
Table 1
Properties of Silica Rods Used in
This Studya
L, nm
σL, %
D, nm
σD, %
L/D
D⊥/D∥
R2
3591
18
587
10
6.1
0.76
SR29
2290
6
600
6.5
3.8
0.81
L is the end-to-end
length of the rod, D is the diameter of the particle, L/D is the aspect ratio of the particle,
and D⊥ and D∥ are the estimated translational diffusion coefficient
perpendicular and parallel to the rod’s long axis,[33] respectively, calculated from their TEM dimensions.
L is the end-to-end
length of the rod, D is the diameter of the particle, L/D is the aspect ratio of the particle,
and D⊥ and D∥ are the estimated translational diffusion coefficient
perpendicular and parallel to the rod’s long axis,[33] respectively, calculated from their TEM dimensions.For measuring zeta potentials
using cross-correlation, the particles
were suspended in a refractive index matched solvent mixture, nD21 = 1.45, of 85 wt % glycerol (≥99%, Sigma-Aldrich) and 15
wt % ultrapure water (Millipore system), with 0.55 mM lithium chloride
(LiCl). For measuring μ⊥ and μ∥ of silica rods, the R2 silica rods were dispersed in polar solvents:
either 85 wt % glycerol in ultrapure water or 78 wt % dimethyl sulfoxide
(DMSO, ≥99.9%, Sigma-Aldrich) in ultrapure water. LiCl was
added to the more polar solvents to control the thickness of the double
layer. SR29 silica rods were dispersed in the nonpolar solvent cyclohexyl
chloride (CHC, 98% Alfa Aesar GmbH & Co KG). Tetrabutylammonium
chloride (TBAC, ≥97% Sigma-Aldrich) was added to CHC as a salt
to control the double layer thickness. Properties of solvents used
in this study can be found in Table .
Table 2
Properties of Solvents Used in This
Studya
ρ, g mL–1
nD
η, mPa s
ϵm
CHC
0.993 (25 °C)[34]
1.46265 (20 °C)[35]
1.5675 (25 °C)[34]
7.6 (25 °C)[36]
DMSO
1.096 (25 °C)[37]
1.477 (25 °C)[37]
1.99 (25 °C)[37]
48.4 (25 °C)[38]
glycerol
1.258 (25 °C)[39]
1.474 (20 °C)[39]
1410 (20 °C)[39]
41.1 (20 °C)[40]
78 wt % DMSO/water
1.096* (25 °C)[37]
1.45 (22 °C)
3.542* (25 °C)[37]
56.1* (25 °C)[38]
85 wt % glycerol/water
1.219 (25 °C)[39]
1.45 (20 °C)
81 (25 °C)[41]
49.6* (20 °C)[40]
With ρ being the density
of the solution, nD being the refractive
index of the solution, η being the viscosity of the solution,
and ϵm being the dielectric constant of the medium.
Values marked with an asterisk (*) were obtained from linear interpolation
of literature values.
With ρ being the density
of the solution, nD being the refractive
index of the solution, η being the viscosity of the solution,
and ϵm being the dielectric constant of the medium.
Values marked with an asterisk (*) were obtained from linear interpolation
of literature values.
Sample Preparation
For electrophoresis measurements,
a sample cell was constructed in the following way. A rectangular
borosilicate capillary (Vitrocom, U.K.), with dimensions (x × y × z) of
either 40 mm × 1 mm × 0.1 mm or 40 mm × 2 mm ×
0.1 mm was attached to a standard microscope glass slide (Menzel Gläzer)
using adhesive tape (Scotch). Two 50 μm diameter nickel alloy
wires (Good Fellow) were bent into rectangular U-shaped electrodes
and inserted in opposite ends of the glass capillary; see Figure b. The typical spacing
of the electrodes was ∼2 cm. The wires were attached to the
glass slide using adhesive tape. The as-prepared electrophoresis cell
could only be used once.
Figure 1
(a) Schematic showing the orientation dependent
velocity of a negatively
charged spherocylinder subjected to a DC-electric field adapted from
Van den Heuvel et al.[11] (b) Sample cell
used during microelectrophoresis. (c) Schematic of flow inside the
capillary during microelectrophoresis.
(a) Schematic showing the orientation dependent
velocity of a negatively
charged spherocylinder subjected to a DC-electric field adapted from
Van den Heuvel et al.[11] (b) Sample cell
used during microelectrophoresis. (c) Schematic of flow inside the
capillary during microelectrophoresis.The electrophoresis cells were filled with dispersions of
silica
particles, using a glass Pasteur pipet (Wu Mainz). The dispersions
prepared had a volume fraction ϕ ∼ 0.01. It was checked
that no air bubbles were present between the electrodes, before sealing
the cell with UV-curable glue (Norland Optical adhesive no. 68). The
as-prepared electrophoresis cells were homogenized in a slowly rotating
stage (rotation about the long axis) for 10–60 min before measuring.
Electrophoresis Measurements
The measurements were
carried out as follows: a DC-electric field of 1.0–6.3 V mm–1 was applied between the electrodes, using a wide
band amplifier (Krohn-Hite, model 7602M). The electric field strengths
quoted are nominal, that is, based on the applied voltage and the
electrode gap. The real electric field inside the cell is typically
slightly lower due to electrode polarization and electrode reactions.
These potentials are on the order of 1 V, whereas we typically apply
a field of ∼50 V with an electrode spacing of ∼2 cm.
