Literature DB >> 28041783

Thermal Resistance Anastomosis Device for the Percutaneous Creation of Arteriovenous Fistulae for Hemodialysis.

Jeffrey E Hull1, Guillermo Elizondo-Riojas2, Wendy Bishop3, Yesenia L Voneida-Reyna2.   

Abstract

PURPOSE: To evaluate the safety and efficacy of arteriovenous fistula (AVF) creation with a thermal resistance anastomosis device (TRAD).
MATERIALS AND METHODS: From January 2014 to March 2015, 26 patients underwent ultrasound (US)-guided percutaneous creation of proximal radial artery-to-perforating vein AVFs with a TRAD that uses heat and pressure to create a fused anastomosis. Primary endpoints were fistula creation, patent fistula by Doppler US, two-needle dialysis at the prescribed rate, and device-related complications.
RESULTS: Technical success rate of fistula creation was 88% (23 of 26). Procedure time averaged 18.4 minutes (range, 5-34 min), and 96% of anastomoses (22 of 23) were fused. At 6 weeks, 87% of AVFs (20 of 23) were patent, 61% (14 of 23) had 400-mL/min brachial artery flow, 1 patient was receiving dialysis, 2 fistulae had thrombosed, and 1 patient had died unrelated to the procedure. Eighty percent (16 of 20), 70% (14 of 20), and 60% (12 of 20) of patients were receiving dialysis at 3, 6, and 12 months; 4 patients died, 3 fistulae failed, and one patient was lost to follow-up. Overall, 87% of AVFs (20 of 23) had an additional procedure at a mean of 56 days (range, 0-239 d), including balloon dilation in 43% (n = 10), brachial vein embolization in 26% (n = 6), basilic vein ligation in 17% (n = 4), venous transposition in 30% (n = 7), and valvulotomy in 4% (n = 1). There were no major complications related to the device.
CONCLUSIONS: Percutaneous AVFs created with a TRAD met the safety endpoints of this study. Midterm follow-up demonstrated intact anastomoses and fistulae suitable for dialysis.
Copyright © 2016 SIR. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2016        PMID: 28041783     DOI: 10.1016/j.jvir.2016.10.033

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  8 in total

Review 1.  Devices and Techniques for Percutaneous Creation of Dialysis Arteriovenous Fistulas.

Authors:  Ahmed Kamel Abdel Aal; Xavier Jefferson; Carleigh Klusman; Laurynn Garcia; Haydi Hassanein; Talal Abdel Aal; Mohamed M Shahin
Journal:  Semin Intervent Radiol       Date:  2022-02-18       Impact factor: 1.513

Review 2.  A Comparison Between the Efficacy and Safety of Endovascular Arteriovenous Fistula Creation and Surgical Fistula Creation: A Systematic Review and Meta-Analysis.

Authors:  Yoshinosuke Shimamura; Yasutaka Kuniyoshi; Hiroshi Ueta; Takamasa Miyauchi; Hajime Yamazaki; Yasushi Tsujimoto
Journal:  Cureus       Date:  2022-05-17

3.  Two-year cumulative patency of endovascular arteriovenous fistula.

Authors:  Gerald A Beathard; Terry Litchfield; William C Jennings
Journal:  J Vasc Access       Date:  2019-09-28       Impact factor: 2.283

4.  Clinical hemodialysis experience with percutaneous arteriovenous fistulas created using the Ellipsys® vascular access system.

Authors:  Hedia Hebibi; Jedjiga Achiche; Gilbert Franco; Jacques Rottembourg
Journal:  Hemodial Int       Date:  2019-03-01       Impact factor: 1.812

5.  Feasibility for arteriovenous fistula creation with Ellipsys®.

Authors:  Gilbert Franco; Alexandros Mallios; Pierre Bourquelot; Hadia Hebibi; William Jennings; Benoit Boura
Journal:  J Vasc Access       Date:  2020-01-13       Impact factor: 2.283

6.  Ultrasound evaluation of percutaneously created arteriovenous fistulae between radial artery and perforating vein at the elbow.

Authors:  Gilbert Franco; Alexandros Mallios; Pierre Bourquelot; William Jennings; Benoit Boura
Journal:  J Vasc Access       Date:  2020-01-10       Impact factor: 2.283

7.  Endovascular versus surgical creation of arteriovenous fistula in hemodialysis patients: Cost-effectiveness and budget impact analyses.

Authors:  Carla Rognoni; Matteo Tozzi; Rosanna Tarricone
Journal:  J Vasc Access       Date:  2020-05-19       Impact factor: 2.283

8.  Percutaneous endovascular arteriovenous fistula: A systematic review and meta-analysis.

Authors:  Ji-Bo Sun; Chun-Cheng Liu; Xi Shen; Qin Chen; Cheng-Liang Xu; Tian-Lei Cui
Journal:  Front Cardiovasc Med       Date:  2022-09-06
  8 in total