Hence, the difference in nominal field strength used and real electric
field strength is only on the order of a few percent. During electrophoresis
measurements, we did not observe the occurrence of bubbles near the
electrodes due to decomposition reactions such as electrolysis. We
believe this is due to the relatively low water content in the solvents
used. To record the velocity of the colloids during electrophoresis,
we used a laser scanning confocal microscope (Leica TCS SP8) equipped
with a 63× 1.4 NA oil confocal-objective (Leica). The sample
cell was placed with the long axis perpendicular to gravity. To measure
electrophoretic mobility profile parallel to gravity (Figure b), we used a Nikon C1 confocal
microscope equipped with a 40× 1.25NA oil confocal-objective
(Leica) mounted on a microscope body that was tilted 90°. Here,
the sample cell was placed with the long axis parallel to gravity,
which prevented the formation of a density gradient across the depth
(z) of the capillary due to sedimentation of particles
that disturbs the recording of mobility profiles.
Figure 2
Electrophoretic mobility
profiles of R2 rods measured in three
different ways. Particles were dispersed in 0.55 mM LiCl 85 wt % glycerol–water,
κa ≈ 24.5. (a) Measured perpendicular
to gravity, ζ = −47.0 ± 0.5 mV. (b) Measured parallel
to gravity at a field strength of 3 V mm–1, ζ
= −47.0 ± 1.6 mV. The dashed vertical lines indicate the
stationary planes. The lines indicate a parabolic-fit through the
data. We set z = 0 μm at the maximum of the
parabolic profile. Error bars, for xyt cross-correlation
and xyt particle tracking, indicate the standard
error of the measurement, which is typically smaller than the size
of a symbol. The given standard deviation on ζ is the standard
deviation on the three different estimates of ζ obtained from
the parabolic fits through the data.
Electrophoretic mobility
profiles of R2 rods measured in three
different ways. Particles were dispersed in 0.55 mM LiCl 85 wt % glycerol–water,
κa ≈ 24.5. (a) Measured perpendicular
to gravity, ζ = −47.0 ± 0.5 mV. (b) Measured parallel
to gravity at a field strength of 3 V mm–1, ζ
= −47.0 ± 1.6 mV. The dashed vertical lines indicate the
stationary planes. The lines indicate a parabolic-fit through the
data. We set z = 0 μm at the maximum of the
parabolic profile. Error bars, for xyt cross-correlation
and xyt particle tracking, indicate the standard
error of the measurement, which is typically smaller than the size
of a symbol. The given standard deviation on ζ is the standard
deviation on the three different estimates of ζ obtained from
the parabolic fits through the data.We recorded series of confocal images in two ways. Either
we recorded
time series of xy-images (∼50–100 images;
typical images size 56 × 56 μm; typical pixel size ∼0.1–0.3
μm, time between frames Δt = 0.4–1.2
s) at fixed depths; typically ∼10 depths over the total height
of 0.1 mm of the electrophoresis cell (≳10 × 40 s).[3−5] Or we recorded an xyz-stack over the whole depth
of the capillary (the average step size was Δz ∼ 1 μm and the time between frames Δt = 0.75 s, total measurement time ∼75 s). The precise positions
in z of the top and bottom walls were determined
using the reflection mode of the confocal microscope.
Electrophoretic
Mobilities
To derive average particle
velocities from confocal xyt-series, we took two
approaches. In the first approach, we obtained 2D trajectories of
particles using particle fitting and tracking algorithms[27,28] to find the average displacement and velocities between frames at
height z (xyt particle tracking).
Sometimes, the average displacement of particles between two consecutive
images was larger than the average distance between colloids, which
does not allow for reliable single particle tracking. If this happened,
we first used cross-correlation between subsequent frames[42] to estimate the average displacement, in integer
number of pixels, parallel to the electric field (x-direction). After that, the average displacement in x was subtracted from the particle coordinates found. In this “co-moving”
frame of reference, the particles could then be tracked. To obtain
the true particle trajectories, the average displacement was added
back to the x-coordinate after particle tracking.
The second approach was to use the maximum in the cross-correlation
between two consecutive frames to find the average particle displacement
and velocities with subpixel accuracy at height z(43) (xyt cross-correlation).
Next, to derive the average particle velocities from a single xyz-stack of the whole depth of the capillary, we used cross-correlation;
cross-correlation of subsequent frames at height z provides the average particle displacement (xyz cross-correlation). Due to the finite time interval between frames
in the stack, particles will appear at shifted positions in subsequent
frames.The electrophoretic mobility of the particles is related
to the velocity of the particles, and given bywhere v is the mean particle
velocity (m s–1) and E is the electric
field strength (V m–1). During electrophoresis measurements
inside a closed capillary, a parabolic Poiseuille flow sets up in
the sample cell, due to electro-osmosis of the fluid;[44] see Figure c. Thus, the motion of the particles in the sample cell is a result
of the superposition of the electrophoretic velocity and the flow
profile setup the by electro-osmotic flow (EOF). At the stationary
planes, z = zstat, the
net flow of the fluid is zero, and the particles solely translate
because of their electrophoretic mobility μ = μ(zstat) and the superimposed
Brownian motion of the particles. In a rectangular cell, the stationary
planes (zstat) are located at[44]where k is the
ratio of the
minor and major cross-section of the capillary, h is the distance from the center of the cell (z =
0) and either of the walls. We fitted the obtained mobility profiles
to a parabolic function, and set z = 0 at the maximum
of the parabolic profile and assumed this to be the center of the
capillary. From the value of the fit at zstat, we derived the true electrophoretic mobility. For our capillaries,
we measured a depth of 2h ≈ 110 μm and zstat is located at (eq ) zstat ≈
± 32.7 μm (k ≈ 20) ∨ ±
33.7 μm (k ≈ 10).In order to
determine the anisotropy in mobility (μ⊥/μ∥) of our silica rods, we generally recorded
only at zstat image series of typically
∼50–100 frames. An exception was the samples of R2 silica
rods dispersed in 85 wt % glycerol in water; in this case, we took xyt-series at 10 different depths of the sample. The typical
images size was 95 × 95 μm; pixel size 0.37 μm; the
time interval between subsequent frames Δt =
0.2–0.9 s.
Solvent Conductivity
An estimate
for the Debye screening
length, κ–1, was obtained from the measured
conductivities of the solvents used. The conductivities of the polar
solvents were measured with a CDM 230 conductivity meter (Radiometer
Analytical). The conductivity of CHC, an apolar solvent, was measured
using a Scientifica 627 conductivity meter. We assumed that ions contributing
to the conductivity in CHC were H+ and Cl– and independent migrations of ions. The conductivity is related
to the concentration of dissociated salt, given by.σ is the measured conductivity (Sm–1), Λ0 is the limiting molar conductance
of the salt at infinite solution (Sm2 mol–1) in the given solvent, and c is the concentration
of dissociated salt.Using Walden’s rule, one can calculate
Λ0 for different solvents, using literature values
of known species. Walden’s rule state that the product of the
limiting conductance and the viscosity is a constant for the same
species in different solvents.The superscript denotes the particular solvent
and η is the viscosity (Pa s) of the solvent. In order to calculate
Λ0 for CHC, we used known literature values for HCl
in ethanol, Λ0EtOH = 84.3 S cm2 mol–1,[45] ηEtOH = 1.08 mPa s.[45] For polar solvents, we assumed that ions contributing
to the conductivity were Li+ and Cl–.
In order to estimate Λ0 in polar solvents, we used
literature values for LiCl in glycerol: Λ0glycerol = 27.1 × 102 S cm2 mol–1,[46] ηglycerol = 910 mPa s (25 °C).[41] Literature values for viscosities of the different
solvent(-mixtures) used can be found in Table . Using the relation between the ionic strength
and conductivity, we can calculate κ–1 using
the following relation, valid for monovalent ions:with NA being
Avogadro constant and c being the concentration of
dissociated monovalent salt in the solvent.For the polar solvents,
either we used solvents with added LiCl
or we used solvents without added salt. In the 85 wt % glycerol–water
mixture, without any salt added, the ionic strength was controlled
by the trace ions that are always present. For this mixture, we measured
a conductivity of 0.168 μS cm–1. Although
we do not know the composition of the trace-ions, we used the limiting
conductivity for LiCl in glycerol to estimate the screening length,
which we estimated to be κ–1 ∼ 32 nm.
When 0.55 mM LiCl was added to the 85 wt % glycerol–water mixture,
all the trace ions in the solution were masked by this salt. For this
mixture, we measured a conductivity of 1.13 μS cm–1, with a corresponding screening length of κ–1 ∼ 12 nm. For the 78 wt % DMSO–water mixture, without
any salt added, we measured a conductivity of 0.260 μS cm–1, with a corresponding screening length κ–1 ∼ 132 nm. When 0.067 mM LiCL was added to
the solvent, we measured a conductivity of 1.6 μS cm–1, with a corresponding screening length of κ–1 ∼ 53 nm. For silica rods dispersed in CHC, we aimed for large
screening lengths. Therefore, trace ions were removed as much as possible
from the solution.[47] Activated alumina
was added to a vial of CHC that was left on a roller-bench overnight.
The cleaned CHC was then separated from the alumina by centrifugation.
Molecular sieves, with a pore size of 4 Å, were added to the
cleaned CHC. The measured conductivity of the cleaned CHC was 10pS
cm–1, which corresponds to a screening length of
κ–1 ∼ 7 μm. Moreover, to decrease
the double layer in CHC, also a saturated solution of TBAC in CHC
was made. This was done by adding TBAC to cleaned CHC (up to a concentration
of 260 mM),[48] and left to equilibrate for
1 week. We then diluted the solvents to 0.26 μM and 0.026 μM.
The measured conductivity of ∼0.026 μM TBAC in CHC was
108 pS cm–1, which corresponds to a screening length
κ–1 ∼ 2.2 μm. The measured conductivity
of ∼0.26 μM TBAC in CHC was 4800 pS cm–1, which corresponds to a screening length κ–1 ∼ 0.33 μm.
Theoretical Framework
The relation between the zeta potential ζ and electrophoretic
mobility μ is given byHere ϵm is the dielectric
constant of the solvent, ϵ0 is the permittivity of
vacuum, η is the viscosity of the medium (Pa s), and f(κa,ζ) is a correction factor
that also takes the relaxation effect into account.[21−23] In the Smoluchowski
limit (κa ≫ 1), f(κa,ζ) approaches to 3/2 and this relation is then valid
for particles of any shape.[20]For
(infinitely) long rods oriented parallel to the electric field
(κL/2 ≫ 1) and ignoring relaxation effects
(ζ < kBT/e), the relation between ζ and μ∥ is just the Smoluchowski limiting law:[21]For
infinitely long rods oriented perpendicular
to the electric field, the electrophoretic mobility is given by[21−23]Approximations of f⊥(κa, ζ) are depicted in Figure ,[49,50] and f⊥(κa) approaches 1 at κa ≫ 1 and 1/2 at
κa ≪ 1.
Figure 5
Anisotropy in mobility
of silica rods (μ⊥/μ∥) plotted as a function of κa. The black solid
line depicts Ohshima’s analytical
expression of Henry’s solution, for infinitely long rods and
low zeta potential ζ < kBT/e.[49] The dashed,
dotted, and dashed-dotted lines depicts Ohshima’s approximate
analytical expression for infinitely long rods and moderate zeta potentials,[50] taking the relaxation effect into account, for
|ζ| ≈ 25 mV, |ζ| ≈ 50 mV, and |ζ|
≈ 75 mV, respectively. The symbols are experimental data points
of the measured anisotropy in mobility. Squares (green) are SR29 rods
in CHC, from left to right: without TBAC added, with ∼0.026
μM, and with ∼0.26 μM TBAC. Circles (purple) are
R2 rods in polar solvents (DMSO–water and glycerol–water),
from left to right: 78 wt % DMSO–water, 78 wt % DMSO–water
with LiCl, 85 wt % glycerol–water, and 85 wt % glycerol–water
with LiCl. The error bars indicate the estimated standard deviations.
Hence, for κa < 102, the electrophoretic
velocity of a cylinder is orientation dependent, such that the rod
moves at an angle with respect to the electric field. The experimentally
determined anisotropy in mobility can be defined asMeasuring the electrophoretic
mobility using
scattering techniques, one obtains the orientationally averaged mobility:[23,51]By recording the trajectories and corresponding
orientations of rods in an electric field using confocal microscopy,
we can obtain μ∥ and μ⊥, separately. We determine the particle velocity parallel v and perpendicular v to the applied electric field. Figure a shows the frame
of reference; here the subscripts ∥ and ⊥ are taken
with respect to the rod’s long axis. The orientation dependent
velocities of the rods in the x and y direction are related to μ∥ and μ⊥,[11] given bywhere μEOF(z) is the mobility of the particles due to the
electro-osmotic flow
inside the sample.As μ⊥ can only be
derived from a fit of
the data to eq , in
practice we derive both (μ∥ – μ⊥) and μ⊥ from eq . Using the effective mobility eq , we determine the zeta
potential ζ of the rods
Results and Discussion
Figure shows the
electrophoretic mobility profiles obtained for R2 rods dispersed in
a 85 wt % glycerol–water mixture with 0.55 mM LiCl. In this
solvent, κa ≫ 10 and we assume the Smoluchowski
limiting law is valid. Hence, the rods move all at the same speed
parallel to the electric field. We measured the electrophoretic mobility
profiles perpendicular to gravity and compared the different methods.
First, the electrophoretic velocity of the particles was recorded
at 10 different depths inside the capillary by recording confocal xyt-series. The mean particle mobility at height z was determined either by determining particle trajectories
(xyt particle tracking) or by measuring the mean
displacement using cross-correlation between subsequent frames (xyt cross-correlation). Next, a confocal xyz-stack was recorded over the whole depth of the capillary. The mean
displacement of particles between subsequent frames in z was determined by using cross-correlation between subsequent images
in the xyz-stack. Hence, a velocity profile of the
particles as a function of z inside the sample cell
could be obtained from one recorded measurement volume (xyz cross-correlation). In all three experiments, there were around
60 rods present in every recorded frame. All three methods gave similar
electrophoretic mobility profiles and show that the mobilities did
not depend on the electric field strength used. This was also the
case for the fast xyz cross-correlation method.When the viscosity η of the solvent in which the particles
are dispersed is low and the density of the particles does not match
the density of the solvent (ρcolloid ≠ ρsolvent), sedimentation can occur during the time of the measurement
(or creaming for particles with a density lighter than that of the
solvent, which is not the case here). For silica particles > 300
nm,
that is a concern. Although the sedimentation velocity is small compared
to the electrophoretic velocity, accumulation of particles near the
bottom can disturb the measured mobility profiles. To prevent significant
sedimentation during measurements, the sample cell can be placed with
the long axis parallel to gravity and the electrophoretic mobility
profile is then measured parallel to gravity.[5]Figure b shows the
electrophoretic mobility profiles, measured parallel to gravity, obtained
by the same methods as described above. We corrected for the effect
of gravity on the apparent mobility of the particles, by taking the
average of the two mobility profiles with different field direction,
canceling out any effect of gravity. For R2 rods using the Smoluchowski
limiting law we found ζ = −47.0 ± 1.6 mV. This is
the same value within the error as was found when the electrophoretic
mobility was measured perpendicular to gravity: ζ = −47.0
± 0.5 mV (Figure a). We thus found similar results for electrophoretic mobility profiles
measured parallel to gravity and mobility profiles measured perpendicular
to gravity. Measurement on silica spheres also showed that the three
different measurement methods give similar results.[5]To compare the robustness of the different measurement
methods,
we calculated the 95% confidence intervals for the estimated electric
mobility from the fit of the different measurement methods using Matlab
(version R2016a). We calculated these intervals from data in Figure and from van der
Linden et al.;[5] see Table S1. This interval is on average smaller for the xyz cross-correlation method, compared to the xyt particle tracking and xyt cross-correlation method.
Thus, the determination of the mobility profile using the xyz cross-correlation method is the most robust.To
summarize the results thus far, the three different methods
to measure electrophoretic mobilities gave similar results. We also
found similar results for the electrophoretic mobility profiles measured
parallel to gravity and the profiles measured perpendicular to gravity.
The fit of the mobility profile for xyz cross-correlation
is the more robust, compared to the two other measurement methods.
Moreover, the xyz cross-correlation is much faster
(∼75 s) than our earlier measurement technique xyt particle tracking (≳10 × 40 s). Because xyz cross-correlation is much faster, sedimentation during the measurement
is less of a problem. But also the Segre and Silberberg effect or
flow focusing[52] will affect the measurements
less. The Segre and Silberberg effect describes that, during laminar
Poiseuille flow in a capillary, particles tend to position themselves
at 0.6R from the center of the capillary. We observed
this unfavorable flow focusing effect during electrophoresis experiments
after measuring for ≳3 min. Hence, with xyz cross-correlation one can measure the electrophoretic mobility profile
before this effect sets in.The xyz cross-correlation
method of measuring
electrophoretic mobility profiles can be applied to both colloidal
rods and spheres. However, for rods this is only possible in the limit
of high κa ≫ 10, since the method measures
only the mobility component parallel to the field (μ). Because at lower κa rods
have an anisotropic mobility that depends on the angle the rods have
with respect to the applied electric field (as described above in
the Theoretical Framework). In principle, xyz cross-correlation can still be used at low κa by using eq and assuming infinitely long rods, one can obtain ζ from the
measured μeff.[22,50] However, measuring
both μ⊥ and μ∥ to
obtain ζ should be more accurate. The electrophoretic mobility
of the R2 rods was measured at κa ≈
24.5, for which we measured an anisotropy in mobility of μ⊥/μ∥ = 0.98 (see below). Thus,
the rods moved almost parallel to the electric field, and we assumed
that the small anisotropy in mobility did not significantly affect
the measured ζ. Finally, we wish to remark that the cross-correlation
method works best if (long-range) order in the sample is absent, otherwise
one will find several peaks in the cross-correlation matrix, making
it difficult to determine the real displacement of the sample.
Anisotropic
Electrophoretic Mobilities of Silica Rods
As observed earlier
for microtubules,[11] we observed that the
electrophoretic velocity and direction of electrophoretic
motion of the colloidal silica rods was in some cases dependent on
the orientation of the rods with respect to the electric field; see Figure a. This is in line
with eqs and 12. A diagram of the electric field components acting
on a silica rod that is oriented at an angle θ with respect
to the electric field is shown in Figure a. The electric field is along the x-axis to the right. The resulting velocity is not parallel
to the electric field, but points toward the long axis of the rod,
as in general μ⊥ < μ∥.Figure a
shows a sequence of cut outs of confocal images of the electrophoretic
motion of R2 silica rods dispersed in a 78 wt % DMSO in water mixture,
without any salt added, κa = 2.2, taken at
height z = zstat. In
these frames, the rods that are highlighted have different orientations
with respect to the electric field. The movement of the rods is not
exactly parallel to the electric field, but the velocity has a component
in the direction of the long axis of the silica rod; see Supporting Movie 1.
Figure 3
(a) Confocal images from
an xyt-series of the
electrophoresis of R2 silica rods in DMSO-–ater κa = 2.2. Silica rods show an orientation dependent velocity
with respect to the applied electric field. (b) Overlay of frames
from an xyt-series of the electrophoresis of a silica
rod dispersed in CHC κa = 0.04, 30 frames, t = 11 s, Δt = 0.37 s, E = 2 V mm–1. At t = 0, the color
of the rod is blue. The rods still exhibit rotational diffusion, due
to Brownian motion. (c) Several trajectories of SR29 rods dispersed
in CHC, κa = 0.04, during electrophoresis.
Position of the center of mass is plotted. Minimum length of trajectory
displayed here is 20 frames.
(a) Confocal images from
an xyt-series of the
electrophoresis of R2 silica rods in DMSO-–ater κa = 2.2. Silica rods show an orientation dependent velocity
with respect to the applied electric field. (b) Overlay of frames
from an xyt-series of the electrophoresis of a silica
rod dispersed in CHC κa = 0.04, 30 frames, t = 11 s, Δt = 0.37 s, E = 2 V mm–1. At t = 0, the color
of the rod is blue. The rods still exhibit rotational diffusion, due
to Brownian motion. (c) Several trajectories of SR29 rods dispersed
in CHC, κa = 0.04, during electrophoresis.
Position of the center of mass is plotted. Minimum length of trajectory
displayed here is 20 frames.Figure b
shows
an overlay of subsequent confocal images taken of the electrophoretic
motion of SR29 silica rods in CHC, κa = 0.04,
taken at height z = zstat. The time interval between the images was Δt = 0.37 s. In total, 30 frames were used to make the overlay. At t = 0 s, the color of the rod is blue. The rod moved in
a more or less straight line in the direction opposite to the electric
field direction, and due to rotational Brownian motion its orientation
fluctuated; see Supporting Movie 3.The orientation dependent velocity of silica rods, dispersed in
78 wt % DMSO in water, κa = 2.2, during electrophoresis
can be clearly observed by eye. However, for rods dispersed in pure
CHC, κa = 0.04, it is harder to observe the
orientation dependent velocity by eye, because the anisotropy in mobility
is smaller.In order to quantify the exact electrophoretic mobilities
of rods
orientated parallel and perpendicular to the electric field, we used
particle tracking. In each frame we determined the orientations of
the rods and recorded their displacements since the previous frame.
The displacements were converted to mobilities parallel and perpendicular
to the electric field, and were averaged in bins of 4° in θ. Figure a shows the result for the measured orientation dependent
mobility of the R2 silica rods dispersed in 78 wt % DMSO–water,
κa = 2.2, at E = 1.14 V mm–1. Note that the mean square angular displacement,
due to the rotational Brownian motion, of the rods in the interval
between frames (0.374 s) is about 12°. The electric field was
pointing in the positive x-direction. The green symbols
indicate the mobility of the rods in the y-direction
μ, which is perpendicular to the
applied electric field. The red symbols indicate the mobility of the
rods in the x-direction parallel to the electric
field. When the rod was oriented perpendicular or parallel to the
electric field (θ = −90° V 90° or θ =
0°), only a mobility component μ was observed, parallel to the electric field. For rods that
were not completely oriented parallel or perpendicular to the electric
field (−90° < θ < 0° or 0° <
θ < 90°), a mobility component perpendicular to the
electric field μ was found, which
was either positive or negative. In other words the rods dispersed
in a 78 wt % DMSO–water mixture showed an orientation dependent
velocity, and depending on the orientation, they also had a velocity
component in the positive or negative y-direction.
The solid lines in Figure a are a fit of eqs and 12 through the unbinned data. From
the fitted amplitude (μ∥ – μ⊥) and offset (μ⊥) of eq , we determined μ⊥, μ∥, and the anisotropy in
mobility μ⊥/μ∥. In
this case, the measured anisotropy was μ⊥/μ∥ = 0.664 ± 0.006. The zeta potential was determined
using eqs , 10, and 13. In all cases, we
used eq to obtain
(μ∥ – μ⊥),
and values obtained from eq were 0–20% off from the value found by eq . The difference may arise because
the fitted values for (μ∥ – μ⊥) and μ⊥ in eq are dependent on each other, whereas
μ⊥ is not of any influence in eq to determine (μ∥ – μ⊥). Figure b shows the corresponding histogram of displacements
of R2 rods in the x and y direction,
respectively parallel or perpendicular to the applied electric field.
Because the rods behave Brownian, their orientation fluctuates and
hence we find a distribution of displacements (see eqs and 12).
Moreover, the main displacement due to the electrophoresis is parallel
to electric field, and hence the histogram of displacement in the x-direction is centered around a negative displacement.
Figure 4
Electrophoresis
measurement of R2 silica rods dispersed in DMSO–water,
without salt added, κa = 2.2. (a) Orientation dependent mobility
in x-direction (red symbols), parallel to electric
field, and y-direction (green symbols), perpendicular
to applied electric field. Raw data is binned in 4° wide bins.
The error bars indicate the standard error on the binned data points.
The solid lines are a fit through the raw data using eq (μ) and eq (μ). (b) Histogram of displacements of rods
in x and y direction. An anisotropy
of μ⊥/μ∥ = 0.664 ±
0.006 was found and ζ = −75 mV. The direction of the
electric field was in the positive x-direction, E = 1.14 V mm–1. Δt = 0.374 s. The error on μ⊥/μ∥ is the estimated standard error obtained from the covariance matrix
of the fitted parameters.
Electrophoresis
measurement of R2 silica rods dispersed in DMSO–water,
without salt added, κa = 2.2. (a) Orientation dependent mobility
in x-direction (red symbols), parallel to electric
field, and y-direction (green symbols), perpendicular
to applied electric field. Raw data is binned in 4° wide bins.
The error bars indicate the standard error on the binned data points.
The solid lines are a fit through the raw data using eq (μ) and eq (μ). (b) Histogram of displacements of rods
in x and y direction. An anisotropy
of μ⊥/μ∥ = 0.664 ±
0.006 was found and ζ = −75 mV. The direction of the
electric field was in the positive x-direction, E = 1.14 V mm–1. Δt = 0.374 s. The error on μ⊥/μ∥ is the estimated standard error obtained from the covariance matrix
of the fitted parameters.We measured the orientation-dependent electrophoretic mobilities
for silica rods over a range of κa. We tuned
κa by dispersing the rods in solvents with
different dielectric constants, and by the addition of salt to change
the concentration of free ions in solutions. Although using different
solvents affects ζ, the value of μ⊥/μ∥ is much less affected. The details of the results
on the measured orientation dependent mobilities at different κa can be found in the Supporting Information (Figures S2–S7). An overview on these results is presented
in Figure . The graph shows the anisotropy in mobility of the
two types of silica rods μ⊥/μ∥ as a function of κa. The symbols in the graph are experimental
data points of the measured μ⊥/μ∥ for the SR29 rods in CHC (green squares) and for the
R2 rods in polar solvents (purple circles). The black line depicts
Ohshima’s analytical expression, for Henry’s solution
of infinitely long rods and low zeta potential ζ < kBT/e.[49] The dashed, dotted, and dashed-dotted lines
depict Ohshima’s approximate analytical expression for infinitely
long rods and at moderate zeta potentials,[50] taking the relaxation effect into account.Anisotropy in mobility
of silica rods (μ⊥/μ∥) plotted as a function of κa. The black solid
line depicts Ohshima’s analytical
expression of Henry’s solution, for infinitely long rods and
low zeta potential ζ < kBT/e.[49] The dashed,
dotted, and dashed-dotted lines depicts Ohshima’s approximate
analytical expression for infinitely long rods and moderate zeta potentials,[50] taking the relaxation effect into account, for
|ζ| ≈ 25 mV, |ζ| ≈ 50 mV, and |ζ|
≈ 75 mV, respectively. The symbols are experimental data points
of the measured anisotropy in mobility. Squares (green) are SR29 rods
in CHC, from left to right: without TBAC added, with ∼0.026
μM, and with ∼0.26 μM TBAC. Circles (purple) are
R2 rods in polar solvents (DMSO–water and glycerol–water),
from left to right: 78 wt % DMSO–water, 78 wt % DMSO–water
with LiCl, 85 wt % glycerol–water, and 85 wt % glycerol–water
with LiCl. The error bars indicate the estimated standard deviations.For relatively small double layers
(R2 rods, purple spheres), when
κa > 2, our experimental data show a similar
trend as theoretical predictions for infinitely long rods.[21,50] For SR29 silica rods in apolar solvents with relatively large double
layers (green squares), the experimental data do not agree with theoretical
predictions for infinitely long cylinders. Rather, μ⊥/μ∥ increases as κa decreases, and the orientation dependent velocity during electrophoresis
becomes less pronounced. The difference arises because end effects
start to play a role.[24,29] In this case, the Smoluchowski
equation does not hold anymore for rods orientated parallel to an
electric field. Calculations on short DNA strands (κa ∼ 1) orientated parallel to the electric field,
showed that the mobility is significantly reduced compared to longer
DNA strands,[24] while the mobility of rods
perpendicular to the electric field remains roughly the same (compared
to infinitely long rods). For theoretical calculations on fd-virus, a similar trend was observed when the aspect ratio
decreased at values of κa = 3.56 × 10–4.[25]Because the Smoluchowski
limiting law is not valid anymore at κa <
2 for short rods oriented parallel to the electric
field, this would lead to an underestimation of ζ if we used eq . Instead, for μ∥, one should expect a dependence on length, but it
is not known analytically. A first estimate is to replace μ∥ by a function with the same κa dependence as that for a rod oriented perpendicular to the field,
or f⊥(κL/2). In other words, we expect μ⊥/μ∥ = f⊥(κa)/f⊥(κL/2). This is shown in Figure for aspect ratios L/D =
6.1 and L/D = 3.8 by using Ohshima’s
approximate analytical expression for infinitely long rods and moderate
zeta potentials.[50] It shows indeed an expected
increase of μ⊥/μ∥ at
κa ≲ 3. This captures the trend in the
experimental data for SR29 rods dispersed in CHC reasonably.
Figure 6
Anisotropy
in electrophoretic mobility μ⊥/μ∥ plotted as a function of κa. Experimental data from Figure are plotted together with a modified expression
for “finite” rods and moderate zeta potentials. This
was done by assuming that rods oriented parallel to the electric field
are subjected to the same retardation function as rods oriented perpendicular
to the electric field: μ⊥/μ∥ = f⊥(κa)/f⊥(κL/2). This assumption predicts that for finite rod sizes μ⊥/μ∥ increases as the size of
the double layer increases (κa < 1).
Anisotropy
in electrophoretic mobility μ⊥/μ∥ plotted as a function of κa. Experimental data from Figure are plotted together with a modified expression
for “finite” rods and moderate zeta potentials. This
was done by assuming that rods oriented parallel to the electric field
are subjected to the same retardation function as rods oriented perpendicular
to the electric field: μ⊥/μ∥ = f⊥(κa)/f⊥(κL/2). This assumption predicts that for finite rod sizes μ⊥/μ∥ increases as the size of
the double layer increases (κa < 1).For κa >
1, we find larger values of μ⊥/μ∥ than the predictions for
infinitely long rods. Although our rods have finite length it may
perhaps also partly be attributed to a projection bias. We measured
the projected angles of the rods onto the 2D image plane. This is
mainly of influence for rods that appear to have an orientation parallel
to the electric the field, as their actual angle with the electric
field may be larger. Hence, when these rods would be completely parallel
to the electric field, their mobility would be larger. Thus, the actual
value of μ⊥/μ∥ is
probably smaller than the value we measured.The anisotropy
in mobility in this study was in most cases determined
by recording confocal xyt-series only at a height z = zstat, because the rods
sediment to the bottom during the electrophoresis measurements before
the whole parabolic electro-osmotic flow profile can be measured.
SR29 rods dispersed in CHC sediment 100 μm in 7 min, the depth
of the capillary. R2 rods dispersed in DMSO–water sediment
100 μm in 10 min. The drawback of measuring only at zstat is that zstat is located in a steep part of the parabolic electro-osmotic flow
profile. A small error in the position of zstat already causes a relatively large error in the determination of
μ⊥. Only when the rods were dispersed in a
viscous 85 wt % glycerol–water mixture, the rods did not sediment
appreciably during the time scale of the measurements. In this solvent
mixture rods take 6 h to sediment over 100 μm. In this case,
we recorded confocal xyt-series at 10 different depths
inside the capillary; through this determination of the electro-osmotic
flow profile, we could more accurately determine μ⊥. To overcome the problem of sedimentation and measure the whole
flow profile, a confocal microscope with a tilted frame can be used,[5] such that the electric field is parallel to gravity.
However, due to the relatively long measurement time needed to obtain
μ⊥/μ∥, we encountered
other difficulties. At a relatively high volume fraction (ϕ
≈ 0.04) of rods swirls occurred in the sample, and nonuniform
settling was observed.[53,54] However, when decreasing the
volume fraction of rods flow focusing[52] occurred in the sample within the time scale of the measurement,
as described above. Therefore, we determined the anisotropy in mobility
at height z = zstat for
rods dispersed in CHC or DMSO–water.The visual appearance
of the motion of silica rods during electrophoresis
was dependent on the solvent the silica rods were dispersed in. For
instance, in a pure 78 wt % DMSO–water mixture (κa = 2.2) a large anisotropy in mobility of μ⊥/μ∥ ∼ 0.664 was measured for the R2
rods. Hence, the rods showed a clear orientation dependent velocity
during electrophoresis. When we look at Supporting Movie 1 of the electrophoretic motion of these silica rods
in 78 wt % DMSO–water, we observed that the silica rods created
the impression of “swimming”. The direction of motion
of the silica rods was approximately in the direction of their long
axis. Because the silica rods still behaved Brownian, the angle of
each rod with respect to the applied electric field fluctuated, so
that the mobility direction changed on a time scale of seconds. The
relatively low viscosity (3.54 mPa s) of the solvent mixture used
in this case results in a calculated rotational diffusion coefficient Dr = 0.032 rad2 s–1. During 1 s, the rods are expected to rotate = 0.35
rad (20°). The combination
of Brownian motion and an orientation dependent electrophoretic motion
led to an apparent swimming behavior of these silica rods dispersed
in DMSO when a DC-electric field was applied to the sample. On the
contrary, a sample of silica rods in 85 wt % glycerol–water
showed a completely different behavior (Supporting Movie 2). The silica rods were pulled in a more of less straight
line toward the positive electrode, without much rotational diffusion
of the rods. This is partly due to the relatively high viscosity of
the medium (81 mPa s) that makes the rotational diffusion of the rods
rather slow Dr = 0.0014 rad2 s–1. Furthermore, the measured anisotropy in mobility
μ⊥/μ∥ ∼ 0.97
did not give rise to a clear orientation dependent motion. Supporting Movie 3 then again shows that silica
rods in CHC translated at an angle during electrophoresis, without
too much “swimming” motion observed because of fast
rotational diffusion. In CHC, the SR29 silica rods had a measured
anisotropy in mobility μ⊥/μ∥ ∼ 0.89. The much lower viscosity (1.57 mPa s) of CHC, and
the slightly smaller dimension of the SR29 silica rods, compared to
R2 rods, results in a higher Dr = 0.195
rad2 s–1 of the rods. Hence, rotational
diffusion of the rods was clearly visible on top of the electrophoretic
motion the silica rods.In summary, differences in the observed
motions of the silica rods
during electrophoresis were due to differences in viscosity and the
κ–1 of the particles due to the solvent the
particles were dispersed in. These parameters make for an intriguing
interplay between rotational diffusion (Brownian motion) and the orientation
dependent electrophoretic mobility.
Conclusions
We
showed that measuring the electrophoretic mobility, as well
as the electro-osmotic flow profile, of colloidal particles that all
move at the same speed parallel to the direction of the electric field,
can be done by recording an xyz-stack over the whole
depth of a capillary using confocal microscopy. Recording a xyz-stack is much faster compared to the method that was
used previously in which xyt-series were recorded
at 10 different depths inside the capillary. In the case of rods that
did not move all at the same speed through an electric field, we were
able to measure the orientation dependent velocity of silica rods
during electrophoresis. From this we could determine the anisotropy
in the mobility of rods, with an aspect ratio around 5, during electrophoresis
and compare this with theoretical values from literature. We showed
that even at relatively high values of κa,
where the Smoluchowski limiting law was assumed to be valid, a small
anisotropy in mobility was found. Furthermore, for the R2 rods dispersed
in polar solvent our data agree qualitatively with the theoretical
prediction for infinitely long cylinders and low zeta potential (κa ≳ 2). However, this theory starts to break down
at small values of κa, where κa ≲ 1. At these values, end effects start to play
a role. Also, the value of the measured anisotropy in mobility was
observed to decrease again at smaller values of κa. This was already predicted in theoretical calculations for finite
cylinders.[24,25] If this end effect is not taken
into account while calculating zeta potentials, this will lead to
an underestimation of the experimentally established zeta potential
by up to a factor two.
Authors: Gerwin H Gelinck; H Edzer A Huitema; Erik van Veenendaal; Eugenio Cantatore; Laurens Schrijnemakers; Jan B P H van der Putten; Tom C T Geuns; Monique Beenhakkers; Jacobus B Giesbers; Bart-Hendrik Huisman; Eduard J Meijer; Estrella Mena Benito; Fred J Touwslager; Albert W Marsman; Bas J E van Rens; Dago M de Leeuw Journal: Nat Mater Date: 2004-01-25 Impact factor: 43.841
Authors: Marjolein N van der Linden; Johan C P Stiefelhagen; Gülşen Heessels-Gürboğa; Jessi E S van der Hoeven; Nina A Elbers; Marjolein Dijkstra; Alfons van Blaaderen Journal: Langmuir Date: 2014-12-23 Impact factor: 3.882
Authors: Vincenzo Calabrese; Davide Califano; Marcelo A da Silva; Julien Schmitt; Saffron J Bryant; Kazi M Zakir Hossain; Ana M Percebom; Aurora Pérez Gramatges; Janet L Scott; Karen J Edler Journal: ACS Appl Polym Mater Date: 2020-02-12